Purpose: We analyzed heart rate variability (HRV) taken by ECG and photoplethysmography (PPG) to assess their agreement. We also analyzed the sensitivity and specificity of PPG to identify subjects with low HRV as an example of its potential use for clinical applications.Methods: The HRV parameters: mean heart rate (HR), amplitude, and ratio of heart rate oscillation (E–I difference, E/I ratio), RMSSD, SDNN, and Power LF, were measured during 1-min deep breathing tests (DBT) in 343 individuals, followed by a 5-min short-term HRV (s-HRV), where the HRV parameters: HR, SD1, SD2, SDNN, Stress Index, Power HF, Power LF, Power VLF, and Total Power, were determined as well. Parameters were compared through correlation analysis and agreement analysis by Bland–Altman plots. Results: PPG derived parameters HR and SD2 in s-HRV showed better agreement than SD1, Power HF, and stress index, whereas in DBT HR, E/I ratio and SDNN were superior to Power LF and RMSSD. DBT yielded stronger agreement than s-HRV. A slight overestimation of PPG HRV over HCG HRV was found. HR, Total Power, and SD2 in the s-HRV, HR, Power LF, and SDNN in the DBT showed high sensitivity and specificity to detect individuals with poor HRV. Cutoff percentiles are given for the future development of PPG-based devices. Conclusion:HRV measured by PPG shows good agreement with ECG HRV when appropriate parameters are used, and PPG-based devices can be employed as an easy screening tool to detect individuals with poor HRV, especially in the 1-min DBT test.

We analyzed heart rate variability (HRV) taken by ECG and photoplethysmography (PPG) to assess their agreement. We also analyzed the sensitivity and specificity of PPG to identify subjects with low HRV as an example of its potential use for clinical applications.

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Weinschenk, Stefan (VerfasserIn) , Beise, Reinhard (VerfasserIn) , Lorenz, Jürgen (VerfasserIn)
Dokumenttyp: Article (Journal)
Sprache:Englisch
Veröffentlicht: 08 June 2016
In: European journal of applied physiology
Year: 2016, Jahrgang: 116, Heft: 8, Pages: 1527-1535
ISSN:1439-6327
DOI:10.1007/s00421-016-3401-3
Online-Zugang:Verlag, lizenzpflichtig, Volltext: https://doi.org/10.1007/s00421-016-3401-3
Volltext
Verfasserangaben:Stefan W. Weinschenk, Reinhard D. Beise & Jürgen Lorenz

MARC

LEADER 00000caa a2200000 c 4500
001 1700531301
003 DE-627
005 20230803145754.0
007 cr uuu---uuuuu
008 200612s2016 xx |||||o 00| ||eng c
024 7 |a 10.1007/s00421-016-3401-3  |2 doi 
035 |a (DE-627)1700531301 
035 |a (DE-599)KXP1700531301 
035 |a (OCoLC)1341339597 
040 |a DE-627  |b ger  |c DE-627  |e rda 
041 |a eng 
084 |a 33  |2 sdnb 
100 1 |a Weinschenk, Stefan  |d 1957-  |e VerfasserIn  |0 (DE-588)111430062  |0 (DE-627)624912507  |0 (DE-576)322312973  |4 aut 
245 1 0 |a Purpose: We analyzed heart rate variability (HRV) taken by ECG and photoplethysmography (PPG) to assess their agreement. We also analyzed the sensitivity and specificity of PPG to identify subjects with low HRV as an example of its potential use for clinical applications.Methods: The HRV parameters: mean heart rate (HR), amplitude, and ratio of heart rate oscillation (E–I difference, E/I ratio), RMSSD, SDNN, and Power LF, were measured during 1-min deep breathing tests (DBT) in 343 individuals, followed by a 5-min short-term HRV (s-HRV), where the HRV parameters: HR, SD1, SD2, SDNN, Stress Index, Power HF, Power LF, Power VLF, and Total Power, were determined as well. Parameters were compared through correlation analysis and agreement analysis by Bland–Altman plots. Results: PPG derived parameters HR and SD2 in s-HRV showed better agreement than SD1, Power HF, and stress index, whereas in DBT HR, E/I ratio and SDNN were superior to Power LF and RMSSD. DBT yielded stronger agreement than s-HRV. A slight overestimation of PPG HRV over HCG HRV was found. HR, Total Power, and SD2 in the s-HRV, HR, Power LF, and SDNN in the DBT showed high sensitivity and specificity to detect individuals with poor HRV. Cutoff percentiles are given for the future development of PPG-based devices. Conclusion:HRV measured by PPG shows good agreement with ECG HRV when appropriate parameters are used, and PPG-based devices can be employed as an easy screening tool to detect individuals with poor HRV, especially in the 1-min DBT test.  |c Stefan W. Weinschenk, Reinhard D. Beise & Jürgen Lorenz 
246 3 0 |a five 
264 1 |c 08 June 2016 
300 |a 9 
336 |a Text  |b txt  |2 rdacontent 
337 |a Computermedien  |b c  |2 rdamedia 
338 |a Online-Ressource  |b cr  |2 rdacarrier 
520 |a We analyzed heart rate variability (HRV) taken by ECG and photoplethysmography (PPG) to assess their agreement. We also analyzed the sensitivity and specificity of PPG to identify subjects with low HRV as an example of its potential use for clinical applications. 
