Tomosynthesis implementation with adaptive online calibration on clinical C-arm systems

PurposeCone beam computed tomography (CBCT) systems offer physicians crucial 3D and 2D imaging capabilities during interventions. However, certain medical applications only require very specific information from the CBCTs (e.g., determination of the position of high-contrast objects). In diagnostics...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Chung, Khanlian (VerfasserIn) , Schad, Lothar R. (VerfasserIn) , Zöllner, Frank G. (VerfasserIn)
Dokumenttyp: Article (Journal)
Sprache:Englisch
Veröffentlicht: [October 2018]
In: International journal of computer assisted radiology and surgery
Year: 2018, Jahrgang: 13, Heft: 10, Pages: 1481-1495
ISSN:1861-6429
DOI:10.1007/s11548-018-1782-y
Online-Zugang:Verlag, Volltext: https://doi.org/10.1007/s11548-018-1782-y
Volltext
Verfasserangaben:Khanlian Chung, Lothar R. Schad, Frank G. Zöllner

MARC

LEADER 00000caa a2200000 c 4500
001 1665916184
003 DE-627
005 20220816154441.0
007 cr uuu---uuuuu
008 190520s2018 xx |||||o 00| ||eng c
024 7 |a 10.1007/s11548-018-1782-y  |2 doi 
035 |a (DE-627)1665916184 
035 |a (DE-599)KXP1665916184 
035 |a (OCoLC)1341225002 
040 |a DE-627  |b ger  |c DE-627  |e rda 
041 |a eng 
084 |a 33  |2 sdnb 
100 1 |a Chung, Khanlian  |d 1988-  |e VerfasserIn  |0 (DE-588)1178448363  |0 (DE-627)1049381440  |0 (DE-576)517783150  |4 aut 
245 1 0 |a Tomosynthesis implementation with adaptive online calibration on clinical C-arm systems  |c Khanlian Chung, Lothar R. Schad, Frank G. Zöllner 
264 1 |c [October 2018] 
300 |b Illustrationen 
300 |a 15 
336 |a Text  |b txt  |2 rdacontent 
337 |a Computermedien  |b c  |2 rdamedia 
338 |a Online-Ressource  |b cr  |2 rdacarrier 
500 |a Published online: 8 May 2018 
500 |a Gesehen am 20.05.2019 
520 |a PurposeCone beam computed tomography (CBCT) systems offer physicians crucial 3D and 2D imaging capabilities during interventions. However, certain medical applications only require very specific information from the CBCTs (e.g., determination of the position of high-contrast objects). In diagnostics, tomosynthesis techniques can be used in these cases to minimize dose exposure. Therefore, integrating such techniques on CBCT systems could also be beneficial for interventions. In this paper, we investigate the performance of our implementation of circular tomosynthesis on a CBCT device.MethodsThe tomosynthesis scan trajectory is realized with step-and-shoot on a clinical C-arm device. The online calibration algorithm uses conventionally acquired 3D CBCT of the scanned object as prior knowledge to correct the imaging geometries. The online calibration algorithm was compared to an offline calibration to test its performance. A ball bearing phantom was used to evaluate the reconstructions with respect to geometric distortions. The evaluation was done for three different scenarios to test the robustness of our tomosynthesis implementation against object deviations (e.g., pen) and different object positioning.ResultsThe circular tomosynthesis was tested on a ball bearing and an anthropomorphic phantom. The results show that the calibration is robust against isocenter shifts and object deviations in the CBCT. All reconstructions used 100 projections and displayed limited angle artifacts. The accuracy of the positions and shapes of high-contrast objects were, however, determined precisely. (The maximal center position deviation is 0.31 mm.)ConclusionFor medical procedures that primarily determine the precise position of high-contrast objects, circular tomosynthesis could offer an approach to reduce dose exposure. 
