The novel Tenzing 7 delivery catheter designed to deliver intermediate catheters to the face of embolus without crossing: clinical performance predicted in anatomically challenging model
<h3>Background</h3> <p>In large vessel occlusionstroke, navigation of aspiration catheters (AC) can be impeded by vessel tortuosity and the ophthalmic artery origin. A novel tapered delivery catheter was designed to facilitate delivery without disturbing the embolus. We assessed AC...
Saved in:
| Main Authors: | , , , , , , , , , , , |
|---|---|
| Format: | Article (Journal) |
| Language: | English |
| Published: |
July 14, 2021
|
| In: |
Journal of neuroInterventional surgery
Year: 2021, Volume: 13, Issue: 8, Pages: 722-726 |
| ISSN: | 1759-8486 |
| DOI: | 10.1136/neurintsurg-2020-016412 |
| Online Access: | Verlag, lizenzpflichtig, Volltext: https://doi.org/10.1136/neurintsurg-2020-016412 Verlag, lizenzpflichtig, Volltext: https://jnis.bmj.com/content/13/8/722 |
| Author Notes: | Andreas Maximilian Frölich, Warren Kim, Knut Stribrny, Olav Jansen, Markus Möhlenbruch, Istvan Szikora, Fritz Wodarg, Jens Fiehler, Kim Otto, Tony Chou, Jan Hendrik Buhk, Joey English |
MARC
| LEADER | 00000caa a2200000 c 4500 | ||
|---|---|---|---|
| 001 | 1774580381 | ||
| 003 | DE-627 | ||
| 005 | 20220820061519.0 | ||
| 007 | cr uuu---uuuuu | ||
| 008 | 211020s2021 xx |||||o 00| ||eng c | ||
| 024 | 7 | |a 10.1136/neurintsurg-2020-016412 |2 doi | |
| 035 | |a (DE-627)1774580381 | ||
| 035 | |a (DE-599)KXP1774580381 | ||
| 035 | |a (OCoLC)1341422109 | ||
| 040 | |a DE-627 |b ger |c DE-627 |e rda | ||
| 041 | |a eng | ||
| 084 | |a 33 |2 sdnb | ||
| 100 | 1 | |a Frölich, Andreas M. J. |e VerfasserIn |0 (DE-588)1171138687 |0 (DE-627)1040379044 |0 (DE-576)513599827 |4 aut | |
| 245 | 1 | 4 | |a The novel Tenzing 7 delivery catheter designed to deliver intermediate catheters to the face of embolus without crossing |b clinical performance predicted in anatomically challenging model |c Andreas Maximilian Frölich, Warren Kim, Knut Stribrny, Olav Jansen, Markus Möhlenbruch, Istvan Szikora, Fritz Wodarg, Jens Fiehler, Kim Otto, Tony Chou, Jan Hendrik Buhk, Joey English |
| 264 | 1 | |c July 14, 2021 | |
| 300 | |a 5 | ||
| 336 | |a Text |b txt |2 rdacontent | ||
| 337 | |a Computermedien |b c |2 rdamedia | ||
| 338 | |a Online-Ressource |b cr |2 rdacarrier | ||
| 500 | |a Gesehen am 20.10.2021 | ||
| 520 | |a <h3>Background</h3> <p>In large vessel occlusionstroke, navigation of aspiration catheters (AC) can be impeded by vessel tortuosity and the ophthalmic artery origin. A novel tapered delivery catheter was designed to facilitate delivery without disturbing the embolus. We assessed AC deliverability in vitro and validated the observations in a first-in-human experience.</p><h3>Methods</h3> <p>In a vascular model with three challenging craniocervical scenarios, two commercial AC were advanced from the carotid to the middle cerebral artery by four neurointerventionalists. Catheter deliverability with standard microwire and microcatheter (MC) combinations and the Tenzing 7 (T7) Delivery Catheter (Route 92 Medical, San Mateo, CA) were compared. Operators rated aspects of catheter deliverability on a 5-point scale. Results were compared with device delivery patterns at a neurovascular center before and after clinical introduction of T7.</p><h3>Results</h3> <p>In vitro, success rate and speed were higher with T7 (96%; mean 30±10 s) than with MC (65%; 72±47 s, p<0.001 each), with fewer interactions with the occlusion site (T7: 54% vs MC: 77%, p=0.004). T7 received superior ratings regarding carotid artery deflection (T7: 2, IQR1-3 vs MC: 3, IQR2-3, p<0.001), guide catheter pushback (T7: 2, IQR1-3 vs MC: 3, IQR3-3, p<0.001) and ophthalmic artery passage (T7: 1.5, IQR1-2 vs MC: 4, IQR3-5, p<0.001). Before introduction of T7 at a single center, delivery of AC to a large vessel occlusion without crossing was achieved in 15/123 cases (12%). With T7, this rate was 28/31 patients (90.3%).</p><h3>Conclusion</h3> <p>Compared with microcatheter and microwire combinations, T7 improves aspiration catheter delivery in vitro, minimizing the need to cross the occlusion. Initial clinical experience appears to validate the model’s observations.</p> | ||
| 700 | 1 | |a Kim, Warren |e VerfasserIn |4 aut | |
| 700 | 1 | |a Stribrny, Knut |e VerfasserIn |4 aut | |
| 700 | 1 | |a Jansen, Olav |e VerfasserIn |4 aut | |
| 700 | 1 | |a Möhlenbruch, Markus Alfred |d 1979- |e VerfasserIn |0 (DE-588)137693591 |0 (DE-627)594845041 |0 (DE-576)304845655 |4 aut | |
