Impact of needle positioning on ablation success of irreversible electroporation: a unicentric retrospective analysis

Irreversible electroporation (IRE) is an ablation procedure in which cell death is induced by ultrashort electrical pulses. In this unicentric retrospective study we investigated the influence of needle positioning on ablation success. 15 IREs with residual tumor after ablation, detected in the firs...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Mathy, René Michael (VerfasserIn) , Tinoush, Parham (VerfasserIn) , Florencia, Ricardo Daniel da (VerfasserIn) , Braun, Alexander (VerfasserIn) , Ghamarnejad, Omid (VerfasserIn) , Radeleff, Boris (VerfasserIn) , Kauczor, Hans-Ulrich (VerfasserIn) , Chang, De-Hua (VerfasserIn)
Dokumenttyp: Article (Journal)
Sprache:Englisch
Veröffentlicht: 14 December 2020
In: Scientific reports
Year: 2020, Jahrgang: 10
ISSN:2045-2322
DOI:10.1038/s41598-020-78660-0
Online-Zugang:Verlag, kostenfrei, Volltext: https://doi.org/10.1038/s41598-020-78660-0
Verlag, lizenzpflichtig, Volltext: http://www.redi-bw.de/db/ebsco.php/search.ebscohost.com/login.aspx%3fdirect%3dtrue%26db%3da9h%26AN%3d147605846%26lang%3dde%26site%3deds-live
Volltext
Verfasserangaben:René Michael Mathy, Parham Tinoush, Ricardo Daniel da Florencia, Alexander Braun, Omid Ghamarnejad, Boris Radeleff, Hans-Ulrich Kauczor & De-Hua Chang

MARC

LEADER 00000caa a2200000 c 4500
001 1745571965
003 DE-627
005 20221123152420.0
007 cr uuu---uuuuu
008 210125s2020 xx |||||o 00| ||eng c
024 7 |a 10.1038/s41598-020-78660-0  |2 doi 
035 |a (DE-627)1745571965 
035 |a (DE-599)KXP1745571965 
035 |a (OCoLC)1341389232 
040 |a DE-627  |b ger  |c DE-627  |e rda 
041 |a eng 
084 |a 33  |2 sdnb 
100 1 |a Mathy, René Michael  |e VerfasserIn  |0 (DE-588)1225699282  |0 (DE-627)1745562877  |4 aut 
245 1 0 |a Impact of needle positioning on ablation success of irreversible electroporation  |b a unicentric retrospective analysis  |c René Michael Mathy, Parham Tinoush, Ricardo Daniel da Florencia, Alexander Braun, Omid Ghamarnejad, Boris Radeleff, Hans-Ulrich Kauczor & De-Hua Chang 
264 1 |c 14 December 2020 
300 |a 10 
336 |a Text  |b txt  |2 rdacontent 
337 |a Computermedien  |b c  |2 rdamedia 
338 |a Online-Ressource  |b cr  |2 rdacarrier 
500 |a Gesehen am 25.01.2021 
520 |a Irreversible electroporation (IRE) is an ablation procedure in which cell death is induced by ultrashort electrical pulses. In this unicentric retrospective study we investigated the influence of needle positioning on ablation success. 15 IREs with residual tumor after ablation, detected in the first follow-up MRI, were included, and compared with 30 successful ablations. Evaluation of needle geometry revealed significantly higher values for needle divergence (NDiv, 7.0° vs. 3.7°, p = 0.02), tumor-center-to-ablation-center distance (TACD, 11.6 vs. 3.2 mm, p < 0.001), tumor-to-needle distance (4.7 vs. 1.9 mm, p = 0.04), and tumor diameter per needle (7.5 vs. 5.9 mm/needle, p = 0.01) in patients with residual tumor. The average number of needles used was higher in the group without residual tumor after ablation (3.1 vs. 2.4, p = 0.04). In many cases with residual tumor, needle depth was too short (2.1 vs. 6.8 mm tumor overlap beyond the most proximal needle tip, p < 0.01). The use of a stereotactic navigation system in 10 cases resulted in a lower NDiv value (2.1° vs. 5.6°, p < 0.01). Thus, correct needle placement seems to be a crucial factor for success and the assistance of a stereotactic navigation system might be helpful. As most important geometrical parameter TACD could be identified. Main reasons for high TACD were insufficient needle depth and a lesion location out of the needle plane. 
