Demonstration of a scalable frequency-domain readout of metallic magnetic calorimeters by means of a microwave SQUID multiplexer

We report on the first demonstration of a scalable GHz frequency-domain readout of metallic magnetic calorimeters (MMCs) using a 64 pixel detector array that is read out by an integrated, on-chip microwave SQUID multiplexer. The detector array is optimized for detecting soft X-ray photons and the mu...

Full description

Saved in:
Bibliographic Details
Main Authors: Kempf, Sebastian (Author) , Wegner, Mathias (Author) , Fleischmann, Andreas (Author) , Gastaldo, Loredana (Author) , Papst, Maximilian (Author) , Herrmann, Felix (Author) , Richter, Daniel Philipp (Author) , Enss, Christian (Author)
Format: Article (Journal)
Language:English
Published: 5 January 2017
In: AIP Advances
Year: 2017, Volume: 7, Issue: 1
ISSN:2158-3226
DOI:10.1063/1.4973872
Online Access:Verlag, kostenfrei, Volltext: http://dx.doi.org/10.1063/1.4973872
Verlag, kostenfrei, Volltext: https://aip.scitation.org/doi/10.1063/1.4973872
Get full text
Author Notes:Sebastian Kempf, Mathias Wegner, Andreas Fleischmann, Loredana Gastaldo, Felix Herrmann, Maximilian Papst, Daniel Richter, and Christian Enss

