Electrocautery damage can reduce implant fatigue strength: cases and in vitro investigation

Background: The risk of femoral stem fracture after total hip replacement is low and can often be associated with a specific implant system or other factors that may reduce the fatigue strength. Additionally, damage to a metal component during revision surgery by an electrocautery device may further...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Sonntag, Robert (VerfasserIn) , Gibmeier, Jens (VerfasserIn) , Pulvermacher, Samuel (VerfasserIn) , Müller, Ulrike (VerfasserIn) , Eckert, Johannes A. (VerfasserIn) , Braun, Steffen (VerfasserIn) , Reichkendler, Markus (VerfasserIn) , Kretzer, Jan Philippe (VerfasserIn)
Dokumenttyp: Article (Journal)
Sprache:Englisch
Veröffentlicht: May 15, 2019
In: The journal of bone & joint surgery. A, American volume
Year: 2019, Jahrgang: 101, Heft: 10, Pages: 868-878
ISSN:1535-1386
DOI:10.2106/JBJS.18.00259
Online-Zugang:Verlag, lizenzpflichtig, Volltext: https://doi.org/10.2106/JBJS.18.00259
Verlag, lizenzpflichtig, Volltext: https://journals.lww.com/jbjsjournal/Fulltext/2019/05150/Electrocautery_Damage_Can_Reduce_Implant_Fatigue.2.aspx
Volltext
Verfasserangaben:Robert Sonntag, Jens Gibmeier, Samuel Pulvermacher, Ulrike Mueller, Johannes Eckert, Steffen Braun, Markus Reichkendler, and J. Philippe Kretzer