650 4 |a ABLATION techniques 
650 4 |a ELECTROPORATION 
650 4 |a MAGNETIC resonance imaging 
650 4 |a NEEDLE biopsy 
650 4 |a STEREOTACTIC radiotherapy 
650 4 |a TUMOR treatment 
650 4 |a ULTRASHORT laser pulses 
700 1 |a Tinoush, Parham  |e VerfasserIn  |0 (DE-588)1167826876  |0 (DE-627)1031340661  |0 (DE-576)51130160X  |4 aut 
700 1 |a Florencia, Ricardo Daniel da  |e VerfasserIn  |0 (DE-588)122570068X  |0 (DE-627)1745566112  |4 aut 
700 1 |a Braun, Alexander  |d 1985-  |e VerfasserIn  |0 (DE-588)1098192923  |0 (DE-627)857480561  |0 (DE-576)469015187  |4 aut 
700 1 |a Ghamarnejad, Omid  |e VerfasserIn  |0 (DE-588)1163283258  |0 (DE-627)102748879X  |0 (DE-576)507869648  |4 aut 
700 1 |a Radeleff, Boris  |d 1969-  |e VerfasserIn  |0 (DE-588)123910846  |0 (DE-627)706470826  |0 (DE-576)293938504  |4 aut 
700 1 |a Kauczor, Hans-Ulrich  |d 1962-  |e VerfasserIn  |0 (DE-588)139267123  |0 (DE-627)70327113X  |0 (DE-576)310955327  |4 aut 
700 1 |a Chang, De-Hua  |d 1978-  |e VerfasserIn  |0 (DE-588)132043939  |0 (DE-627)517295210  |0 (DE-576)298911752  |4 aut 
773 0 8 |i Enthalten in  |t Scientific reports  |d [London] : Springer Nature, 2011  |g 10(2020) Artikel-Nummer 21902, 10 Seiten  |h Online-Ressource  |w (DE-627)663366712  |w (DE-600)2615211-3  |w (DE-576)346641179  |x 2045-2322  |7 nnas  |a Impact of needle positioning on ablation success of irreversible electroporation a unicentric retrospective analysis 
773 1 8 |g volume:10  |g year:2020  |g extent:10  |a Impact of needle positioning on ablation success of irreversible electroporation a unicentric retrospective analysis 
856 4 0 |u https://doi.org/10.1038/s41598-020-78660-0  |x Verlag  |x Resolving-System  |z kostenfrei  |3 Volltext 
856 4 0 |u http://www.redi-bw.de/db/ebsco.php/search.ebscohost.com/login.aspx%3fdirect%3dtrue%26db%3da9h%26AN%3d147605846%26lang%3dde%26site%3deds-live  |x Verlag  |z lizenzpflichtig  |3 Volltext 
951 |a AR 
992 |a 20210125 
993 |a Article 
994 |a 2020 
998 |g 132043939  |a Chang, De-Hua  |m 132043939:Chang, De-Hua  |d 910000  |d 911400  |d 50000  |e 910000PC132043939  |e 911400PC132043939  |e 50000PC132043939  |k 0/910000/  |k 1/910000/911400/  |k 0/50000/  |p 8  |y j 
998 |g 139267123  |a Kauczor, Hans-Ulrich  |m 139267123:Kauczor, Hans-Ulrich  |d 910000  |d 911400  |e 910000PK139267123  |e 911400PK139267123  |k 0/910000/  |k 1/910000/911400/  |p 7 
998 |g 123910846  |a Radeleff, Boris  |m 123910846:Radeleff, Boris  |d 50000  |e 50000PR123910846  |k 0/50000/  |p 6 
998 |g 1098192923  |a Braun, Alexander  |m 1098192923:Braun, Alexander  |d 910000  |d 911400  |e 910000PB1098192923  |e 911400PB1098192923  |k 0/910000/  |k 1/910000/911400/  |p 4 