MARC

LEADER 00000caa a2200000 c 4500
001 1571426183
003 DE-627
005 20220814101650.0
007 cr uuu---uuuuu
008 180326s2017 xx |||||o 00| ||eng c
024 7 |a 10.1063/1.4973872  |2 doi 
035 |a (DE-627)1571426183 
035 |a (DE-576)501426183 
035 |a (DE-599)BSZ501426183 
035 |a (OCoLC)1340995486 
040 |a DE-627  |b ger  |c DE-627  |e rda 
041 |a eng 
084 |a 29  |2 sdnb 
100 1 |a Kempf, Sebastian  |d 1983-  |e VerfasserIn  |0 (DE-588)1020440449  |0 (DE-627)69122143X  |0 (DE-576)360448690  |4 aut 
245 1 0 |a Demonstration of a scalable frequency-domain readout of metallic magnetic calorimeters by means of a microwave SQUID multiplexer  |c Sebastian Kempf, Mathias Wegner, Andreas Fleischmann, Loredana Gastaldo, Felix Herrmann, Maximilian Papst, Daniel Richter, and Christian Enss 
264 1 |c 5 January 2017 
300 |a 9 
336 |a Text  |b txt  |2 rdacontent 
337 |a Computermedien  |b c  |2 rdamedia 
338 |a Online-Ressource  |b cr  |2 rdacarrier 
500 |a Gesehen am 26.03.2018 
520 |a We report on the first demonstration of a scalable GHz frequency-domain readout of metallic magnetic calorimeters (MMCs) using a 64 pixel detector array that is read out by an integrated, on-chip microwave SQUID multiplexer. The detector array is optimized for detecting soft X-ray photons and the multiplexer is designed to provide a signal rise time τrise<400nsτrise<400ns<math display="inline" overflow="scroll" altimg="eq-00001.gif"><mrow><msub><mi>τ</mi><mi>rise</mi></msub><mo><</mo><mrow><mn>400</mn><mi>ns</mi></mrow></mrow></math> and an intrinsic energy sensitivity ϵ<30hϵ<30h<math display="inline" overflow="scroll" altimg="eq-00002.gif"><mrow><mi>ϵ</mi><mo><</mo><mrow><mn>30</mn><mi>h</mi></mrow></mrow></math>. This results in an expected energy resolution ΔEFWHM<10eVΔEFWHM<10eV<math display="inline" overflow="scroll" altimg="eq-00003.gif"><mrow><mrow><mi mathvariant="normal">Δ</mi><msub><mi>E</mi><mi>FWHM</mi></msub></mrow><mo><</mo><mrow><mn>10</mn><mi>eV</mi></mrow></mrow></math>. We measured a signal rise time τriseτrise<math display="inline" overflow="scroll" altimg="eq-00004.gif"><msub><mi>τ</mi><mi>rise</mi></msub></math> as low as 90ns90ns<math display="inline" overflow="scroll" altimg="eq-00005.gif"><mrow><mn>90</mn><mi>ns</mi></mrow></math> and an energy resolution ΔEFWHMΔEFWHM<math display="inline" overflow="scroll" altimg="eq-00006.gif"><mrow><mi mathvariant="normal">Δ</mi><msub><mi>E</mi><mi>FWHM</mi></msub></mrow></math> as low as 50eV50eV<math display="inline" overflow="scroll" altimg="eq-00007.gif"><mrow><mn>50</mn><mi>eV</mi></mrow></math> for 5.9keV5.9keV<math display="inline" overflow="scroll" altimg="eq-00008.gif"><mrow><mn>5.9</mn><mi>keV</mi></mrow></math> photons. The rise time is about an order of magnitude faster compared to other multiplexed low-temperature microcalorimeters and close to the intrinsic value set by the coupling between electron and spins. The energy resolution is degraded with respect to our design value due to a rather low intrinsic quality factor of the microwave resonators that is caused by the quality of the Josephson junction of the associated rf-SQUID as well as an elevated chip temperature as compared to the heat bath. Though the achieved energy resolution is not yet compatible with state-of-the-art single-channel MMCs, this demonstration of a scalable readout approach for MMCs in combination with the full understanding of the device performance showing ways how to improve represents an important milestone for the development of future large-scale MMC detector arrays. 
700 1 |a Wegner, Mathias  |d 1986-  |e VerfasserIn  |0 (DE-588)1176685597  |0 (DE-627)1047656418  |0 (DE-576)516756893  |4 aut 
700 1 |a Fleischmann, Andreas  |e VerfasserIn  |0 (DE-588)1020467568  |0 (DE-627)691226873  |0 (DE-576)360449409  |4 aut 
700 1 |a Gastaldo, Loredana  |e VerfasserIn  |0 (DE-588)1020440317  |0 (DE-627)691221340  |0 (DE-576)360449603  |4 aut 
700 1 |a Papst, Maximilian  |e VerfasserIn  |0 (DE-588)1155184068  |0 (DE-627)1016517394  |0 (DE-576)501456252  |4 aut 
700 1 |a Herrmann, Felix  |d 1994-  |e VerfasserIn  |0 (DE-588)1155184351  |0 (DE-627)1016517750  |0 (DE-576)501456619  |4 aut 
700 1 |a Richter, Daniel Philipp  |d 1992-  |e VerfasserIn  |0 (DE-588)115518372X  |0 (DE-627)1016516797  |0 (DE-576)501425225  |4 aut 
700 1 |a Enss, Christian  |d 1960-  |e VerfasserIn  |0 (DE-588)1020467878  |0 (DE-627)691227020  |0 (DE-576)360449530  |4 aut 