MARC

LEADER 00000caa a2200000 c 4500
001 169480383X
003 DE-627
005 20240405193217.0
007 cr uuu---uuuuu
008 200416s2019 xx |||||o 00| ||eng c
024 7 |a 10.2106/JBJS.18.00259  |2 doi 
035 |a (DE-627)169480383X 
035 |a (DE-599)KXP169480383X 
035 |a (OCoLC)1341315193 
040 |a DE-627  |b ger  |c DE-627  |e rda 
041 |a eng 
084 |a 33  |2 sdnb 
100 1 |a Sonntag, Robert  |e VerfasserIn  |0 (DE-588)1068843756  |0 (DE-627)82081699X  |0 (DE-576)428211135  |4 aut 
245 1 0 |a Electrocautery damage can reduce implant fatigue strength  |b cases and in vitro investigation  |c Robert Sonntag, Jens Gibmeier, Samuel Pulvermacher, Ulrike Mueller, Johannes Eckert, Steffen Braun, Markus Reichkendler, and J. Philippe Kretzer 
264 1 |c May 15, 2019 
300 |a 11 
336 |a Text  |b txt  |2 rdacontent 
337 |a Computermedien  |b c  |2 rdamedia 
338 |a Online-Ressource  |b cr  |2 rdacarrier 
500 |a Gesehen am 16.04.2020 
520 |a Background: The risk of femoral stem fracture after total hip replacement is low and can often be associated with a specific implant system or other factors that may reduce the fatigue strength. Additionally, damage to a metal component during revision surgery by an electrocautery device may further affect the fatigue behavior. - Methods: Two clinical cases of stem failure after revision of fractured ceramic components are presented; the retrieved components were analyzed for the cause of failure. In vitro cyclic load-to-failure testing of titanium alloy femoral stems after electrocautery application at 2 different locations (at the base and about midway on the femoral neck) was performed using a stepwise increase in load until implant fracture occurred. In addition, a detailed characterization of the local material structure around the electrocautery marks was performed. Results: Superficial discoloration and melting marks were found on the retrieved components, including at the location of crack initiation in the anterolateral region, which may have reduced the fatigue strength of the material. In addition, elemental analysis indicated material transfer from the electrocautery tip. Damage to the surface by the electrocautery device significantly reduced the in vitro load to failure by up to 47% compared with that of undamaged femoral neck specimens. Material analysis revealed a relevant modification in microstructure, with an extension of approximately 2.7 mm and a depth of 550 µm, which could be divided in 3 structural zones. Conclusions: Intraoperative electrocautery device contact with the implant during surgical revision of a total hip replacement cannot always be avoided. However, on the basis of our findings, the risk of implant failure is increased due to a change in microstructure and a potential reduction of the implant’s fatigue strength. Surgeons and manufacturers of electrocautery devices should be aware of this concern. Clinical Relevance: During revision surgery, contact between an electrocautery device and the femoral component should be avoided to reduce the chance of subsequent femoral neck fracture. 
700 1 |a Gibmeier, Jens  |e VerfasserIn  |4 aut 
700 1 |a Pulvermacher, Samuel  |e VerfasserIn  |4 aut 
700 1 |a Müller, Ulrike  |d 1988-  |e VerfasserIn  |0 (DE-588)1116644002  |0 (DE-627)870607081  |0 (DE-576)478555504  |4 aut 
700 1 |a Eckert, Johannes A.  |d 1985-  |e VerfasserIn  |0 (DE-588)1116642697  |0 (DE-627)870604651  |0 (DE-576)478553854  |4 aut 
700 1 |a Braun, Steffen  |d 1988-  |e VerfasserIn  |0 (DE-588)1149869429  |0 (DE-627)1010183974  |0 (DE-576)496872729  |4 aut 
700 1 |a Reichkendler, Markus  |e VerfasserIn  |4 aut 
700 1 |a Kretzer, Jan Philippe  |d 1978-  |e VerfasserIn  |0 (DE-588)136998348  |0 (DE-627)589003690  |0 (DE-576)301383103  |4 aut 
773 0 8 |i Enthalten in  |t The journal of bone & joint surgery. A, American volume  |d Boston, Mass. : Journal, 1889  |g 101(2019), 10, Seite 868-878  |h Online-Ressource  |w (DE-627)268761485  |w (DE-600)1473137-X  |w (DE-576)077610555  |x 1535-1386  |7 nnas  |a Electrocautery damage can reduce implant fatigue strength cases and in vitro investigation 
773 1 8 |g volume:101  |g year:2019  |g number:10  |g pages:868-878  |g extent:11  |a Electrocautery damage can reduce implant fatigue strength cases and in vitro investigation 
856 4 0 |u https://doi.org/10.2106/JBJS.18.00259  |x Verlag  |x Resolving-System  |z lizenzpflichtig  |3 Volltext 
856 4 0 |u https://journals.lww.com/jbjsjournal/Fulltext/2019/05150/Electrocautery_Damage_Can_Reduce_Implant_Fatigue.2.aspx  |x Verlag  |z lizenzpflichtig  |3 Volltext 
951 |a AR 
992 |a 20200416 
993 |a Article 
994 |a 2019 
998 |g 136998348  |a Kretzer, Jan Philippe  |m 136998348:Kretzer, Jan Philippe  |d 910000  |d 911460  |d 50000  |e 910000PK136998348  |e 911460PK136998348  |e 50000PK136998348  |k 0/910000/  |k 1/910000/911460/  |k 0/50000/  |p 8  |y j 
998 |g 1149869429  |a Braun, Steffen  |m 1149869429:Braun, Steffen  |d 910000  |d 911460  |e 910000PB1149869429  |e 911460PB1149869429  |k 0/910000/  |k 1/910000/911460/  |p 6 
998 |g 1116642697  |a Eckert, Johannes A.  |m 1116642697:Eckert, Johannes A.  |d 910000  |d 911460  |e 910000PE1116642697  |e 911460PE1116642697  |k 0/910000/  |k 1/910000/911460/  |p 5 
998 |g 1116644002  |a Müller, Ulrike  |m 1116644002:Müller, Ulrike  |d 910000  |d 911460  |e 910000PM1116644002  |e 911460PM1116644002  |k 0/910000/  |k 1/910000/911460/  |p 4 
998 |g 1068843756  |a Sonntag, Robert  |m 1068843756:Sonntag, Robert  |d 910000  |d 911460  |e 910000PS1068843756  |e 911460PS1068843756  |k 0/910000/  |k 1/910000/911460/  |p 1  |x j 
999 |a KXP-PPN169480383X  |e 3624520938 
BIB |a Y 
SER |a journal 
JSO |a {"type":{"media":"Online-Ressource","bibl":"article-journal"},"id":{"eki":["169480383X"],"doi":["10.2106/JBJS.18.00259"]},"physDesc":[{"extent":"11 S."}],"relHost":[{"note":["Gesehen am 22.08.2023"],"physDesc":[{"extent":"Online-Ressource"}],"id":{"eki":["268761485"],"issn":["1535-1386"],"zdb":["1473137-X"]},"part":{"extent":"11","issue":"10","volume":"101","text":"101(2019), 10, Seite 868-878","year":"2019","pages":"868-878"},"title":[{"partname":"American volume","title_sort":"journal of bone & joint surgery","subtitle":"JBJS","title":"The journal of bone & joint surgery"}],"origin":[{"publisher":"Journal","dateIssuedDisp":"1889-","dateIssuedKey":"1889","publisherPlace":"Boston, Mass."}],"name":{"displayForm":["American Orthopaedic Association ; British Orthopaedic Association ; American Academy of Orthopaedic Surgeons"]},"corporate":[{"display":"American Orthopaedic Association","role":"isb"},{"display":"British Orthopaedic Association","role":"isb"}],"pubHistory":["1.1889 -"],"type":{"bibl":"periodical","media":"Online-Ressource"},"language":["eng"],"titleAlt":[{"title":"The journal of bone & joint surgery / A"},{"title":"JB&JS"},{"title":"The journal of bone and joint surgery"},{"title":"JBJS"},{"title":"The journal of bone and joint surgery"}],"recId":"268761485","disp":"Electrocautery damage can reduce implant fatigue strength cases and in vitro investigationThe journal of bone & joint surgery. A, American volume"}],"note":["Gesehen am 16.04.2020"],"recId":"169480383X","origin":[{"dateIssuedKey":"2019","dateIssuedDisp":"May 15, 2019"}],"name":{"displayForm":["Robert Sonntag, Jens Gibmeier, Samuel Pulvermacher, Ulrike Mueller, Johannes Eckert, Steffen Braun, Markus Reichkendler, and J. Philippe Kretzer"]},"person":[{"display":"Sonntag, Robert","given":"Robert","family":"Sonntag","role":"aut"},{"role":"aut","display":"Gibmeier, Jens","given":"Jens","family":"Gibmeier"},{"role":"aut","family":"Pulvermacher","given":"Samuel","display":"Pulvermacher, Samuel"},{"display":"Müller, Ulrike","given":"Ulrike","family":"Müller","role":"aut"},{"role":"aut","display":"Eckert, Johannes A.","given":"Johannes A.","family":"Eckert"},{"role":"aut","given":"Steffen","display":"Braun, Steffen","family":"Braun"},{"role":"aut","given":"Markus","display":"Reichkendler, Markus","family":"Reichkendler"},{"role":"aut","display":"Kretzer, Jan Philippe","given":"Jan Philippe","family":"Kretzer"}],"language":["eng"],"title":[{"title":"Electrocautery damage can reduce implant fatigue strength","title_sort":"Electrocautery damage can reduce implant fatigue strength","subtitle":"cases and in vitro investigation"}]} 
SRT |a SONNTAGROBELECTROCAU1520