998 |g 122570068X  |a Florencia, Ricardo Daniel da  |m 122570068X:Florencia, Ricardo Daniel da  |d 910000  |d 911400  |e 910000PF122570068X  |e 911400PF122570068X  |k 0/910000/  |k 1/910000/911400/  |p 3 
998 |g 1167826876  |a Tinoush, Parham  |m 1167826876:Tinoush, Parham  |d 910000  |d 911400  |e 910000PT1167826876  |e 911400PT1167826876  |k 0/910000/  |k 1/910000/911400/  |p 2 
998 |g 1225699282  |a Mathy, René Michael  |m 1225699282:Mathy, René Michael  |d 910000  |d 911400  |e 910000PM1225699282  |e 911400PM1225699282  |k 0/910000/  |k 1/910000/911400/  |p 1  |x j 
999 |a KXP-PPN1745571965  |e 3841982484 
BIB |a Y 
SER |a journal 
JSO |a {"recId":"1745571965","physDesc":[{"extent":"10 S."}],"origin":[{"dateIssuedKey":"2020","dateIssuedDisp":"14 December 2020"}],"person":[{"display":"Mathy, René Michael","family":"Mathy","role":"aut","given":"René Michael"},{"display":"Tinoush, Parham","given":"Parham","role":"aut","family":"Tinoush"},{"display":"Florencia, Ricardo Daniel da","family":"Florencia","role":"aut","given":"Ricardo Daniel da"},{"display":"Braun, Alexander","given":"Alexander","role":"aut","family":"Braun"},{"display":"Ghamarnejad, Omid","family":"Ghamarnejad","role":"aut","given":"Omid"},{"display":"Radeleff, Boris","role":"aut","given":"Boris","family":"Radeleff"},{"display":"Kauczor, Hans-Ulrich","family":"Kauczor","given":"Hans-Ulrich","role":"aut"},{"display":"Chang, De-Hua","given":"De-Hua","role":"aut","family":"Chang"}],"relHost":[{"physDesc":[{"extent":"Online-Ressource"}],"recId":"663366712","disp":"Impact of needle positioning on ablation success of irreversible electroporation a unicentric retrospective analysisScientific reports","pubHistory":["1, article number 1 (2011)-"],"origin":[{"dateIssuedKey":"2011","publisherPlace":"[London] ; London","dateIssuedDisp":"2011-","publisher":"Springer Nature ; Nature Publishing Group"}],"note":["Gesehen am 12.07.24"],"language":["eng"],"type":{"media":"Online-Ressource","bibl":"periodical"},"part":{"extent":"10","text":"10(2020) Artikel-Nummer 21902, 10 Seiten","volume":"10","year":"2020"},"title":[{"title_sort":"Scientific reports","title":"Scientific reports"}],"id":{"zdb":["2615211-3"],"issn":["2045-2322"],"eki":["663366712"]}}],"title":[{"subtitle":"a unicentric retrospective analysis","title":"Impact of needle positioning on ablation success of irreversible electroporation","title_sort":"Impact of needle positioning on ablation success of irreversible electroporation"}],"type":{"media":"Online-Ressource","bibl":"article-journal"},"language":["eng"],"note":["Gesehen am 25.01.2021"],"name":{"displayForm":["René Michael Mathy, Parham Tinoush, Ricardo Daniel da Florencia, Alexander Braun, Omid Ghamarnejad, Boris Radeleff, Hans-Ulrich Kauczor & De-Hua Chang"]},"id":{"doi":["10.1038/s41598-020-78660-0"],"eki":["1745571965"]}} 
SRT |a MATHYRENEMIMPACTOFNE1420