MRI safety of a programmable shunt assistant at 3 and 7 Tesla

Objective. Several new shunt technologies have been developed to optimize hydrocephalus treatment within the past few years. Overdrainage, however, still remains an unresolved problem. One new technology which may reduce the frequency of this complication is the use of a programmable shunt assistant...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Mirzayan, Mohammad Javad (VerfasserIn) , Samii, Madjid (VerfasserIn) , Goetz, Friedrich (VerfasserIn) , Krauss, Joachim K. (VerfasserIn)
Weitere Verfasser: Klinge, Petra Margarete (BerichterstatterIn)
Dokumenttyp: Article (Journal)
Sprache:Englisch
Veröffentlicht: 2012
In: British journal of neurosurgery
Year: 2012, Jahrgang: 26, Heft: 3, Pages: 397-400
ISSN:1360-046X
DOI:10.3109/02688697.2011.625060
Online-Zugang:Verlag, Volltext: http://dx.doi.org/10.3109/02688697.2011.625060
Verlag, Volltext: https://doi.org/10.3109/02688697.2011.625060
Volltext
Verfasserangaben:M. Javad Mirzayan, Petra M. Klinge, Madjid Samii, Friedrich Goetz, Joachim K. Krauss
Beschreibung
Zusammenfassung:Objective. Several new shunt technologies have been developed to optimize hydrocephalus treatment within the past few years. Overdrainage, however, still remains an unresolved problem. One new technology which may reduce the frequency of this complication is the use of a programmable shunt assistant (proSA). Inactive in a horizontal position, it impedes CSF flow in a vertical position according to a prescribed pressure level ranging from 0 to 40 cm H2O. Methods. We exposed the proSA valve in an ex vivo protocol to MR systems operating at 3 and 7 Tesla to investigate its MRI safety. Results. Following 3 Tesla exposure, no changes in valve settings were noted. Adjustment to any pressure level was possible thereafter. The mean deflection angle was 23 ± 3°. After exposure to 7 Tesla, however, there were unintended pressure changes, and the mechanism for further adjustment of the valves even disintegrated. Conclusion. According to the results of this study, proSA is safe with heteropolar vertical magnet alignment at 3 Tesla. Following 7 Tesla exposure, the valves lost their functional capability.
Beschreibung: Published online: 20 Feb 2012
Gesehen am 21.11.2018
Beschreibung:Online Resource
ISSN:1360-046X
DOI:10.3109/02688697.2011.625060