Comparative Monte Carlo study on the performance of integration- and list-mode detector configurations for carbon ion computed tomography
Ion beam therapy offers the possibility of a highly conformal tumor-dose distribution; however, this technique is extremely sensitive to inaccuracies in the treatment procedures. Ambiguities in the conversion of Hounsfield units of the treatment planning x-ray CT to relative stopping power (RSP) can...
Saved in:
| Main Authors: | , , , , |
|---|---|
| Format: | Article (Journal) |
| Language: | English |
| Published: |
16 January 2017
|
| In: |
Physics in medicine and biology
Year: 2017, Volume: 62, Issue: 3, Pages: 1096-1112 |
| ISSN: | 1361-6560 |
| DOI: | 10.1088/1361-6560/aa5602 |
| Online Access: | Verlag, Volltext: http://dx.doi.org/10.1088/1361-6560/aa5602 Verlag, Volltext: http://stacks.iop.org/0031-9155/62/i=3/a=1096 |
| Author Notes: | Sebastian Meyer, Chiara Gianoli, Lorena Magallanes, Benedikt Kopp, Thomas Tessonnier, Guillaume Landry, George Dedes, Bernd Voss and Katia Parodi |
MARC
| LEADER | 00000caa a2200000 c 4500 | ||
|---|---|---|---|
| 001 | 1582373981 | ||
| 003 | DE-627 | ||
| 005 | 20230427193212.0 | ||
| 007 | cr uuu---uuuuu | ||
| 008 | 181029s2017 xx |||||o 00| ||eng c | ||
| 024 | 7 | |a 10.1088/1361-6560/aa5602 |2 doi | |
| 035 | |a (DE-627)1582373981 | ||
| 035 | |a (DE-576)512373981 | ||
| 035 | |a (DE-599)BSZ512373981 | ||
| 035 | |a (OCoLC)1341021759 | ||
| 040 | |a DE-627 |b ger |c DE-627 |e rda | ||
| 041 | |a eng | ||
| 084 | |a 33 |2 sdnb | ||
| 100 | 1 | |a Meyer, Sebastian |e VerfasserIn |0 (DE-588)1034705458 |0 (DE-627)746184395 |0 (DE-576)382337093 |4 aut | |
| 245 | 1 | 0 | |a Comparative Monte Carlo study on the performance of integration- and list-mode detector configurations for carbon ion computed tomography |c Sebastian Meyer, Chiara Gianoli, Lorena Magallanes, Benedikt Kopp, Thomas Tessonnier, Guillaume Landry, George Dedes, Bernd Voss and Katia Parodi |
| 264 | 1 | |c 16 January 2017 | |
| 300 | |a 18 | ||
| 336 | |a Text |b txt |2 rdacontent | ||
| 337 | |a Computermedien |b c |2 rdamedia | ||
| 338 | |a Online-Ressource |b cr |2 rdacarrier | ||
| 500 | |a Gesehen am 29.10.2018 | ||
| 520 | |a Ion beam therapy offers the possibility of a highly conformal tumor-dose distribution; however, this technique is extremely sensitive to inaccuracies in the treatment procedures. Ambiguities in the conversion of Hounsfield units of the treatment planning x-ray CT to relative stopping power (RSP) can cause uncertainties in the estimated ion range of up to several millimeters. Ion CT (iCT) represents a favorable solution allowing to directly assess the RSP. In this simulation study we investigate the performance of the integration-mode configuration for carbon iCT, in comparison with a single-particle approach under the same set-up. The experimental detector consists of a stack of 61 air-filled parallel-plate ionization chambers, interleaved with 3 mm thick PMMA absorbers. By means of Monte Carlo simulations, this design was applied to acquire iCTs of phantoms of tissue-equivalent materials. An optimization of the acquisition parameters was performed to reduce the dose exposure, and the implications of a reduced absorber thickness were assessed. In order to overcome limitations of integration-mode detection in the presence of lateral tissue heterogeneities a dedicated post-processing method using a linear decomposition of the detector signal was developed and its performance was compared to the list-mode acquisition. For the current set-up, the phantom dose could be reduced to below 30 mGy with only minor image quality degradation. By using the decomposition method a correct identification of the components and a RSP accuracy improvement of around 2.0% was obtained. The comparison of integration- and list-mode indicated a slightly better image quality of the latter, with an average median RSP error below 1.8% and 1.0%, respectively. With a decreased absorber thickness a reduced RSP error was observed. Overall, these findings support the potential of iCT for low dose RSP estimation, showing that integration-mode detectors with dedicated post-processing strategies can provide a RSP accuracy comparable to list-mode configurations. | ||
