Characterization and clinical evaluation of a novel 2D detector array for conventional and flattening filter free (FFF) IMRT pre-treatment verification

Background and purpose: The novel MatriXXFFF (IBA Dosimetry, Germany) detector is a new 2D ionization chamber detector array designed for patient specific IMRT-plan verification including flattening-filter-free (FFF) beams. This study provides a detailed analysis of the characterization and clinical...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Sekar, Yuvaraj (VerfasserIn) , Thölking, Johannes (VerfasserIn) , Eckl, Miriam (VerfasserIn) , Kalichava, Irakli (VerfasserIn) , Sihono, Dwi Seno Kuncoro (VerfasserIn) , Wenz, Frederik (VerfasserIn) , Wertz, Hansjörg (VerfasserIn)
Dokumenttyp: Article (Journal)
Sprache:Englisch
Veröffentlicht: 2018
In: Zeitschrift für medizinische Physik
Year: 2018, Jahrgang: 28, Heft: 2, Pages: 134-141
ISSN:1876-4436
DOI:10.1016/j.zemedi.2017.08.003
Online-Zugang:Verlag, Volltext: http://dx.doi.org/10.1016/j.zemedi.2017.08.003
Verlag, Volltext: http://www.sciencedirect.com/science/article/pii/S0939388917300466
Volltext
Verfasserangaben:Yuvaraj Sekar, Johannes Thoelking, Miriam Eckl, Irakli Kalichava, Dwi Seno Kuncoro Sihono, Frank Lohr, Frederik Wenz, Hansjoerg Wertz

MARC

LEADER 00000caa a2200000 c 4500
001 1577186826
003 DE-627
005 20230508120554.0
007 cr uuu---uuuuu
008 180702s2018 xx |||||o 00| ||eng c
024 7 |a 10.1016/j.zemedi.2017.08.003  |2 doi 
035 |a (DE-627)1577186826 
035 |a (DE-576)507186826 
035 |a (DE-599)BSZ507186826 
035 |a (OCoLC)1341013215 
040 |a DE-627  |b ger  |c DE-627  |e rda 
041 |a eng 
084 |a 33  |2 sdnb 
100 1 |a Sekar, Yuvaraj  |e VerfasserIn  |0 (DE-588)1162098341  |0 (DE-627)1025541200  |0 (DE-576)507156498  |4 aut 
245 1 0 |a Characterization and clinical evaluation of a novel 2D detector array for conventional and flattening filter free (FFF) IMRT pre-treatment verification  |c Yuvaraj Sekar, Johannes Thoelking, Miriam Eckl, Irakli Kalichava, Dwi Seno Kuncoro Sihono, Frank Lohr, Frederik Wenz, Hansjoerg Wertz 
264 1 |c 2018 
300 |a 8 
336 |a Text  |b txt  |2 rdacontent 
337 |a Computermedien  |b c  |2 rdamedia 
338 |a Online-Ressource  |b cr  |2 rdacarrier 
500 |a Available online 11 October 2017 
500 |a Gesehen am 02.07.2018 
520 |a Background and purpose: The novel MatriXXFFF (IBA Dosimetry, Germany) detector is a new 2D ionization chamber detector array designed for patient specific IMRT-plan verification including flattening-filter-free (FFF) beams. This study provides a detailed analysis of the characterization and clinical evaluation of the new detector array. Material and methods: The verification of the MatriXXFFF was subdivided into (i) physical dosimetric tests including dose linearity, dose rate dependency and output factor measurements and (ii) patient specific IMRT pre-treatment plan verifications. The MatriXXFFF measurements were compared to the calculated dose distribution of a commissioned treatment planning system by gamma index and dose difference evaluations for 18 IMRT-sequences. All IMRT-sequences were measured with original gantry angles and with collapsing all beams to 0° gantry angle to exclude the influence of the detector's angle dependency. Results: The MatriXXFFF was found to be linear and dose rate independent for all investigated modalities (deviations ≤0.6%). Furthermore, the output measurements of the MatriXXFFF were in very good agreement to reference measurements (deviations ≤1.8%). For the clinical evaluation an average pixel passing rate for γ(3%,3mm) of (98.5±1.5)% was achieved when applying a gantry angle correction. Also, with collapsing all beams to 0° gantry angle an excellent agreement to the calculated dose distribution was observed (γ(3%,3mm)=(99.1±1.1)%). Conclusions: The MatriXXFFF fulfills all physical requirements in terms of dosimetric accuracy. Furthermore, the evaluation of the IMRT-plan measurements showed that the detector particularly together with the gantry angle correction is a reliable device for IMRT-plan verification including FFF. 