| 700 | 1 | |a Szikora, Istvan |e VerfasserIn |4 aut | |
| 700 | 1 | |a Wodarg, Fritz |e VerfasserIn |4 aut | |
| 700 | 1 | |a Fiehler, Jens |e VerfasserIn |4 aut | |
| 700 | 1 | |a Otto, Kim |e VerfasserIn |4 aut | |
| 700 | 1 | |a Chou, Tony |e VerfasserIn |4 aut | |
| 700 | 1 | |a Buhk, Jan Hendrik |e VerfasserIn |4 aut | |
| 700 | 1 | |a English, Joey |e VerfasserIn |4 aut | |
| 773 | 0 | 8 | |i Enthalten in |t Journal of neuroInterventional surgery |d London : BMJ Journals, 2009 |g 13(2021), 8, Seite 722-726 |h Online-Ressource |w (DE-627)605217548 |w (DE-600)2506028-4 |w (DE-576)308906721 |x 1759-8486 |7 nnas |a The novel Tenzing 7 delivery catheter designed to deliver intermediate catheters to the face of embolus without crossing clinical performance predicted in anatomically challenging model |
| 773 | 1 | 8 | |g volume:13 |g year:2021 |g number:8 |g pages:722-726 |g extent:5 |a The novel Tenzing 7 delivery catheter designed to deliver intermediate catheters to the face of embolus without crossing clinical performance predicted in anatomically challenging model |
| 856 | 4 | 0 | |u https://doi.org/10.1136/neurintsurg-2020-016412 |x Verlag |x Resolving-System |z lizenzpflichtig |3 Volltext |
| 856 | 4 | 0 | |u https://jnis.bmj.com/content/13/8/722 |x Verlag |z lizenzpflichtig |3 Volltext |
| 951 | |a AR | ||
| 992 | |a 20211020 | ||
| 993 | |a Article | ||
| 994 | |a 2021 | ||
| 998 | |g 137693591 |a Möhlenbruch, Markus Alfred |m 137693591:Möhlenbruch, Markus Alfred |d 910000 |d 911100 |d 50000 |e 910000PM137693591 |e 911100PM137693591 |e 50000PM137693591 |k 0/910000/ |k 1/910000/911100/ |k 0/50000/ |p 5 | ||
| 999 | |a KXP-PPN1774580381 |e 399156999X | ||
| BIB | |a Y | ||
| SER | |a journal | ||
| JSO | |a {"physDesc":[{"extent":"5 S."}],"id":{"doi":["10.1136/neurintsurg-2020-016412"],"eki":["1774580381"]},"type":{"bibl":"article-journal","media":"Online-Ressource"},"title":[{"title_sort":"novel Tenzing 7 delivery catheter designed to deliver intermediate catheters to the face of embolus without crossing","subtitle":"clinical performance predicted in anatomically challenging model","title":"The novel Tenzing 7 delivery catheter designed to deliver intermediate catheters to the face of embolus without crossing"}],"origin":[{"dateIssuedKey":"2021","dateIssuedDisp":"July 14, 2021"}],"relHost":[{"title":[{"title_sort":"Journal of neuroInterventional surgery","title":"Journal of neuroInterventional surgery","subtitle":"JNIS ; the journal of the Society of NeuroInterventional Surgery"}],"id":{"zdb":["2506028-4"],"issn":["1759-8486"],"eki":["605217548"]},"physDesc":[{"extent":"Online-Ressource"}],"origin":[{"dateIssuedDisp":"2009-","publisher":"BMJ Journals","publisherPlace":"London","dateIssuedKey":"2009"}],"recId":"605217548","disp":"The novel Tenzing 7 delivery catheter designed to deliver intermediate catheters to the face of embolus without crossing clinical performance predicted in anatomically challenging modelJournal of neuroInterventional surgery","type":{"media":"Online-Ressource","bibl":"periodical"},"titleAlt":[{"title":"JNIS"}],"language":["eng"],"part":{"extent":"5","pages":"722-726","issue":"8","text":"13(2021), 8, Seite 722-726","year":"2021","volume":"13"},"pubHistory":["1.2009 -"],"note":["Gesehen am 03.01.11"]}],"name":{"displayForm":["Andreas Maximilian Frölich, Warren Kim, Knut Stribrny, Olav Jansen, Markus Möhlenbruch, Istvan Szikora, Fritz Wodarg, Jens Fiehler, Kim Otto, Tony Chou, Jan Hendrik Buhk, Joey English"]},"language":["eng"],"note":["Gesehen am 20.10.2021"],"person":[{"given":"Andreas M. J.","display":"Frölich, Andreas M. J.","family":"Frölich","role":"aut"},{"family":"Kim","role":"aut","display":"Kim, Warren","given":"Warren"},{"role":"aut","family":"Stribrny","given":"Knut","display":"Stribrny, Knut"},{"family":"Jansen","role":"aut","given":"Olav","display":"Jansen, Olav"},{"given":"Markus Alfred","display":"Möhlenbruch, Markus Alfred","role":"aut","family":"Möhlenbruch"},{"display":"Szikora, Istvan","given":"Istvan","role":"aut","family":"Szikora"},{"role":"aut","family":"Wodarg","display":"Wodarg, Fritz","given":"Fritz"},{"given":"Jens","display":"Fiehler, Jens","family":"Fiehler","role":"aut"},{"display":"Otto, Kim","given":"Kim","family":"Otto","role":"aut"},{"role":"aut","family":"Chou","display":"Chou, Tony","given":"Tony"},{"display":"Buhk, Jan Hendrik","given":"Jan Hendrik","role":"aut","family":"Buhk"},{"role":"aut","family":"English","given":"Joey","display":"English, Joey"}],"recId":"1774580381"} | ||
| SRT | |a FROELICHANNOVELTENZI1420 | ||