Evaluation of machine log files/MC-based treatment planning and delivery QA as compared to ArcCHECK QA
Purpose: A treatment planning/delivery QA tool using linac log files (LF) and Monte Carlo (MC) dose calculation is investigated as a standalone alternative to phantom-based patient-specific QA (ArcCHECK (AC)). Methods Delivering a variety of fields onto MapCHECK2 and ArcCHECK, diode sensitivity depe...
Gespeichert in:
| Hauptverfasser: | , , , , , , , |
|---|---|
| Dokumenttyp: | Article (Journal) |
| Sprache: | Englisch |
| Veröffentlicht: |
20 April 2018
|
| In: |
Medical physics
Year: 2018, Jahrgang: 45, Heft: 7, Pages: 2864-2874 |
| ISSN: | 2473-4209 |
| DOI: | 10.1002/mp.12926 |
| Online-Zugang: | Verlag, lizenzpflichtig, Volltext: https://doi.org/10.1002/mp.12926 Verlag, lizenzpflichtig, Volltext: https://aapm.onlinelibrary.wiley.com/doi/abs/10.1002/mp.12926 |
| Verfasserangaben: | Carl W. Stanhope, Douglas G. Drake, Jian Liang, Markus Alber, Matthias Söhn, Charbel Habib, Virgil Willcut, Di Yan |
MARC
| LEADER | 00000caa a2200000 c 4500 | ||
|---|---|---|---|
| 001 | 1694254178 | ||
| 003 | DE-627 | ||
| 005 | 20230427094944.0 | ||
| 007 | cr uuu---uuuuu | ||
| 008 | 200408s2018 xx |||||o 00| ||eng c | ||
| 024 | 7 | |a 10.1002/mp.12926 |2 doi | |
| 035 | |a (DE-627)1694254178 | ||
| 035 | |a (DE-599)KXP1694254178 | ||
| 035 | |a (OCoLC)1341314251 | ||
| 040 | |a DE-627 |b ger |c DE-627 |e rda | ||
| 041 | |a eng | ||
| 084 | |a 33 |2 sdnb | ||
| 100 | 1 | |a Stanhope, Carl |e VerfasserIn |0 (DE-588)1207935522 |0 (DE-627)1694255549 |4 aut | |
| 245 | 1 | 0 | |a Evaluation of machine log files/MC-based treatment planning and delivery QA as compared to ArcCHECK QA |c Carl W. Stanhope, Douglas G. Drake, Jian Liang, Markus Alber, Matthias Söhn, Charbel Habib, Virgil Willcut, Di Yan |
| 264 | 1 | |c 20 April 2018 | |
| 300 | |a 11 | ||
| 336 | |a Text |b txt |2 rdacontent | ||
| 337 | |a Computermedien |b c |2 rdamedia | ||
| 338 | |a Online-Ressource |b cr |2 rdacarrier | ||
| 500 | |a Gesehen am 08.04.2020 | ||
| 520 | |a Purpose: A treatment planning/delivery QA tool using linac log files (LF) and Monte Carlo (MC) dose calculation is investigated as a standalone alternative to phantom-based patient-specific QA (ArcCHECK (AC)). Methods Delivering a variety of fields onto MapCHECK2 and ArcCHECK, diode sensitivity dependence on dose rate (in-field) and energy (primarily out-of-field) was quantified. AC and LF QAs were analyzed with respect to delivery complexity by delivering 12 × 12 cm static fields/arcs comprised of varying numbers of abutting sub-fields onto ArcCHECK. About 11 clinical dual-arc VMAT patients planned using Pinnacle's convolution-superposition (CS) were delivered on ArcCHECK and log file dose (LF-CS and LF-MC) calculated. To minimize calculation time, reduced LF-CS sampling (1/2/3/4° control point spacing) was investigated. Planned (“Plan”) and LF-reconstructed CS and MC doses were compared with each other and AC measurement via statistical [mean ± StdDev(σ)] and gamma analyses to isolate dosimetric uncertainties and quantify the relative accuracies of AC QA and MC-based LF QA. Results Calculation and ArcCHECK measurement differed by up to 1.5% in-field due to variation in dose rate and up to 5% out-of-field. For the experimental segment-varying plans, despite CS calculation deviating by as much as 13% from measurement, Plan-MC and LF-MC doses generally matched AC measurement within 3%. Utilizing 1° control point spacing, 2%/2 mm LF-CS vs AC pass rates (97%) were slightly lower than Plan-CS vs AC pass rates (97.5%). Utilizing all log file samples, 2%/2 mm LF-MC vs AC pass rates (97.3%) were higher than Plan-MC vs AC (96.5%). Phantom-dependent, calculation algorithm-dependent (MC vs CS), and delivery error-dependent dose uncertainties were 0.8 ± 1.2%, 0.2 ± 1.1%, and 0.1 ± 0.9% respectively. Conclusion Reconstructing every log file sample with no increase in computational cost, MC-based LF QA is faster and more accurate than CS-based LF QA. Offering similar dosimetric accuracy compared to AC measurement, MC-based log files can be used for treatment planning QA. | ||