650 4 |a Arbitrary scan trajectories 
650 4 |a C-arm systems 
650 4 |a Circular tomosynthesis 
650 4 |a Cone beam computed tomography 
650 4 |a Image-guided intervention 
650 4 |a Online calibration 
700 1 |a Schad, Lothar R.  |d 1956-  |e VerfasserIn  |0 (DE-588)1028817630  |0 (DE-627)731640241  |0 (DE-576)376271221  |4 aut 
700 1 |a Zöllner, Frank G.  |d 1976-  |e VerfasserIn  |0 (DE-588)129580015  |0 (DE-627)473357054  |0 (DE-576)297732587  |4 aut 
773 0 8 |i Enthalten in  |t International journal of computer assisted radiology and surgery  |d Berlin : Springer, 2006  |g 13(2018), 10, Seite 1481-1495  |h Online-Ressource  |w (DE-627)512299250  |w (DE-600)2235881-X  |w (DE-576)283119322  |x 1861-6429  |7 nnas  |a Tomosynthesis implementation with adaptive online calibration on clinical C-arm systems 
773 1 8 |g volume:13  |g year:2018  |g number:10  |g pages:1481-1495  |g extent:15  |a Tomosynthesis implementation with adaptive online calibration on clinical C-arm systems 
856 4 0 |u https://doi.org/10.1007/s11548-018-1782-y  |x Verlag  |x Resolving-System  |3 Volltext 
856 4 0 |u https://doi.org/10.1007/s11548-018-1782-y  |x Verlag  |3 Volltext 
951 |a AR 
992 |a 20190520 
993 |a Article 
994 |a 2018 
998 |g 129580015  |a Zöllner, Frank G.  |m 129580015:Zöllner, Frank G.  |p 3  |y j 
998 |g 1028817630  |a Schad, Lothar R.  |m 1028817630:Schad, Lothar R.  |p 2 
998 |g 1178448363  |a Chung, Khanlian  |m 1178448363:Chung, Khanlian  |p 1  |x j 
999 |a KXP-PPN1665916184  |e 347822098X 
BIB |a Y 
SER |a journal 
JSO |a {"language":["eng"],"person":[{"family":"Chung","display":"Chung, Khanlian","role":"aut","given":"Khanlian"},{"given":"Lothar R.","role":"aut","display":"Schad, Lothar R.","family":"Schad"},{"display":"Zöllner, Frank G.","family":"Zöllner","given":"Frank G.","role":"aut"}],"id":{"doi":["10.1007/s11548-018-1782-y"],"eki":["1665916184"]},"relHost":[{"part":{"extent":"15","volume":"13","text":"13(2018), 10, Seite 1481-1495","issue":"10","pages":"1481-1495","year":"2018"},"language":["eng"],"type":{"media":"Online-Ressource","bibl":"periodical"},"disp":"Tomosynthesis implementation with adaptive online calibration on clinical C-arm systemsInternational journal of computer assisted radiology and surgery","pubHistory":["1.2006/07 -"],"note":["Gesehen am 26.02.14","Fortsetzung der Druck-Ausgabe"],"origin":[{"publisherPlace":"Berlin ; Heidelberg [u.a.]","dateIssuedKey":"2006","publisher":"Springer","dateIssuedDisp":"2006-"}],"title":[{"subtitle":"a journal for interdisciplinary research, development and applications of image guided diagnosis and therapy","title_sort":"International journal of computer assisted radiology and surgery","title":"International journal of computer assisted radiology and surgery"}],"id":{"eki":["512299250"],"issn":["1861-6429"],"zdb":["2235881-X"]},"recId":"512299250","physDesc":[{"extent":"Online-Ressource"}]}],"recId":"1665916184","name":{"displayForm":["Khanlian Chung, Lothar R. Schad, Frank G. Zöllner"]},"physDesc":[{"noteIll":"Illustrationen","extent":"15 S."}],"type":{"media":"Online-Ressource","bibl":"article-journal"},"note":["Published online: 8 May 2018","Gesehen am 20.05.2019"],"origin":[{"dateIssuedDisp":"[October 2018]","dateIssuedKey":"2018"}],"title":[{"title":"Tomosynthesis implementation with adaptive online calibration on clinical C-arm systems","title_sort":"Tomosynthesis implementation with adaptive online calibration on clinical C-arm systems"}]} 
SRT |a CHUNGKHANLTOMOSYNTHE2018