Test bed for assessment of the kinematical determination of navigation systems for total knee arthroplasty. Does a limited range of motion of the hip joint influence the accuracy of the determination?

Objective: To locate the rotational center of the hip joint, CT-less navigation systems for artificial knee-joint replacement use movements of the femur with a rigid body attached. It cannot be assumed that the hip joint provides free mobility at all times. The purpose of the present study was: 1) T...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Schwarz, Markus (VerfasserIn) , Pott, Peter Paul (VerfasserIn) , Heute, Steffen (VerfasserIn) , Scharf, Hanns-Peter (VerfasserIn)
Dokumenttyp: Article (Journal)
Sprache:Englisch
Veröffentlicht: 06 Jan 2010
In: Computer aided surgery
Year: 2005, Jahrgang: 10, Heft: 1, Pages: 51-57
ISSN:1097-0150
DOI:10.3109/10929080500228597
Online-Zugang:Verlag, Volltext: https://doi.org/10.3109/10929080500228597
Volltext
Verfasserangaben:M.L.R. Schwarz, P.P. Pott, A. Beck, S. Heute, & H.-P. Scharf

MARC

LEADER 00000caa a2200000 c 4500
001 1685474292
003 DE-627
005 20220817184658.0
007 cr uuu---uuuuu
008 191213r20102005xx |||||o 00| ||eng c
024 7 |a 10.3109/10929080500228597  |2 doi 
035 |a (DE-627)1685474292 
035 |a (DE-599)KXP1685474292 
035 |a (OCoLC)1341280725 
040 |a DE-627  |b ger  |c DE-627  |e rda 
041 |a eng 
084 |a 33  |2 sdnb 
100 1 |a Schwarz, Markus  |e VerfasserIn  |0 (DE-588)1058900536  |0 (DE-627)797731733  |0 (DE-576)414981456  |4 aut 
245 1 0 |a Test bed for assessment of the kinematical determination of navigation systems for total knee arthroplasty. Does a limited range of motion of the hip joint influence the accuracy of the determination?  |c M.L.R. Schwarz, P.P. Pott, A. Beck, S. Heute, & H.-P. Scharf 
264 1 |c 06 Jan 2010 
300 |a 7 
336 |a Text  |b txt  |2 rdacontent 
337 |a Computermedien  |b c  |2 rdamedia 
338 |a Online-Ressource  |b cr  |2 rdacarrier 
500 |a Elektronische Reproduktion der Druckausgabe 
500 |a Gesehen am 13.12.2019 
520 |a Objective: To locate the rotational center of the hip joint, CT-less navigation systems for artificial knee-joint replacement use movements of the femur with a rigid body attached. It cannot be assumed that the hip joint provides free mobility at all times. The purpose of the present study was: 1) To build a mechanical model to assess the system's accuracy in locating the rotational center of the hip by simulating a step-wise reduction of the range of motion (ROM) of the hip joint. 2) To determine the system's resolution by assessing a critical distance between two positions of the same femoral rigid body during the process of locating the rotational center of the hip. 3) To determine the sensitivity of the navigation system to the rotation of a femoral rigid body relative to the femoral bone while locating the rotational center of the hip joint.Material and Methods: To assess the impact that a limited ROM of the hip joint has on the accuracy of determination of the hip joint's rotational center, a test bed was built. This enables validation of the algorithm used by a CT-less navigation system.Results: In the first part of the study, it was shown that a reduction of the ROM of the hip joint to 30% of its initial value had no evident influence on the accuracy of locating the rotational center of the joint. In the second part of the study, it was determined that the limit of resolution between two spatial points of the pivoting process is between 4.4 and 8.7 cm. The third part of the study showed that the examined system rejected the determination of the hip center even when the rigid body was only rotated through 22.5°.Conclusions: The results show that osteoarthritis of the hip with a limited ROM, for example, cannot be taken as a contraindication for the use of the evaluated CT-less navigation system. However, the surgeon should ensure that the pivoting of the femur is performed without hindrance within the free range of motion of the hip joint. In accordance with the vendor's recommendation, a minimum distance of 10 cm should be maintained between two spatial points. To ensure safe and unconstrained operation, the rigid body must be firmly attached to the bone and must not be dislocated. 
534 |c 2005 
650 4 |a accuracy 
650 4 |a CT-less navigation 
650 4 |a limited range of motion 
