How do gait frequency and serum-replacement interval affect polyethylene wear in knee-wear simulator tests?

Polyethylene wear (PE) is known to be a limiting factor in total joint replacements. However, a standardized wear test (e.g. ISO standard) can only replicate the complex in vivo loading condition in a simplified form. In this study, two different parameters were analyzed: (a) Bovine serum, as a subs...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Reinders, Jörn (VerfasserIn) , Sonntag, Robert (VerfasserIn) , Kretzer, Jan Philippe (VerfasserIn)
Dokumenttyp: Article (Journal)
Sprache:Englisch
Veröffentlicht: 12 July 2014
In: Journal of materials science
Year: 2014, Jahrgang: 25, Heft: 11, Pages: 2463-2469
ISSN:1573-4838
DOI:10.1007/s10856-014-5271-8
Online-Zugang:Verlag, lizenzpflichtig, Volltext: https://doi.org/10.1007/s10856-014-5271-8
Volltext
Verfasserangaben:Jörn Reinders, Robert Sonntag, Jan Philippe Kretzer

MARC

LEADER 00000caa a2200000 c 4500
001 1743367635
003 DE-627
005 20230427215720.0
007 cr uuu---uuuuu
008 201221s2014 xx |||||o 00| ||eng c
024 7 |a 10.1007/s10856-014-5271-8  |2 doi 
035 |a (DE-627)1743367635 
035 |a (DE-599)KXP1743367635 
035 |a (OCoLC)1341384194 
040 |a DE-627  |b ger  |c DE-627  |e rda 
041 |a eng 
084 |a 33  |2 sdnb 
100 1 |a Reinders, Jörn  |d 1982-  |e VerfasserIn  |0 (DE-588)1068844167  |0 (DE-627)820817708  |0 (DE-576)428211445  |4 aut 
245 1 0 |a How do gait frequency and serum-replacement interval affect polyethylene wear in knee-wear simulator tests?  |c Jörn Reinders, Robert Sonntag, Jan Philippe Kretzer 
264 1 |c 12 July 2014 
300 |a 7 
336 |a Text  |b txt  |2 rdacontent 
337 |a Computermedien  |b c  |2 rdamedia 
338 |a Online-Ressource  |b cr  |2 rdacarrier 
500 |a Gesehen am 18.12.2020 
520 |a Polyethylene wear (PE) is known to be a limiting factor in total joint replacements. However, a standardized wear test (e.g. ISO standard) can only replicate the complex in vivo loading condition in a simplified form. In this study, two different parameters were analyzed: (a) Bovine serum, as a substitute for synovial fluid, is typically replaced every 500,000 cycles. However, a continuous regeneration takes place in vivo. How does serum-replacement interval affect the wear rate of total knee replacements? (b) Patients with an artificial joint show reduced gait frequencies compared to standardized testing. What is the influence of a reduced frequency? Three knee wear tests were run: (a) reference test (ISO), (b) testing with a shortened lubricant replacement interval, (c) testing with reduced frequency. The wear behavior was determined based on gravimetric measurements and wear particle analysis. The results showed that the reduced test frequency only had a small effect on wear behavior. Testing with 1 Hz frequency is therefore a valid method for wear testing. However, testing with a shortened replacement interval nearly doubled the wear rate. Wear particle analysis revealed only small differences in wear particle size between the different tests. Wear particles were not linearly released within one replacement interval. The ISO standard should be revised to address the marked effects of lubricant replacement interval on wear rate. 
650 4 |a Animals 
650 4 |a Biomimetics 
650 4 |a Cattle 
650 4 |a Equipment Failure Analysis 
650 4 |a Gait 
650 4 |a Humans 
650 4 |a Knee Prosthesis 
650 4 |a Lubricants 
650 4 |a Physical Exertion 
650 4 |a Polyethylene 
650 4 |a Prosthesis Design 