| 700 | 1 | |a Magallanes Herńandez, Lorena |e VerfasserIn |0 (DE-588)1156758181 |0 (DE-627)1019695447 |0 (DE-576)50242382X |4 aut | |
| 700 | 1 | |a Kopp, Benedikt |d 1990- |e VerfasserIn |0 (DE-588)1170097081 |0 (DE-627)1037079787 |0 (DE-576)512373809 |4 aut | |
| 700 | 1 | |a Tessonnier, Thomas |e VerfasserIn |0 (DE-588)1075079195 |0 (DE-627)83292685X |0 (DE-576)443252270 |4 aut | |
| 700 | 1 | |a Parodi, Katia |d 1975- |e VerfasserIn |0 (DE-588)129736996 |0 (DE-627)707308917 |0 (DE-576)188641688 |4 aut | |
| 773 | 0 | 8 | |i Enthalten in |t Physics in medicine and biology |d Bristol : IOP Publ., 1956 |g 62(2017), 3, Seite 1096-1112 |h Online-Ressource |w (DE-627)269016163 |w (DE-600)1473501-5 |w (DE-576)088704130 |x 1361-6560 |7 nnas |a Comparative Monte Carlo study on the performance of integration- and list-mode detector configurations for carbon ion computed tomography |
| 773 | 1 | 8 | |g volume:62 |g year:2017 |g number:3 |g pages:1096-1112 |g extent:18 |a Comparative Monte Carlo study on the performance of integration- and list-mode detector configurations for carbon ion computed tomography |
| 856 | 4 | 0 | |u http://dx.doi.org/10.1088/1361-6560/aa5602 |x Verlag |x Resolving-System |3 Volltext |
| 856 | 4 | 0 | |u http://stacks.iop.org/0031-9155/62/i=3/a=1096 |x Verlag |3 Volltext |
| 951 | |a AR | ||
| 992 | |a 20181029 | ||
| 993 | |a Article | ||
| 994 | |a 2017 | ||
| 998 | |g 129736996 |a Parodi, Katia |m 129736996:Parodi, Katia |p 9 |y j | ||
| 998 | |g 1075079195 |a Tessonnier, Thomas |m 1075079195:Tessonnier, Thomas |p 5 | ||
| 998 | |g 1170097081 |a Kopp, Benedikt |m 1170097081:Kopp, Benedikt |d 910000 |e 910000PK1170097081 |k 0/910000/ |p 4 | ||
| 998 | |g 1156758181 |a Magallanes Hernández, Lorena |m 1156758181:Magallanes Hernández, Lorena |d 910000 |d 911400 |e 910000PM1156758181 |e 911400PM1156758181 |k 0/910000/ |k 1/910000/911400/ |p 3 | ||
| 999 | |a KXP-PPN1582373981 |e 3029955907 | ||
| BIB | |a Y | ||
| SER | |a journal | ||
| JSO | |a {"origin":[{"dateIssuedDisp":"16 January 2017","dateIssuedKey":"2017"}],"relHost":[{"part":{"volume":"62","pages":"1096-1112","year":"2017","extent":"18","issue":"3","text":"62(2017), 3, Seite 1096-1112"},"note":["Gesehen am 28.01.2019","Fortsetzung der Druck-Ausgabe"],"title":[{"title":"Physics in medicine and biology","subtitle":"an official journal of the Institute of Physics and Engineering in Medicine","title_sort":"Physics in medicine and biology"}],"type":{"media":"Online-Ressource","bibl":"periodical"},"id":{"eki":["269016163"],"zdb":["1473501-5"],"issn":["1361-6560"]},"language":["eng"],"physDesc":[{"extent":"Online-Ressource"}],"origin":[{"dateIssuedKey":"1956","publisherPlace":"Bristol","publisher":"IOP Publ.","dateIssuedDisp":"1956-"}],"recId":"269016163","pubHistory":["1.1956 -"],"disp":"Comparative Monte Carlo study on the performance of integration- and list-mode detector configurations for carbon ion computed tomographyPhysics in medicine and biology"}],"name":{"displayForm":["Sebastian Meyer, Chiara Gianoli, Lorena Magallanes, Benedikt Kopp, Thomas Tessonnier, Guillaume Landry, George Dedes, Bernd Voss and Katia Parodi"]},"physDesc":[{"extent":"18 S."}],"id":{"doi":["10.1088/1361-6560/aa5602"],"eki":["1582373981"]},"language":["eng"],"recId":"1582373981","person":[{"role":"aut","given":"Sebastian","roleDisplay":"VerfasserIn","display":"Meyer, Sebastian","family":"Meyer"},{"roleDisplay":"VerfasserIn","display":"Magallanes Herńandez, Lorena","family":"Magallanes Herńandez","role":"aut","given":"Lorena"},{"display":"Kopp, Benedikt","family":"Kopp","roleDisplay":"VerfasserIn","role":"aut","given":"Benedikt"},{"family":"Tessonnier","display":"Tessonnier, Thomas","roleDisplay":"VerfasserIn","role":"aut","given":"Thomas"},{"given":"Katia","role":"aut","roleDisplay":"VerfasserIn","family":"Parodi","display":"Parodi, Katia"}],"title":[{"title":"Comparative Monte Carlo study on the performance of integration- and list-mode detector configurations for carbon ion computed tomography","title_sort":"Comparative Monte Carlo study on the performance of integration- and list-mode detector configurations for carbon ion computed tomography"}],"note":["Gesehen am 29.10.2018"],"type":{"bibl":"article-journal","media":"Online-Ressource"}} | ||
| SRT | |a MEYERSEBASCOMPARATIV1620 | ||