650 4 |a test bed 
700 1 |a Pott, Peter Paul  |e VerfasserIn  |0 (DE-588)13771727X  |0 (DE-627)594857686  |0 (DE-576)303964480  |4 aut 
700 1 |a Heute, Steffen  |e VerfasserIn  |0 (DE-588)1160210403  |0 (DE-627)1023340194  |0 (DE-576)505557983  |4 aut 
700 1 |a Scharf, Hanns-Peter  |d 1953-  |e VerfasserIn  |0 (DE-588)109675266  |0 (DE-627)587792329  |0 (DE-576)168550776  |4 aut 
773 0 8 |i Enthalten in  |t Computer aided surgery  |d Abingdon : Taylor & Francis Group, 1997  |g 10(2005), 1, Seite 51-57  |h Online-Ressource  |w (DE-627)320455807  |w (DE-600)2006641-7  |w (DE-576)090890418  |x 1097-0150  |7 nnas  |a Test bed for assessment of the kinematical determination of navigation systems for total knee arthroplasty. Does a limited range of motion of the hip joint influence the accuracy of the determination? 
773 1 8 |g volume:10  |g year:2005  |g number:1  |g pages:51-57  |g extent:7  |a Test bed for assessment of the kinematical determination of navigation systems for total knee arthroplasty. Does a limited range of motion of the hip joint influence the accuracy of the determination? 
856 4 0 |u https://doi.org/10.3109/10929080500228597  |x Verlag  |x Resolving-System  |3 Volltext 
951 |a AR 
992 |a 20191213 
993 |a Article 
994 |a 2005 
998 |g 109675266  |a Scharf, Hanns-Peter  |m 109675266:Scharf, Hanns-Peter  |d 60000  |d 62400  |e 60000PS109675266  |e 62400PS109675266  |k 0/60000/  |k 1/60000/62400/  |p 5  |y j 
998 |g 1160210403  |a Heute, Steffen  |m 1160210403:Heute, Steffen  |d 60000  |d 62400  |e 60000PH1160210403  |e 62400PH1160210403  |k 0/60000/  |k 1/60000/62400/  |p 4 
998 |g 13771727X  |a Pott, Peter Paul  |m 13771727X:Pott, Peter Paul  |p 2 
998 |g 1058900536  |a Schwarz, Markus  |m 1058900536:Schwarz, Markus  |d 60000  |d 62400  |e 60000PS1058900536  |e 62400PS1058900536  |k 0/60000/  |k 1/60000/62400/  |p 1  |x j 
999 |a KXP-PPN1685474292  |e 3564630333 
BIB |a Y 
SER |a journal 
JSO |a {"relHost":[{"physDesc":[{"extent":"Online-Ressource"}],"origin":[{"publisherPlace":"Abingdon ; London [u.a.] ; New York, NY","publisher":"Taylor & Francis Group ; Informa Healthcare ; Wiley","dateIssuedKey":"1997","dateIssuedDisp":"1997-2015"}],"id":{"zdb":["2006641-7"],"doi":["10.1002/(ISSN)1097-0150"],"eki":["320455807"],"issn":["1097-0150"]},"note":["Gesehen am 04.09.15"],"type":{"media":"Online-Ressource","bibl":"periodical"},"disp":"Test bed for assessment of the kinematical determination of navigation systems for total knee arthroplasty. Does a limited range of motion of the hip joint influence the accuracy of the determination?Computer aided surgery","language":["eng"],"recId":"320455807","pubHistory":["2.1997-volume 20, issue 1 (2015)"],"part":{"issue":"1","pages":"51-57","year":"2005","extent":"7","volume":"10","text":"10(2005), 1, Seite 51-57"},"title":[{"title_sort":"Computer aided surgery","subtitle":"official journal of the International Society for Computer Aided Surgery (ISCAS)","title":"Computer aided surgery"}]}],"physDesc":[{"extent":"7 S."}],"name":{"displayForm":["M.L.R. Schwarz, P.P. Pott, A. Beck, S. Heute, & H.-P. Scharf"]},"id":{"doi":["10.3109/10929080500228597"],"eki":["1685474292"]},"origin":[{"dateIssuedDisp":"06 Jan 2010","dateIssuedKey":"2010"}],"recId":"1685474292","language":["eng"],"note":["Elektronische Reproduktion der Druckausgabe","Gesehen am 13.12.2019"],"type":{"media":"Online-Ressource","bibl":"article-journal"},"person":[{"given":"Markus","family":"Schwarz","role":"aut","display":"Schwarz, Markus","roleDisplay":"VerfasserIn"},{"role":"aut","roleDisplay":"VerfasserIn","display":"Pott, Peter Paul","given":"Peter Paul","family":"Pott"},{"family":"Heute","given":"Steffen","roleDisplay":"VerfasserIn","display":"Heute, Steffen","role":"aut"},{"given":"Hanns-Peter","family":"Scharf","role":"aut","display":"Scharf, Hanns-Peter","roleDisplay":"VerfasserIn"}],"title":[{"title":"Test bed for assessment of the kinematical determination of navigation systems for total knee arthroplasty. Does a limited range of motion of the hip joint influence the accuracy of the determination?","title_sort":"Test bed for assessment of the kinematical determination of navigation systems for total knee arthroplasty. Does a limited range of motion of the hip joint influence the accuracy of the determination?"}]} 
SRT |a SCHWARZMARTESTBEDFOR0620