Accuracy assessment for navigated maxillo-facial surgery using an electromagnetic tracking device
Purpose: To evaluate the accuracy and the usability of an electromagnetic tracking device in maxillo-facial surgery through testing on a phantom skull under operating room (OR) conditions. Material and methods. A standard plastic skull phantom was equipped with a custom made model of the maxilla and...
Saved in:
| Main Authors: | , , , |
|---|---|
| Format: | Article (Journal) |
| Language: | English |
| Published: |
2012
|
| In: |
Journal of cranio-maxillofacial surgery
Year: 2011, Volume: 40, Issue: 2, Pages: 156-161 |
| ISSN: | 1878-4119 |
| DOI: | 10.1016/j.jcms.2011.03.003 |
| Online Access: | Verlag, Volltext: http://dx.doi.org/10.1016/j.jcms.2011.03.003 Verlag, Volltext: http://www.sciencedirect.com/science/article/pii/S1010518211000461 |
| Author Notes: | Robin Seeberger, Gavin Kane, Juergen Hoffmann, Georg Eggers |
MARC
| LEADER | 00000caa a2200000 c 4500 | ||
|---|---|---|---|
| 001 | 1576888118 | ||
| 003 | DE-627 | ||
| 005 | 20220814175256.0 | ||
| 007 | cr uuu---uuuuu | ||
| 008 | 180627r20122011xx |||||o 00| ||eng c | ||
| 024 | 7 | |a 10.1016/j.jcms.2011.03.003 |2 doi | |
| 035 | |a (DE-627)1576888118 | ||
| 035 | |a (DE-576)506888118 | ||
| 035 | |a (DE-599)BSZ506888118 | ||
| 035 | |a (OCoLC)1341012957 | ||
| 040 | |a DE-627 |b ger |c DE-627 |e rda | ||
| 041 | |a eng | ||
| 084 | |a 33 |2 sdnb | ||
| 100 | 1 | |a Seeberger, Robin |d 1976- |e VerfasserIn |0 (DE-588)136014216 |0 (DE-627)574409211 |0 (DE-576)300779240 |4 aut | |
| 245 | 1 | 0 | |a Accuracy assessment for navigated maxillo-facial surgery using an electromagnetic tracking device |c Robin Seeberger, Gavin Kane, Juergen Hoffmann, Georg Eggers |
| 264 | 1 | |c 2012 | |
| 300 | |a 6 | ||
| 336 | |a Text |b txt |2 rdacontent | ||
| 337 | |a Computermedien |b c |2 rdamedia | ||
| 338 | |a Online-Ressource |b cr |2 rdacarrier | ||
| 500 | |a Published online: 1 April 2011 | ||
| 500 | |a Gesehen am 27.06.2018 | ||
| 520 | |a Purpose: To evaluate the accuracy and the usability of an electromagnetic tracking device in maxillo-facial surgery through testing on a phantom skull under operating room (OR) conditions. Material and methods. A standard plastic skull phantom was equipped with a custom made model of the maxilla and with target markers and dental brackets. Imaging was performed with a computed tomography (CT) scanner. The extent and robustness of the electromagnetic tracking system’s target registration error (TRE) was evaluated under various conditions. Results: For each measurement a total of 243 registrations were performed with 5 point registration and 4374 registrations with 6 point registration. The average target registration error for the 5 point registration under OR conditions was 2.1mm (SD 0.86) and 1.03 (SD 0.53) for the 6 point registration. Metallic instruments applied to the skull increased the TRE significantly in both registration methods. Conclusion: The electromagnetic tracking device showed a high accuracy and performed stable in both registration methods. Electromagnetic interference due to metallic instruments was significant but the extent of TRE was still acceptable in comparison to optical navigation devices. A benefit of EM tracking is the absence of line-of-sight hindrance. The test setting simulating OR conditions has proven suitable for further studies. | ||
| 534 | |c 2011 | ||
| 650 | 4 | |a Electromagnetic tracing | |
| 650 | 4 | |a Image guided surgery | |
