Relative Biologic Effectiveness (RBE) of 50 kV X-rays measured in a phantom for intraoperative tumor-bed irradiation

Purpose - Intraoperative radiation therapy (IORT) with low-energy x-rays is used to treat the tumor bed during breast-conserving surgery. The purpose was to determine the relative biologic effectiveness (RBE) of 50-kV x-rays for inactivation of cells irradiated in a tumor-bed phantom. - Methods and...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Liu, Qi (VerfasserIn) , Schneider, Frank (VerfasserIn) , Ma, Lin (VerfasserIn) , Wenz, Frederik (VerfasserIn) , Herskind, Carsten (VerfasserIn)
Dokumenttyp: Article (Journal)
Sprache:Englisch
Veröffentlicht: 15 March 2013
In: International journal of radiation oncology, biology, physics
Year: 2013, Jahrgang: 85, Heft: 4, Pages: 1127-1133
ISSN:1879-355X
DOI:10.1016/j.ijrobp.2012.08.005
Online-Zugang:Verlag, Volltext: http://dx.doi.org/10.1016/j.ijrobp.2012.08.005
Verlag, Volltext: http://www.sciencedirect.com/science/article/pii/S0360301612033913
Volltext
Verfasserangaben:Qi Liu, Frank Schneider, Lin Ma, Frederik Wenz, and Carsten Herskind
Beschreibung
Zusammenfassung:Purpose - Intraoperative radiation therapy (IORT) with low-energy x-rays is used to treat the tumor bed during breast-conserving surgery. The purpose was to determine the relative biologic effectiveness (RBE) of 50-kV x-rays for inactivation of cells irradiated in a tumor-bed phantom. - Methods and Materials - The RBE was determined for clonogenic inactivation of human tumor and normal cells (MCF7, human umbilical vein endothelial cells, normal skin fibroblasts), and hamster V79 cells. The 50-kV x-rays from the Intrabeam machine (Carl Zeiss Surgical) with a spherical 4-cm applicator were used. Cells were irradiated in a water-equivalent phantom at defined distances (8.1-22.9 mm) from the applicator surface. The 50-kV x-rays from a surface therapy machine (Dermopan, Siemens) were included for comparison; 6-MV x-rays were used as reference radiation. - Results - At 8.1-mm depth in the phantom (dose rate 15.1 Gy/h), mean RBE values of 50-kV x-rays from Intrabeam were 1.26 to 1.42 for the 4 cell types at doses yielding surviving fractions in the range of 0.01 to 0.5. Confidence intervals were in the range of 1.2 and 1.5. Similar RBE values were found for 50-kV x-rays from Dermopan for V79 (1.30, CI 1.25-1.36, P=.74) and GS4 (1.42, CI 1.30-1.54, P=.67). No significant dependence of RBE on dose was found for Intrabeam, but RBE decreased at a larger distance (12.7 mm; 9.8 Gy/h). - Conclusions - An increased clinically relevant RBE was found for cell irradiation with Intrabeam at depths in the tumor bed targeted by IORT. The reduced RBE values at larger distances may be related to increased repair of sublethal damage during protracted irradiation or to hardening of the photon beam energy.
Beschreibung:Available online: 14 September 2012
Gesehen am 21.11.2018
Beschreibung:Online Resource
ISSN:1879-355X
DOI:10.1016/j.ijrobp.2012.08.005