Optimal implantation depth and adherence to guidelines on permanent pacing to improve the results of transcatheter aortic valve replacement with the medtronic corevalve system
The aim of the CoreValve prospective, international, post-market ADVANCE-II study was to define the rates of conduction disturbances and permanent pacemaker implantation (PPI) after transcatheter aortic valve replacement with the Medtronic CoreValve System (Minneapolis, Minnesota) using optimized im...
Gespeichert in:
| Hauptverfasser: | , |
|---|---|
| Dokumenttyp: | Article (Journal) |
| Sprache: | Englisch |
| Veröffentlicht: |
18 May 2015
|
| In: |
JACC Cardiovascular interventions
Year: 2015, Jahrgang: 8, Heft: 6, Pages: 837-846$10 |
| ISSN: | 1876-7605 |
| DOI: | 10.1016/j.jcin.2015.02.005 |
| Online-Zugang: | Verlag, kostenfrei, Volltext: http://dx.doi.org/10.1016/j.jcin.2015.02.005 Verlag, kostenfrei, Volltext: http://www.sciencedirect.com/science/article/pii/S1936879815002514 |
| Verfasserangaben: | Anna S. Petronio, MD, Jan-Malte Sinning, MD, Nicolas Van Mieghem, MD, Giulio Zucchelli, MD, PhD, Georg Nickenig, MD, Raffi Bekeredjian, MD, PhD, Johan Bosmans, MD, PhD, Francesco Bedogni, MD, Marian Branny, MD, Karl Stangl, MD, Jan Kovac, MD, Molly Schiltgen, MS, Stacia Kraus, MPH, Peter de Jaegere, MD,PhD |
MARC
| LEADER | 00000caa a2200000 c 4500 | ||
|---|---|---|---|
| 001 | 1562336584 | ||
| 003 | DE-627 | ||
| 005 | 20250206093958.0 | ||
| 007 | cr uuu---uuuuu | ||
| 008 | 170810s2015 xx |||||o 00| ||eng c | ||
| 024 | 7 | |a 10.1016/j.jcin.2015.02.005 |2 doi | |
| 035 | |a (DE-627)1562336584 | ||
| 035 | |a (DE-576)49233658X | ||
| 035 | |a (DE-599)BSZ49233658X | ||
| 035 | |a (OCoLC)1340978453 | ||
| 040 | |a DE-627 |b ger |c DE-627 |e rda | ||
| 041 | |a eng | ||
| 084 | |a 33 |2 sdnb | ||
| 100 | 1 | |a Petronio, Anna Sonia |e VerfasserIn |0 (DE-588)1138058726 |0 (DE-627)895428911 |0 (DE-576)49233577X |4 aut | |
| 245 | 1 | 0 | |a Optimal implantation depth and adherence to guidelines on permanent pacing to improve the results of transcatheter aortic valve replacement with the medtronic corevalve system |c Anna S. Petronio, MD, Jan-Malte Sinning, MD, Nicolas Van Mieghem, MD, Giulio Zucchelli, MD, PhD, Georg Nickenig, MD, Raffi Bekeredjian, MD, PhD, Johan Bosmans, MD, PhD, Francesco Bedogni, MD, Marian Branny, MD, Karl Stangl, MD, Jan Kovac, MD, Molly Schiltgen, MS, Stacia Kraus, MPH, Peter de Jaegere, MD,PhD |
| 264 | 1 | |c 18 May 2015 | |
| 336 | |a Text |b txt |2 rdacontent | ||
| 337 | |a Computermedien |b c |2 rdamedia | ||
| 338 | |a Online-Ressource |b cr |2 rdacarrier | ||
| 500 | |a Gesehen am 10.08.2017 | ||
| 520 | |a The aim of the CoreValve prospective, international, post-market ADVANCE-II study was to define the rates of conduction disturbances and permanent pacemaker implantation (PPI) after transcatheter aortic valve replacement with the Medtronic CoreValve System (Minneapolis, Minnesota) using optimized implantation techniques and application of international guidelines on cardiac pacing. Conduction disturbances are a frequent complication of transcatheter aortic valve replacement. The rates of PPI in the published reports vary according to bioprosthesis type and the indications for PPI. The primary endpoint was the 30-day incidence of PPI with Class I/II indications when the Medtronic CoreValve System was implanted at an optimal depth (≤6 mm below the aortic annulus). The timing and resolution of all new-onset conduction disturbances were analyzed. A total of 194 patients were treated. The overall rate of PPI for Class I/II indications was 18.2%. An optimal depth was reached in 43.2% of patients, with a nonsignificantly lower incidence of PPI in patients with depths ≤6 mm, compared with those with deeper implants (13.3% vs. 21.1%; p = 0.14). In a paired analysis, new-onset left bundle branch block and first-degree atrioventricular block occurred in 45.4% and 39.0% of patients, respectively, and resolved spontaneously within 30 days in 43.2% and 73.9%, respectively. In patients with new PPI, the rate of intrinsic sinus rhythm increased from 25.9% at 7 days to 59.3% at 30 days (p = 0.004). Optimal Medtronic CoreValve System deployment and adherence to international guidelines on cardiac pacing are associated with a lower rate of new PPI after transcatheter aortic valve replacement, compared with results reported in previous studies. (CoreValve Advance-II Study: Prospective International Post-Market Study [ADVANCE II]; NCT01624870) | ||