650 4 |a 2D detector array 
650 4 |a Flattening filter free (FFF) 
650 4 |a IMRT-plan verification 
650 4 |a Quality assurance 
700 1 |a Thölking, Johannes  |d 1987-  |e VerfasserIn  |0 (DE-588)1069150924  |0 (DE-627)821387855  |0 (DE-576)428448615  |4 aut 
700 1 |a Eckl, Miriam  |d 1992-  |e VerfasserIn  |0 (DE-588)1162098538  |0 (DE-627)1025541596  |0 (DE-576)507186419  |4 aut 
700 1 |a Kalichava, Irakli  |d 1986-  |e VerfasserIn  |0 (DE-588)1083050540  |0 (DE-627)847862208  |0 (DE-576)455641552  |4 aut 
700 1 |a Sihono, Dwi Seno Kuncoro  |d 1979-  |e VerfasserIn  |0 (DE-588)1073337642  |0 (DE-627)829090754  |0 (DE-576)434936057  |4 aut 
700 1 |a Wenz, Frederik  |d 1966-  |e VerfasserIn  |0 (DE-588)113310390  |0 (DE-627)663837766  |0 (DE-576)346656281  |4 aut 
700 1 |a Wertz, Hansjörg  |d 1976-  |e VerfasserIn  |0 (DE-588)143861247  |0 (DE-627)70462401X  |0 (DE-576)339135778  |4 aut 
773 0 8 |i Enthalten in  |t Zeitschrift für medizinische Physik  |d Amsterdam [u.a.] : Elsevier, 1990  |g 28(2018), 2, Seite 134-141  |w (DE-627)510617662  |w (DE-600)2231492-1  |w (DE-576)272350818  |x 1876-4436  |7 nnas  |a Characterization and clinical evaluation of a novel 2D detector array for conventional and flattening filter free (FFF) IMRT pre-treatment verification 
773 1 8 |g volume:28  |g year:2018  |g number:2  |g pages:134-141  |g extent:8  |a Characterization and clinical evaluation of a novel 2D detector array for conventional and flattening filter free (FFF) IMRT pre-treatment verification 
856 4 0 |u http://dx.doi.org/10.1016/j.zemedi.2017.08.003  |x Verlag  |x Resolving-System  |3 Volltext 
856 4 0 |u http://www.sciencedirect.com/science/article/pii/S0939388917300466  |x Verlag  |3 Volltext 
951 |a AR 
992 |a 20180702 
993 |a Article 
994 |a 2018 
998 |g 143861247  |a Wertz, Hansjörg  |m 143861247:Wertz, Hansjörg  |d 60000  |d 63000  |e 60000PW143861247  |e 63000PW143861247  |k 0/60000/  |k 1/60000/63000/  |p 8  |y j 
998 |g 113310390  |a Wenz, Frederik  |m 113310390:Wenz, Frederik  |d 60000  |d 63000  |e 60000PW113310390  |e 63000PW113310390  |k 0/60000/  |k 1/60000/63000/  |p 7 
998 |g 1073337642  |a Sihono, Dwi Seno Kuncoro  |m 1073337642:Sihono, Dwi Seno Kuncoro  |d 60000  |d 63000  |e 60000PS1073337642  |e 63000PS1073337642  |k 0/60000/  |k 1/60000/63000/  |p 5 
998 |g 1083050540  |a Kalichava, Irakli  |m 1083050540:Kalichava, Irakli  |d 60000  |d 63000  |e 60000PK1083050540  |e 63000PK1083050540  |k 0/60000/  |k 1/60000/63000/  |p 4 
998 |g 1162098538  |a Eckl, Miriam  |m 1162098538:Eckl, Miriam  |d 60000  |d 63000  |e 60000PE1162098538  |e 63000PE1162098538  |k 0/60000/  |k 1/60000/63000/  |p 3 
998 |g 1069150924  |a Thölking, Johannes  |m 1069150924:Thölking, Johannes  |d 60000  |d 63000  |e 60000PT1069150924  |e 63000PT1069150924  |k 0/60000/  |k 1/60000/63000/  |p 2 