| 650 | 4 | |a ArcCHECK | |
| 650 | 4 | |a log file | |
| 650 | 4 | |a Monte Carlo | |
| 650 | 4 | |a quality assurance | |
| 700 | 1 | |a Drake, Douglas G. |e VerfasserIn |4 aut | |
| 700 | 1 | |a Liang, Jian |e VerfasserIn |4 aut | |
| 700 | 1 | |a Alber, Markus |e VerfasserIn |0 (DE-588)1145057101 |0 (DE-627)1005461767 |0 (DE-576)495602507 |4 aut | |
| 700 | 1 | |a Söhn, Matthias |e VerfasserIn |4 aut | |
| 700 | 1 | |a Habib, Charbel |e VerfasserIn |4 aut | |
| 700 | 1 | |a Willcut, Virgil |e VerfasserIn |4 aut | |
| 700 | 1 | |a Yan, Di |e VerfasserIn |4 aut | |
| 773 | 0 | 8 | |i Enthalten in |t Medical physics |d Hoboken, NJ : Wiley, 1974 |g 45(2018), 7, Seite 2864-2874 |h Online-Ressource |w (DE-627)265784867 |w (DE-600)1466421-5 |w (DE-576)074891243 |x 2473-4209 |7 nnas |a Evaluation of machine log files/MC-based treatment planning and delivery QA as compared to ArcCHECK QA |
| 773 | 1 | 8 | |g volume:45 |g year:2018 |g number:7 |g pages:2864-2874 |g extent:11 |a Evaluation of machine log files/MC-based treatment planning and delivery QA as compared to ArcCHECK QA |
| 856 | 4 | 0 | |u https://doi.org/10.1002/mp.12926 |x Verlag |x Resolving-System |z lizenzpflichtig |3 Volltext |
| 856 | 4 | 0 | |u https://aapm.onlinelibrary.wiley.com/doi/abs/10.1002/mp.12926 |x Verlag |z lizenzpflichtig |3 Volltext |
| 951 | |a AR | ||
| 992 | |a 20200408 | ||
| 993 | |a Article | ||
| 994 | |a 2018 | ||
| 998 | |g 1145057101 |a Alber, Markus |m 1145057101:Alber, Markus |d 910000 |d 911400 |e 910000PA1145057101 |e 911400PA1145057101 |k 0/910000/ |k 1/910000/911400/ |p 4 | ||
| 999 | |a KXP-PPN1694254178 |e 3620463972 | ||
| BIB | |a Y | ||
| SER | |a journal | ||
| JSO | |a {"person":[{"given":"Carl","role":"aut","display":"Stanhope, Carl","family":"Stanhope"},{"given":"Douglas G.","role":"aut","display":"Drake, Douglas G.","family":"Drake"},{"display":"Liang, Jian","family":"Liang","role":"aut","given":"Jian"},{"family":"Alber","display":"Alber, Markus","given":"Markus","role":"aut"},{"given":"Matthias","role":"aut","display":"Söhn, Matthias","family":"Söhn"},{"given":"Charbel","role":"aut","family":"Habib","display":"Habib, Charbel"},{"family":"Willcut","display":"Willcut, Virgil","role":"aut","given":"Virgil"},{"display":"Yan, Di","family":"Yan","role":"aut","given":"Di"}],"language":["eng"],"origin":[{"dateIssuedKey":"2018","dateIssuedDisp":"20 April 2018"}],"note":["Gesehen am 08.04.2020"],"type":{"bibl":"article-journal","media":"Online-Ressource"},"title":[{"title_sort":"Evaluation of machine log files/MC-based treatment planning and delivery QA as compared to ArcCHECK QA","title":"Evaluation of machine log files/MC-based treatment planning and delivery QA as compared to ArcCHECK QA"}],"recId":"1694254178","name":{"displayForm":["Carl W. Stanhope, Douglas G. Drake, Jian Liang, Markus Alber, Matthias Söhn, Charbel Habib, Virgil Willcut, Di Yan"]},"physDesc":[{"extent":"11 S."}],"relHost":[{"recId":"265784867","pubHistory":["1.1974 -"],"physDesc":[{"extent":"Online-Ressource"}],"name":{"displayForm":["American Association of Physicists in Medicine ; American Institute of Physics"]},"id":{"eki":["265784867"],"issn":["2473-4209","1522-8541"],"zdb":["1466421-5"]},"disp":"Evaluation of machine log files/MC-based treatment planning and delivery QA as compared to ArcCHECK QAMedical physics","language":["eng"],"part":{"extent":"11","volume":"45","year":"2018","pages":"2864-2874","text":"45(2018), 7, Seite 2864-2874","issue":"7"},"note":["Gesehen am 01.08.2025"],"title":[{"title_sort":"Medical physics","title":"Medical physics"}],"origin":[{"publisher":"Wiley ; AAPM ; [Verlag nicht ermittelbar]","dateIssuedKey":"1974","publisherPlace":"Hoboken, NJ ; College Park, Md. ; New York, NY","dateIssuedDisp":"1974-"}],"titleAlt":[{"title":"Medical physics online"}],"type":{"media":"Online-Ressource","bibl":"periodical"}}],"id":{"doi":["10.1002/mp.12926"],"eki":["1694254178"]}} | ||
| SRT | |a STANHOPECAEVALUATION2020 | ||