650 4 |a Serum 
700 1 |a Sonntag, Robert  |e VerfasserIn  |0 (DE-588)1068843756  |0 (DE-627)82081699X  |0 (DE-576)428211135  |4 aut 
700 1 |a Kretzer, Jan Philippe  |d 1978-  |e VerfasserIn  |0 (DE-588)136998348  |0 (DE-627)589003690  |0 (DE-576)301383103  |4 aut 
773 0 8 |i Enthalten in  |t Journal of materials science  |d Dordrecht : Springer Science + Business Media B.V, 1990  |g 25(2014), 11, Seite 2463-2469  |h Online-Ressource  |w (DE-627)317827316  |w (DE-600)2016995-4  |w (DE-576)107736721  |x 1573-4838  |7 nnas  |a How do gait frequency and serum-replacement interval affect polyethylene wear in knee-wear simulator tests? 
773 1 8 |g volume:25  |g year:2014  |g number:11  |g pages:2463-2469  |g extent:7  |a How do gait frequency and serum-replacement interval affect polyethylene wear in knee-wear simulator tests? 
856 4 0 |u https://doi.org/10.1007/s10856-014-5271-8  |x Verlag  |x Resolving-System  |z lizenzpflichtig  |3 Volltext 
951 |a AR 
992 |a 20201221 
993 |a Article 
994 |a 2014 
998 |g 136998348  |a Kretzer, Jan Philippe  |m 136998348:Kretzer, Jan Philippe  |d 910000  |d 911460  |e 910000PK136998348  |e 911460PK136998348  |k 0/910000/  |k 1/910000/911460/  |p 3  |y j 
998 |g 1068843756  |a Sonntag, Robert  |m 1068843756:Sonntag, Robert  |d 910000  |d 911460  |e 910000PS1068843756  |e 911460PS1068843756  |k 0/910000/  |k 1/910000/911460/  |p 2 
998 |g 1068844167  |a Reinders, Jörn  |m 1068844167:Reinders, Jörn  |d 50000  |e 50000PR1068844167  |k 0/50000/  |p 1  |x j 
999 |a KXP-PPN1743367635  |e 3827955394 
BIB |a Y 
SER |a journal 
JSO |a {"language":["eng"],"person":[{"given":"Jörn","role":"aut","display":"Reinders, Jörn","family":"Reinders"},{"family":"Sonntag","display":"Sonntag, Robert","role":"aut","given":"Robert"},{"display":"Kretzer, Jan Philippe","family":"Kretzer","given":"Jan Philippe","role":"aut"}],"physDesc":[{"extent":"7 S."}],"name":{"displayForm":["Jörn Reinders, Robert Sonntag, Jan Philippe Kretzer"]},"relHost":[{"language":["eng"],"titleAlt":[{"title":"Journal of materials science / Materials in medicine"},{"title":"JSMMEL"},{"title":"Materials in medicine"},{"title":"JMS"}],"part":{"extent":"7","volume":"25","pages":"2463-2469","year":"2014","issue":"11","text":"25(2014), 11, Seite 2463-2469"},"physDesc":[{"extent":"Online-Ressource"}],"recId":"317827316","id":{"eki":["317827316"],"issn":["1573-4838"],"zdb":["2016995-4"]},"origin":[{"dateIssuedDisp":"1990-","dateIssuedKey":"1990","publisher":"Springer Science + Business Media B.V ; Chapman and Hall ; Kluwer","publisherPlace":"Dordrecht ; London ; Dordrecht"}],"note":["Gesehen am 28.10.05"],"pubHistory":["1.1990 -"],"title":[{"subtitle":"JMS","title_sort":"Journal of materials science","title":"Journal of materials science"}],"disp":"How do gait frequency and serum-replacement interval affect polyethylene wear in knee-wear simulator tests?Journal of materials science","type":{"bibl":"periodical","media":"Online-Ressource"}}],"recId":"1743367635","id":{"doi":["10.1007/s10856-014-5271-8"],"eki":["1743367635"]},"title":[{"title_sort":"How do gait frequency and serum-replacement interval affect polyethylene wear in knee-wear simulator tests?","title":"How do gait frequency and serum-replacement interval affect polyethylene wear in knee-wear simulator tests?"}],"origin":[{"dateIssuedKey":"2014","dateIssuedDisp":"12 July 2014"}],"note":["Gesehen am 18.12.2020"],"type":{"bibl":"article-journal","media":"Online-Ressource"}} 
SRT |a REINDERSJOHOWDOGAITF1220