| 650 | 4 | |a Maxillo-facial surgery | |
| 650 | 4 | |a Navigation | |
| 700 | 1 | |a Kane, Gavin J. |e VerfasserIn |0 (DE-588)1026155495 |0 (DE-627)726203548 |0 (DE-576)371429498 |4 aut | |
| 700 | 1 | |a Hoffmann, Jürgen |d 1965- |e VerfasserIn |0 (DE-588)1032524456 |0 (DE-627)738520063 |0 (DE-576)380087189 |4 aut | |
| 700 | 1 | |a Eggers, Georg |e VerfasserIn |4 aut | |
| 773 | 0 | 8 | |i Enthalten in |t Journal of cranio-maxillofacial surgery |d Oxford [u.a.] : Elsevier, 1987 |g 40(2012), 2, Seite 156-161 |h Online-Ressource |w (DE-627)320479218 |w (DE-600)2009565-X |w (DE-576)10374696X |x 1878-4119 |7 nnas |a Accuracy assessment for navigated maxillo-facial surgery using an electromagnetic tracking device |
| 773 | 1 | 8 | |g volume:40 |g year:2012 |g number:2 |g pages:156-161 |g extent:6 |a Accuracy assessment for navigated maxillo-facial surgery using an electromagnetic tracking device |
| 856 | 4 | 0 | |u http://dx.doi.org/10.1016/j.jcms.2011.03.003 |x Verlag |x Resolving-System |3 Volltext |
| 856 | 4 | 0 | |u http://www.sciencedirect.com/science/article/pii/S1010518211000461 |x Verlag |3 Volltext |
| 951 | |a AR | ||
| 992 | |a 20180627 | ||
| 993 | |a Article | ||
| 994 | |a 2012 | ||
| 998 | |g 1032524456 |a Hoffmann, Jürgen |m 1032524456:Hoffmann, Jürgen |d 910000 |d 910800 |e 910000PH1032524456 |e 910800PH1032524456 |k 0/910000/ |k 1/910000/910800/ |p 3 | ||
| 998 | |g 136014216 |a Seeberger, Robin |m 136014216:Seeberger, Robin |d 50000 |e 50000PS136014216 |k 0/50000/ |p 1 |x j | ||
| 999 | |a KXP-PPN1576888118 |e 301386298X | ||
| BIB | |a Y | ||
| SER | |a journal | ||
| JSO | |a {"language":["eng"],"physDesc":[{"extent":"6 S."}],"type":{"media":"Online-Ressource","bibl":"article-journal"},"person":[{"display":"Seeberger, Robin","role":"aut","given":"Robin","family":"Seeberger"},{"given":"Gavin J.","role":"aut","family":"Kane","display":"Kane, Gavin J."},{"family":"Hoffmann","role":"aut","given":"Jürgen","display":"Hoffmann, Jürgen"},{"family":"Eggers","given":"Georg","role":"aut","display":"Eggers, Georg"}],"title":[{"title":"Accuracy assessment for navigated maxillo-facial surgery using an electromagnetic tracking device","title_sort":"Accuracy assessment for navigated maxillo-facial surgery using an electromagnetic tracking device"}],"name":{"displayForm":["Robin Seeberger, Gavin Kane, Juergen Hoffmann, Georg Eggers"]},"recId":"1576888118","id":{"eki":["1576888118"],"doi":["10.1016/j.jcms.2011.03.003"]},"origin":[{"dateIssuedKey":"2012","dateIssuedDisp":"2012"}],"note":["Published online: 1 April 2011","Gesehen am 27.06.2018"],"relHost":[{"note":["Gesehen am 11.06.18"],"type":{"media":"Online-Ressource","bibl":"periodical"},"language":["eng"],"pubHistory":["Nachgewiesen 15.1987 -"],"physDesc":[{"extent":"Online-Ressource"}],"title":[{"title":"Journal of cranio-maxillofacial surgery","subtitle":"official publication of the European Association for Cranio-Maxillofacial Surgery","title_sort":"Journal of cranio-maxillofacial surgery"}],"recId":"320479218","disp":"Accuracy assessment for navigated maxillo-facial surgery using an electromagnetic tracking deviceJournal of cranio-maxillofacial surgery","origin":[{"dateIssuedKey":"1987","publisher":"Elsevier ; Churchill Livingstone","dateIssuedDisp":"1987-","publisherPlace":"Oxford [u.a.] ; Edinburgh [u.a.]"}],"id":{"eki":["320479218"],"issn":["1878-4119"],"zdb":["2009565-X"]},"part":{"text":"40(2012), 2, Seite 156-161","extent":"6","pages":"156-161","volume":"40","year":"2012","issue":"2"}}]} | ||
| SRT | |a SEEBERGERRACCURACYAS2012 | ||