773 0 8 |i Enthalten in  |t AIP Advances  |d New York, NY : American Inst. of Physics, 2011  |g 7(2017,1) Artikel-Nummer 015007, 9 Seiten  |h Online-Ressource  |w (DE-627)641391706  |w (DE-600)2583909-3  |w (DE-576)33467087X  |x 2158-3226  |7 nnas  |a Demonstration of a scalable frequency-domain readout of metallic magnetic calorimeters by means of a microwave SQUID multiplexer 
773 1 8 |g volume:7  |g year:2017  |g number:1  |g extent:9  |a Demonstration of a scalable frequency-domain readout of metallic magnetic calorimeters by means of a microwave SQUID multiplexer 
856 4 0 |u http://dx.doi.org/10.1063/1.4973872  |x Verlag  |x Resolving-System  |z kostenfrei  |3 Volltext 
856 4 0 |u https://aip.scitation.org/doi/10.1063/1.4973872  |x Verlag  |z kostenfrei  |3 Volltext 
951 |a AR 
992 |a 20180326 
993 |a Article 
994 |a 2017 
998 |g 1155184351  |a Herrmann, Felix  |m 1155184351:Herrmann, Felix  |d 130000  |d 130700  |e 130000PH1155184351  |e 130700PH1155184351  |k 0/130000/  |k 1/130000/130700/  |p 5 
998 |g 1155184068  |a Papst, Maximilian  |m 1155184068:Papst, Maximilian  |d 130000  |d 130700  |e 130000PP1155184068  |e 130700PP1155184068  |k 0/130000/  |k 1/130000/130700/  |p 6 
998 |g 1020467878  |a Enss, Christian  |m 1020467878:Enss, Christian  |d 130000  |d 130700  |e 130000PE1020467878  |e 130700PE1020467878  |k 0/130000/  |k 1/130000/130700/  |p 8  |y j 
998 |g 115518372X  |a Richter, Daniel Philipp  |m 115518372X:Richter, Daniel Philipp  |d 130000  |d 130700  |e 130000PR115518372X  |e 130700PR115518372X  |k 0/130000/  |k 1/130000/130700/  |p 7 
998 |g 1020440317  |a Gastaldo, Loredana  |m 1020440317:Gastaldo, Loredana  |d 130000  |d 130700  |e 130000PG1020440317  |e 130700PG1020440317  |k 0/130000/  |k 1/130000/130700/  |p 4 
998 |g 1020467568  |a Fleischmann, Andreas  |m 1020467568:Fleischmann, Andreas  |d 130000  |d 130700  |e 130000PF1020467568  |e 130700PF1020467568  |k 0/130000/  |k 1/130000/130700/  |p 3 
998 |g 1176685597  |a Wegner, Mathias  |m 1176685597:Wegner, Mathias  |d 130000  |d 130700  |e 130000PW1176685597  |e 130700PW1176685597  |k 0/130000/  |k 1/130000/130700/  |p 2 
998 |g 1020440449  |a Kempf, Sebastian  |m 1020440449:Kempf, Sebastian  |d 130000  |d 130700  |e 130000PK1020440449  |e 130700PK1020440449  |k 0/130000/  |k 1/130000/130700/  |p 1  |x j 
999 |a KXP-PPN1571426183  |e 3004308665 
BIB |a Y 
SER |a journal 
JSO |a {"note":["Gesehen am 26.03.2018"],"relHost":[{"recId":"641391706","disp":"Demonstration of a scalable frequency-domain readout of metallic magnetic calorimeters by means of a microwave SQUID multiplexerAIP Advances","language":["eng"],"pubHistory":["1.2011 -"],"type":{"bibl":"periodical","media":"Online-Ressource"},"origin":[{"publisher":"American Inst. of Physics","dateIssuedDisp":"2011-","dateIssuedKey":"2011","publisherPlace":"New York, NY"}],"name":{"displayForm":["American Institute of Physics"]},"corporate":[{"display":"American Institute of Physics","role":"isb"}],"title":[{"title":"AIP Advances","title_sort":"AIP Advances"}],"id":{"issn":["2158-3226"],"zdb":["2583909-3"],"eki":["641391706"]},"part":{"issue":"1","extent":"9","year":"2017","volume":"7","text":"7(2017,1) Artikel-Nummer 015007, 9 Seiten"},"physDesc":[{"extent":"Online-Ressource"}],"note":["Gesehen am 25.04.2023"]}],"physDesc":[{"extent":"9 S."}],"id":{"eki":["1571426183"],"doi":["10.1063/1.4973872"]},"type":{"bibl":"article-journal","media":"Online-Ressource"},"title":[{"title":"Demonstration of a scalable frequency-domain readout of metallic magnetic calorimeters by means of a microwave SQUID multiplexer","title_sort":"Demonstration of a scalable frequency-domain readout of metallic magnetic calorimeters by means of a microwave SQUID multiplexer"}],"language":["eng"],"person":[{"role":"aut","family":"Kempf","given":"Sebastian","display":"Kempf, Sebastian"},{"family":"Wegner","given":"Mathias","display":"Wegner, Mathias","role":"aut"},{"role":"aut","family":"Fleischmann","given":"Andreas","display":"Fleischmann, Andreas"},{"role":"aut","family":"Gastaldo","given":"Loredana","display":"Gastaldo, Loredana"},{"role":"aut","given":"Maximilian","display":"Papst, Maximilian","family":"Papst"},{"role":"aut","family":"Herrmann","given":"Felix","display":"Herrmann, Felix"},{"display":"Richter, Daniel Philipp","given":"Daniel Philipp","family":"Richter","role":"aut"},{"family":"Enss","given":"Christian","display":"Enss, Christian","role":"aut"}],"origin":[{"dateIssuedKey":"2017","dateIssuedDisp":"5 January 2017"}],"name":{"displayForm":["Sebastian Kempf, Mathias Wegner, Andreas Fleischmann, Loredana Gastaldo, Felix Herrmann, Maximilian Papst, Daniel Richter, and Christian Enss"]},"recId":"1571426183"} 
SRT |a KEMPFSEBASDEMONSTRAT5201