700 1 |a Beise, Reinhard  |d 1967-  |e VerfasserIn  |0 (DE-588)120733099  |0 (DE-627)704954338  |0 (DE-576)292359608  |4 aut 
700 1 |a Lorenz, Jürgen  |d 1963-  |e VerfasserIn  |0 (DE-588)118150030  |0 (DE-627)079272762  |0 (DE-576)291740499  |4 aut 
773 0 8 |i Enthalten in  |t European journal of applied physiology  |d Berlin : Springer, 1928  |g 116(2016), 8, Seite 1527-1535  |h Online-Ressource  |w (DE-627)253722780  |w (DE-600)1459054-2  |w (DE-576)072372605  |x 1439-6327  |7 nnas  |a Purpose: We analyzed heart rate variability (HRV) taken by ECG and photoplethysmography (PPG) to assess their agreement. We also analyzed the sensitivity and specificity of PPG to identify subjects with low HRV as an example of its potential use for clinical applications.Methods: The HRV parameters: mean heart rate (HR), amplitude, and ratio of heart rate oscillation (E–I difference, E/I ratio), RMSSD, SDNN, and Power LF, were measured during 1-min deep breathing tests (DBT) in 343 individuals, followed by a 5-min short-term HRV (s-HRV), where the HRV parameters: HR, SD1, SD2, SDNN, Stress Index, Power HF, Power LF, Power VLF, and Total Power, were determined as well. Parameters were compared through correlation analysis and agreement analysis by Bland–Altman plots. Results: PPG derived parameters HR and SD2 in s-HRV showed better agreement than SD1, Power HF, and stress index, whereas in DBT HR, E/I ratio and SDNN were superior to Power LF and RMSSD. DBT yielded stronger agreement than s-HRV. A slight overestimation of PPG HRV over HCG HRV was found. HR, Total Power, and SD2 in the s-HRV, HR, Power LF, and SDNN in the DBT showed high sensitivity and specificity to detect individuals with poor HRV. Cutoff percentiles are given for the future development of PPG-based devices. Conclusion:HRV measured by PPG shows good agreement with ECG HRV when appropriate parameters are used, and PPG-based devices can be employed as an easy screening tool to detect individuals with poor HRV, especially in the 1-min DBT test. 
