Backside wear in total knee replacement: a new quantitative measurement method and a comparison of polished cobalt-chromium tibial trays with titanium tibial trays

Introduction - In modular total knee replacements, polyethylene backside wear can occur due to micro-motion between the tibial tray and polyethylene insert. Tibial trays manufactured from a cobalt-chromium alloy with highly polished surfaces are increasingly used to prevent backside wear. However, t...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Braun, Steffen (VerfasserIn) , Uhler, Maximilian (VerfasserIn) , Bormann, Therese (VerfasserIn) , Schröder, Stefan (VerfasserIn) , Jäger, Sebastian (VerfasserIn) , Sonntag, Robert (VerfasserIn) , Kretzer, Jan Philippe (VerfasserIn)
Dokumenttyp: Article (Journal)
Sprache:Englisch
Veröffentlicht: 2021
In: Wear
Year: 2020, Jahrgang: 466/467
ISSN:1873-2577
DOI:10.1016/j.wear.2020.203552
Online-Zugang:Verlag, lizenzpflichtig, Volltext: https://doi.org/10.1016/j.wear.2020.203552
Verlag, lizenzpflichtig, Volltext: https://www.sciencedirect.com/science/article/pii/S0043164820310115
Volltext
Verfasserangaben:Steffen Braun, Maximilian Uhler, Therese Bormann, Stefan Schroeder, Sebastian Jaeger, Robert Sonntag, Jan Philippe Kretzer