| 650 | 4 | |a aortic stenosis | |
| 650 | 4 | |a conduction disturbance | |
| 650 | 4 | |a permanent pacemaker implantation | |
| 650 | 4 | |a transcatheter aortic valve replacement | |
| 700 | 1 | |a Bekeredjian, Raffi |d 1972- |e VerfasserIn |0 (DE-588)122088271 |0 (DE-627)705767930 |0 (DE-576)293085781 |4 aut | |
| 773 | 0 | 8 | |i Enthalten in |a American College of Cardiology |t JACC Cardiovascular interventions |d New York, NY : Elsevier, 2008 |g 8(2015), 6, Seite 837-846$10 |h Online-Ressource |w (DE-627)578539160 |w (DE-600)2452163-2 |w (DE-576)294403027 |x 1876-7605 |7 nnas |
| 773 | 1 | 8 | |g volume:8 |g year:2015 |g number:6 |g pages:837-846$10 |a Optimal implantation depth and adherence to guidelines on permanent pacing to improve the results of transcatheter aortic valve replacement with the medtronic corevalve system |
| 856 | 4 | 0 | |u http://dx.doi.org/10.1016/j.jcin.2015.02.005 |x Verlag |x Resolving-System |z kostenfrei |3 Volltext |
| 856 | 4 | 0 | |u http://www.sciencedirect.com/science/article/pii/S1936879815002514 |x Verlag |z kostenfrei |3 Volltext |
| 951 | |a AR | ||
| 992 | |a 20170810 | ||
| 993 | |a Article | ||
| 994 | |a 2015 | ||
| 998 | |g 122088271 |a Bekeredjian, Raffi |m 122088271:Bekeredjian, Raffi |d 910000 |d 910100 |e 910000PB122088271 |e 910100PB122088271 |k 0/910000/ |k 1/910000/910100/ |p 6 | ||
| 999 | |a KXP-PPN1562336584 |e 2977342451 | ||
| BIB | |a Y | ||
| SER | |a journal | ||
| JSO | |a {"note":["Gesehen am 10.08.2017"],"type":{"bibl":"article-journal","media":"Online-Ressource"},"relHost":[{"title":[{"title_sort":"JACC Cardiovascular interventions","subtitle":"a journal of the American College of Cardiology","title":"JACC Cardiovascular interventions"}],"part":{"volume":"8","text":"8(2015), 6, Seite 837-846$10","issue":"6","pages":"837-846$10","year":"2015"},"pubHistory":["1.2008 -"],"recId":"578539160","corporate":[{"roleDisplay":"VerfasserIn","display":"American College of Cardiology","role":"aut"}],"language":["eng"],"disp":"American College of CardiologyJACC Cardiovascular interventions","note":["Gesehen am 05.06.23","Fortsetzung der Druck-Ausgabe"],"type":{"media":"Online-Ressource","bibl":"periodical"},"id":{"issn":["1876-7605"],"zdb":["2452163-2"],"eki":["578539160"]},"origin":[{"dateIssuedDisp":"2008-","publisher":"Elsevier ; American College of Cardiology","dateIssuedKey":"2008","publisherPlace":"New York, NY ; New York, NY"}],"physDesc":[{"extent":"Online-Ressource"}]}],"language":["eng"],"recId":"1562336584","origin":[{"dateIssuedKey":"2015","dateIssuedDisp":"18 May 2015"}],"title":[{"title":"Optimal implantation depth and adherence to guidelines on permanent pacing to improve the results of transcatheter aortic valve replacement with the medtronic corevalve system","title_sort":"Optimal implantation depth and adherence to guidelines on permanent pacing to improve the results of transcatheter aortic valve replacement with the medtronic corevalve system"}],"id":{"doi":["10.1016/j.jcin.2015.02.005"],"eki":["1562336584"]},"name":{"displayForm":["Anna S. Petronio, MD, Jan-Malte Sinning, MD, Nicolas Van Mieghem, MD, Giulio Zucchelli, MD, PhD, Georg Nickenig, MD, Raffi Bekeredjian, MD, PhD, Johan Bosmans, MD, PhD, Francesco Bedogni, MD, Marian Branny, MD, Karl Stangl, MD, Jan Kovac, MD, Molly Schiltgen, MS, Stacia Kraus, MPH, Peter de Jaegere, MD,PhD"]},"person":[{"roleDisplay":"VerfasserIn","display":"Petronio, Anna Sonia","role":"aut","family":"Petronio","given":"Anna Sonia"},{"role":"aut","roleDisplay":"VerfasserIn","display":"Bekeredjian, Raffi","given":"Raffi","family":"Bekeredjian"}]} | ||
| SRT | |a PETRONIOANOPTIMALIMP1820 | ||