773 1 8 |g volume:116  |g year:2016  |g number:8  |g pages:1527-1535  |g extent:9  |a Purpose: We analyzed heart rate variability (HRV) taken by ECG and photoplethysmography (PPG) to assess their agreement. We also analyzed the sensitivity and specificity of PPG to identify subjects with low HRV as an example of its potential use for clinical applications.Methods: The HRV parameters: mean heart rate (HR), amplitude, and ratio of heart rate oscillation (E–I difference, E/I ratio), RMSSD, SDNN, and Power LF, were measured during 1-min deep breathing tests (DBT) in 343 individuals, followed by a 5-min short-term HRV (s-HRV), where the HRV parameters: HR, SD1, SD2, SDNN, Stress Index, Power HF, Power LF, Power VLF, and Total Power, were determined as well. Parameters were compared through correlation analysis and agreement analysis by Bland–Altman plots. Results: PPG derived parameters HR and SD2 in s-HRV showed better agreement than SD1, Power HF, and stress index, whereas in DBT HR, E/I ratio and SDNN were superior to Power LF and RMSSD. DBT yielded stronger agreement than s-HRV. A slight overestimation of PPG HRV over HCG HRV was found. HR, Total Power, and SD2 in the s-HRV, HR, Power LF, and SDNN in the DBT showed high sensitivity and specificity to detect individuals with poor HRV. Cutoff percentiles are given for the future development of PPG-based devices. Conclusion:HRV measured by PPG shows good agreement with ECG HRV when appropriate parameters are used, and PPG-based devices can be employed as an easy screening tool to detect individuals with poor HRV, especially in the 1-min DBT test. 
856 4 0 |u https://doi.org/10.1007/s00421-016-3401-3  |x Verlag  |x Resolving-System  |z lizenzpflichtig  |3 Volltext 
951 |a AR 
992 |a 20200612 
993 |a Article 
994 |a 2016 
998 |g 111430062  |a Weinschenk, Stefan  |m 111430062:Weinschenk, Stefan  |d 50000  |e 50000PW111430062  |k 0/50000/  |p 1  |x j 
999 |a KXP-PPN1700531301  |e 3686701448 
BIB |a Y 
SER |a journal 
JSO |a {"origin":[{"dateIssuedKey":"2016","dateIssuedDisp":"08 June 2016"}],"title":[{"title":"Purpose: We analyzed heart rate variability (HRV) taken by ECG and photoplethysmography (PPG) to assess their agreement. We also analyzed the sensitivity and specificity of PPG to identify subjects with low HRV as an example of its potential use for clinical applications.Methods: The HRV parameters: mean heart rate (HR), amplitude, and ratio of heart rate oscillation (E–I difference, E/I ratio), RMSSD, SDNN, and Power LF, were measured during 1-min deep breathing tests (DBT) in 343 individuals, followed by a 5-min short-term HRV (s-HRV), where the HRV parameters: HR, SD1, SD2, SDNN, Stress Index, Power HF, Power LF, Power VLF, and Total Power, were determined as well. Parameters were compared through correlation analysis and agreement analysis by Bland–Altman plots. Results: PPG derived parameters HR and SD2 in s-HRV showed better agreement than SD1, Power HF, and stress index, whereas in DBT HR, E/I ratio and SDNN were superior to Power LF and RMSSD. DBT yielded stronger agreement than s-HRV. A slight overestimation of PPG HRV over HCG HRV was found. HR, Total Power, and SD2 in the s-HRV, HR, Power LF, and SDNN in the DBT showed high sensitivity and specificity to detect individuals with poor HRV. Cutoff percentiles are given for the future development of PPG-based devices. Conclusion:HRV measured by PPG shows good agreement with ECG HRV when appropriate parameters are used, and PPG-based devices can be employed as an easy screening tool to detect individuals with poor HRV, especially in the 1-min DBT test.","title_sort":"Purpose: We analyzed heart rate variability (HRV) taken by ECG and photoplethysmography (PPG) to assess their agreement. We also analyzed the sensitivity and specificity of PPG to identify subjects with low HRV as an example of its potential use for clinical applications.Methods: The HRV parameters: mean heart rate (HR), amplitude, and ratio of heart rate oscillation (E–I difference, E/I ratio), RMSSD, SDNN, and Power LF, were measured during 1-min deep breathing tests (DBT) in 343 individuals, followed by a 5-min short-term HRV (s-HRV), where the HRV parameters: HR, SD1, SD2, SDNN, Stress Index, Power HF, Power LF, Power VLF, and Total Power, were determined as well. Parameters were compared through correlation analysis and agreement analysis by Bland–Altman plots. Results: PPG derived parameters HR and SD2 in s-HRV showed better agreement than SD1, Power HF, and stress index, whereas in DBT HR, E/I ratio and SDNN were superior to Power LF and RMSSD. DBT yielded stronger agreement than s-HRV. A slight overestimation of PPG HRV over HCG HRV was found. HR, Total Power, and SD2 in the s-HRV, HR, Power LF, and SDNN in the DBT showed high sensitivity and specificity to detect individuals with poor HRV. Cutoff percentiles are given for the future development of PPG-based devices. Conclusion:HRV measured by PPG shows good agreement with ECG HRV when appropriate parameters are used, and PPG-based devices can be employed as an easy screening tool to detect individuals with poor HRV, especially in the 1-min DBT test."}],"id":{"doi":["10.1007/s00421-016-3401-3"],"eki":["1700531301"]},"name":{"displayForm":["Stefan W. Weinschenk, Reinhard D. Beise & Jürgen Lorenz"]},"person":[{"family":"Weinschenk","given":"Stefan","roleDisplay":"VerfasserIn","display":"Weinschenk, Stefan","role":"aut"},{"display":"Beise, Reinhard","roleDisplay":"VerfasserIn","role":"aut","family":"Beise","given":"Reinhard"},{"role":"aut","roleDisplay":"VerfasserIn","display":"Lorenz, Jürgen","given":"Jürgen","family":"Lorenz"}],"physDesc":[{"extent":"9 S."}],"type":{"media":"Online-Ressource","bibl":"article-journal"},"relHost":[{"physDesc":[{"extent":"Online-Ressource"}],"id":{"zdb":["1459054-2"],"eki":["253722780"],"issn":["1439-6327","1432-1025"]},"origin":[{"publisherPlace":"Berlin ; Heidelberg ; Berlin ; Heidelberg ; New York, NY","dateIssuedKey":"1928","publisher":"Springer ; Springer","dateIssuedDisp":"1928-"}],"part":{"extent":"9","text":"116(2016), 8, Seite 1527-1535","volume":"116","pages":"1527-1535","issue":"8","year":"2016"},"titleAlt":[{"title":"Applied physiology"},{"title":"Arbeitsphysiologie"},{"title":"Internationale Zeitschrift für angewandte Physiologie einschließlich Arbeitsphysiologie"},{"title":"European journal of applied physiology and occupational physiology"}],"pubHistory":["1.1928/29 -"],"recId":"253722780","language":["eng"],"type":{"media":"Online-Ressource","bibl":"periodical"},"disp":"Purpose: We analyzed heart rate variability (HRV) taken by ECG and photoplethysmography (PPG) to assess their agreement. We also analyzed the sensitivity and specificity of PPG to identify subjects with low HRV as an example of its potential use for clinical applications.Methods: The HRV parameters: mean heart rate (HR), amplitude, and ratio of heart rate oscillation (E–I difference, E/I ratio), RMSSD, SDNN, and Power LF, were measured during 1-min deep breathing tests (DBT) in 343 individuals, followed by a 5-min short-term HRV (s-HRV), where the HRV parameters: HR, SD1, SD2, SDNN, Stress Index, Power HF, Power LF, Power VLF, and Total Power, were determined as well. Parameters were compared through correlation analysis and agreement analysis by Bland–Altman plots. Results: PPG derived parameters HR and SD2 in s-HRV showed better agreement than SD1, Power HF, and stress index, whereas in DBT HR, E/I ratio and SDNN were superior to Power LF and RMSSD. DBT yielded stronger agreement than s-HRV. A slight overestimation of PPG HRV over HCG HRV was found. HR, Total Power, and SD2 in the s-HRV, HR, Power LF, and SDNN in the DBT showed high sensitivity and specificity to detect individuals with poor HRV. Cutoff percentiles are given for the future development of PPG-based devices. Conclusion:HRV measured by PPG shows good agreement with ECG HRV when appropriate parameters are used, and PPG-based devices can be employed as an easy screening tool to detect individuals with poor HRV, especially in the 1-min DBT test.European journal of applied physiology","note":["Gesehen am 19.08.09"],"title":[{"title":"European journal of applied physiology","title_sort":"European journal of applied physiology"}]}],"language":["eng"],"recId":"1700531301"} 
SRT |a WEINSCHENKPURPOSEWEA0820