MARC

LEADER 00000caa a2200000 c 4500
001 1750138972
003 DE-627
005 20240412193205.0
007 cr uuu---uuuuu
008 210302r20212020xx |||||o 00| ||eng c
024 7 |a 10.1016/j.wear.2020.203552  |2 doi 
035 |a (DE-627)1750138972 
035 |a (DE-599)KXP1750138972 
035 |a (OCoLC)1341396718 
040 |a DE-627  |b ger  |c DE-627  |e rda 
041 |a eng 
084 |a 33  |2 sdnb 
100 1 |a Braun, Steffen  |d 1988-  |e VerfasserIn  |0 (DE-588)1149869429  |0 (DE-627)1010183974  |0 (DE-576)496872729  |4 aut 
245 1 0 |a Backside wear in total knee replacement  |b a new quantitative measurement method and a comparison of polished cobalt-chromium tibial trays with titanium tibial trays  |c Steffen Braun, Maximilian Uhler, Therese Bormann, Stefan Schroeder, Sebastian Jaeger, Robert Sonntag, Jan Philippe Kretzer 
264 1 |c 2021 
300 |a 11 
336 |a Text  |b txt  |2 rdacontent 
337 |a Computermedien  |b c  |2 rdamedia 
338 |a Online-Ressource  |b cr  |2 rdacarrier 
500 |a Available online 1 December 2020 
500 |a Gesehen am 03.03.2021 
520 |a Introduction - In modular total knee replacements, polyethylene backside wear can occur due to micro-motion between the tibial tray and polyethylene insert. Tibial trays manufactured from a cobalt-chromium alloy with highly polished surfaces are increasingly used to prevent backside wear. However, there is no quantitative evidence for this supposed wear advantage because previous evaluations only considered the subjective assessment of surface damage and changes without any information about the characteristics of the generated wear products. The validity about a true wear in the form of material removal is often missing. Therefore, this study used a new method to quantitatively investigate the possible wear advantage of polished cobalt-chromium tibial trays. - Materials and methods - Total knee replacements with titanium and polished cobalt-chromium tibial trays (SIGMA® DePuy Synthes & Johnson&Johnson, USA) were tested in a physiological knee wear simulator according to ISO-14243-1 for 3 x 106 cycles. The articulation side was hermetically separated from the backside of the insert. The serum was analyzed separately by particle analysis according to ASTM-F1877. - Results - With polished cobalt-chromium tibial trays, significantly (p = 0.003) less backside wear particles was generated. The backside wear particles differ significantly in size and morphology from the articulating particles. - Conclusion - In this study, backside wear was measured quantitatively and its wear particles were characterized for the first time. We demonstrated that even with modern inlay locking mechanisms and polished surfaces, wear particles on the backside can account for 15-33% of total wear. With polished cobalt-chromium trays, significantly less backside wear particles was generated. 
534 |c 2020 
650 4 |a Backside wear 
650 4 |a Cobalt-chromium-alloy 
650 4 |a Particle analysis 
650 4 |a Polished tibial tray 
650 4 |a Total knee replacement 
700 1 |a Uhler, Maximilian  |e VerfasserIn  |0 (DE-588)1211349799  |0 (DE-627)1699088438  |4 aut 
700 1 |a Bormann, Therese  |e VerfasserIn  |0 (DE-588)1205894829  |0 (DE-627)1691740136  |4 aut 
700 1 |a Schröder, Stefan  |d 1989-  |e VerfasserIn  |0 (DE-588)1191520447  |0 (DE-627)1670141438  |4 aut 
700 1 |a Jäger, Sebastian  |d 1975-  |e VerfasserIn  |0 (DE-588)1059313731  |0 (DE-627)79828238X  |0 (DE-576)415306647  |4 aut 
700 1 |a Sonntag, Robert  |e VerfasserIn  |0 (DE-588)1068843756  |0 (DE-627)82081699X  |0 (DE-576)428211135  |4 aut 
700 1 |a Kretzer, Jan Philippe  |d 1978-  |e VerfasserIn  |0 (DE-588)136998348  |0 (DE-627)589003690  |0 (DE-576)301383103  |4 aut 
773 0 8 |i Enthalten in  |t Wear  |d Amsterdam [u.a.] : Elsevier Science, 1957  |g 466/467(2021), Artikel-ID 203552  |h Online-Ressource  |w (DE-627)306714027  |w (DE-600)1501123-9  |w (DE-576)098474030  |x 1873-2577  |7 nnas  |a Backside wear in total knee replacement a new quantitative measurement method and a comparison of polished cobalt-chromium tibial trays with titanium tibial trays 
773 1 8 |g volume:466/467  |g year:2021  |g elocationid:203552  |g extent:11  |a Backside wear in total knee replacement a new quantitative measurement method and a comparison of polished cobalt-chromium tibial trays with titanium tibial trays 
856 4 0 |u https://doi.org/10.1016/j.wear.2020.203552  |x Verlag  |x Resolving-System  |z lizenzpflichtig  |3 Volltext 
856 4 0 |u https://www.sciencedirect.com/science/article/pii/S0043164820310115  |x Verlag  |z lizenzpflichtig  |3 Volltext 
951 |a AR 
992 |a 20210302 
993 |a Article 
994 |a 2021 
998 |g 136998348  |a Kretzer, Jan Philippe  |m 136998348:Kretzer, Jan Philippe  |d 910000  |d 911460  |d 50000  |e 910000PK136998348  |e 911460PK136998348  |e 50000PK136998348  |k 0/910000/  |k 1/910000/911460/  |k 0/50000/  |p 7  |y j 
998 |g 1068843756  |a Sonntag, Robert  |m 1068843756:Sonntag, Robert  |d 910000  |d 911460  |e 910000PS1068843756  |e 911460PS1068843756  |k 0/910000/  |k 1/910000/911460/  |p 6 
998 |g 1059313731  |a Jäger, Sebastian  |m 1059313731:Jäger, Sebastian  |d 910000  |d 911460  |e 910000PJ1059313731  |e 911460PJ1059313731  |k 0/910000/  |k 1/910000/911460/  |p 5 
998 |g 1191520447  |a Schröder, Stefan  |m 1191520447:Schröder, Stefan  |d 910000  |d 911460  |e 910000PS1191520447  |e 911460PS1191520447  |k 0/910000/  |k 1/910000/911460/  |p 4 
998 |g 1205894829  |a Bormann, Therese  |m 1205894829:Bormann, Therese  |d 910000  |d 911460  |e 910000PB1205894829  |e 911460PB1205894829  |k 0/910000/  |k 1/910000/911460/  |p 3 
998 |g 1211349799  |a Uhler, Maximilian  |m 1211349799:Uhler, Maximilian  |d 910000  |d 911460  |e 910000PU1211349799  |e 911460PU1211349799  |k 0/910000/  |k 1/910000/911460/  |p 2 
998 |g 1149869429  |a Braun, Steffen  |m 1149869429:Braun, Steffen  |d 50000  |e 50000PB1149869429  |k 0/50000/  |p 1  |x j 
999 |a KXP-PPN1750138972  |e 3880118965 
BIB |a Y 
SER |a journal 
JSO |a {"language":["eng"],"name":{"displayForm":["Steffen Braun, Maximilian Uhler, Therese Bormann, Stefan Schroeder, Sebastian Jaeger, Robert Sonntag, Jan Philippe Kretzer"]},"type":{"bibl":"article-journal","media":"Online-Ressource"},"physDesc":[{"extent":"11 S."}],"id":{"doi":["10.1016/j.wear.2020.203552"],"eki":["1750138972"]},"title":[{"subtitle":"a new quantitative measurement method and a comparison of polished cobalt-chromium tibial trays with titanium tibial trays","title":"Backside wear in total knee replacement","title_sort":"Backside wear in total knee replacement"}],"person":[{"family":"Braun","role":"aut","given":"Steffen","display":"Braun, Steffen"},{"given":"Maximilian","display":"Uhler, Maximilian","role":"aut","family":"Uhler"},{"given":"Therese","display":"Bormann, Therese","role":"aut","family":"Bormann"},{"role":"aut","given":"Stefan","display":"Schröder, Stefan","family":"Schröder"},{"role":"aut","given":"Sebastian","display":"Jäger, Sebastian","family":"Jäger"},{"family":"Sonntag","display":"Sonntag, Robert","given":"Robert","role":"aut"},{"family":"Kretzer","given":"Jan Philippe","display":"Kretzer, Jan Philippe","role":"aut"}],"origin":[{"dateIssuedDisp":"2021","dateIssuedKey":"2021"}],"note":["Available online 1 December 2020","Gesehen am 03.03.2021"],"relHost":[{"physDesc":[{"extent":"Online-Ressource"}],"disp":"Backside wear in total knee replacement a new quantitative measurement method and a comparison of polished cobalt-chromium tibial trays with titanium tibial traysWear","part":{"year":"2021","text":"466/467(2021), Artikel-ID 203552","extent":"11","volume":"466/467"},"title":[{"title_sort":"Wear","subtitle":"an international journal on the science and technology of friction, lubrication and wear","title":"Wear"}],"type":{"media":"Online-Ressource","bibl":"periodical"},"recId":"306714027","pubHistory":["1.1957 - 319.2014; Vol. 320.2014 -"],"origin":[{"publisherPlace":"Amsterdam [u.a.]","dateIssuedDisp":"1957-","publisher":"Elsevier Science","dateIssuedKey":"1957"}],"note":["Gesehen am 21.11.23"],"id":{"issn":["1873-2577"],"eki":["306714027"],"zdb":["1501123-9"]},"language":["eng"]}],"recId":"1750138972"} 
SRT |a BRAUNSTEFFBACKSIDEWE2021