998 |g 1162098341  |a Sekar, Yuvaraj  |m 1162098341:Sekar, Yuvaraj  |d 60000  |d 63000  |e 60000PS1162098341  |e 63000PS1162098341  |k 0/60000/  |k 1/60000/63000/  |p 1  |x j 
999 |a KXP-PPN1577186826  |e 3015521594 
BIB |a Y 
SER |a journal 
JSO |a {"language":["eng"],"type":{"bibl":"article-journal","media":"Online-Ressource"},"note":["Available online 11 October 2017","Gesehen am 02.07.2018"],"origin":[{"dateIssuedKey":"2018","dateIssuedDisp":"2018"}],"title":[{"title_sort":"Characterization and clinical evaluation of a novel 2D detector array for conventional and flattening filter free (FFF) IMRT pre-treatment verification","title":"Characterization and clinical evaluation of a novel 2D detector array for conventional and flattening filter free (FFF) IMRT pre-treatment verification"}],"person":[{"display":"Sekar, Yuvaraj","family":"Sekar","given":"Yuvaraj","role":"aut"},{"given":"Johannes","role":"aut","display":"Thölking, Johannes","family":"Thölking"},{"family":"Eckl","display":"Eckl, Miriam","role":"aut","given":"Miriam"},{"role":"aut","given":"Irakli","display":"Kalichava, Irakli","family":"Kalichava"},{"display":"Sihono, Dwi Seno Kuncoro","family":"Sihono","given":"Dwi Seno Kuncoro","role":"aut"},{"given":"Frederik","role":"aut","display":"Wenz, Frederik","family":"Wenz"},{"family":"Wertz","display":"Wertz, Hansjörg","role":"aut","given":"Hansjörg"}],"id":{"eki":["1577186826"],"doi":["10.1016/j.zemedi.2017.08.003"]},"relHost":[{"corporate":[{"display":"Deutsche Gesellschaft für Medizinische Physik","role":"isb"},{"display":"Österreichische Gesellschaft für Medizinische Physik","role":"isb"}],"disp":"Characterization and clinical evaluation of a novel 2D detector array for conventional and flattening filter free (FFF) IMRT pre-treatment verificationZeitschrift für medizinische Physik","note":["Gesehen am 31.01.2022"],"origin":[{"publisherPlace":"Amsterdam [u.a.]","dateIssuedDisp":"1990-","dateIssuedKey":"1990","publisher":"Elsevier"}],"title":[{"title":"Zeitschrift für medizinische Physik","title_sort":"Zeitschrift für medizinische Physik","subtitle":"offizielles Organ der Deutschen, Österreichischen und Schweizerischen Gesellschaft für Medizinische Physik"}],"titleAlt":[{"title":"Medizinische Physik"}],"type":{"media":"Online-Ressource","bibl":"periodical"},"part":{"issue":"2","extent":"8","volume":"28","year":"2018","text":"28(2018), 2, Seite 134-141","pages":"134-141"},"language":["ger"],"pubHistory":["Volume 1, supplement (1990) ; volume 1, issue 1 (1991)-"],"recId":"510617662","id":{"issn":["1876-4436"],"eki":["510617662"],"zdb":["2231492-1"]}}],"recId":"1577186826","physDesc":[{"extent":"8 S."}],"name":{"displayForm":["Yuvaraj Sekar, Johannes Thoelking, Miriam Eckl, Irakli Kalichava, Dwi Seno Kuncoro Sihono, Frank Lohr, Frederik Wenz, Hansjoerg Wertz"]}} 
SRT |a SEKARYUVARCHARACTERI2018