Navigated renal access using electromagnetic tracking: an initial experience

Background and aim: Navigation systems are promising tools for improving efficacy and safety in surgical endoscopy and other minimally invasive techniques. The aim of the current study is to investigate electromagnetic tracking (EMT) for navigated renal access in a porcine model. Methods: For our pr...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Huber, Johannes (VerfasserIn) , Wegner, Ingmar (VerfasserIn) , Meinzer, Hans-Peter (VerfasserIn) , Hallscheidt, Peter (VerfasserIn) , Hadaschik, Boris (VerfasserIn) , Pahernik, Sascha (VerfasserIn) , Hohenfellner, Markus (VerfasserIn)
Dokumenttyp: Article (Journal)
Sprache:Englisch
Veröffentlicht: 2011
In: Surgical endoscopy and other interventional techniques
Year: 2011, Jahrgang: 25, Heft: 4, Pages: 1307-1312
ISSN:1432-2218
DOI:10.1007/s00464-010-1338-x
Online-Zugang:Verlag, lizenzpflichtig, Volltext: https://doi.org/10.1007/s00464-010-1338-x
Volltext
Verfasserangaben:Johannes Huber, Ingmar Wegner, Hans-Peter Meinzer, Peter Hallscheidt, Boris Hadaschik, Sascha Pahernik, Markus Hohenfellner

MARC

LEADER 00000caa a2200000 c 4500
001 180870438X
003 DE-627
005 20220820211812.0
007 cr uuu---uuuuu
008 220630s2011 xx |||||o 00| ||eng c
024 7 |a 10.1007/s00464-010-1338-x  |2 doi 
035 |a (DE-627)180870438X 
035 |a (DE-599)KXP180870438X 
035 |a (OCoLC)1341463357 
040 |a DE-627  |b ger  |c DE-627  |e rda 
041 |a eng 
084 |a 33  |2 sdnb 
100 1 |a Huber, Johannes  |d 1980-  |e VerfasserIn  |0 (DE-588)1066882967  |0 (DE-627)817999566  |0 (DE-576)426260554  |4 aut 
245 1 0 |a Navigated renal access using electromagnetic tracking  |b an initial experience  |c Johannes Huber, Ingmar Wegner, Hans-Peter Meinzer, Peter Hallscheidt, Boris Hadaschik, Sascha Pahernik, Markus Hohenfellner 
264 1 |c 2011 
300 |a 6 
336 |a Text  |b txt  |2 rdacontent 
337 |a Computermedien  |b c  |2 rdamedia 
338 |a Online-Ressource  |b cr  |2 rdacarrier 
500 |a Published online: 11 September 2010 
500 |a Gesehen am 30.06.2022 
520 |a Background and aim: Navigation systems are promising tools for improving efficacy and safety in surgical endoscopy and other minimally invasive techniques. The aim of the current study is to investigate electromagnetic tracking (EMT) for navigated renal access in a porcine model. Methods: For our proof-of-principle study we modified a recently established porcine ex vivo model. Via a ureteral catheter which was placed into the desired puncture site, a small sensor was introduced and located by EMT. Then, a tracked needle was navigated into the collecting system in a “rendezvous” approach. A total of 90 renal tracts were obtained in six kidneys using EMT, with a maximum of three punctures allowed per intervention. For each puncture, number of attempts to success, final distance to probe, puncture time, and localization were assessed. We compared absolute and relative frequencies using the chi-square test and applied the Mann–Whitney U-test for continuous variables. Results: No major problems were encountered performing the experiment. Access to the collecting system was successfully obtained after a single puncture in 91% (82/90) and within a second attempt in the remaining 9% (8/90). Thus, a 100% success rate was reached after a maximum of two punctures. Location of the calyx did not have a significant effect on success rate (p = 0.637). After a learning phase of 30 punctures, higher success rate (96% versus 83%; p = 0.041) was accomplished within shorter puncture time (14 versus 17 s; p = 0.049) and with higher precision (1.7 versus 2.8 mm; p < 0.001). Conclusions: With respect to other established techniques, use of EMT seems to decrease the number of attempts and procedural time remarkably. This might contribute to greater safety and efficacy when applied clinically. The presented approach appears to be promising, especially in difficult settings, provided that in vivo data support these initial results. 
650 4 |a Electromagnetic tracking 
650 4 |a Endourology 
650 4 |a Navigation 
650 4 |a Percutaneous renal access 
700 1 |a Wegner, Ingmar  |d 1977-  |e VerfasserIn  |0 (DE-588)137032994  |0 (DE-627)69483842X  |0 (DE-576)298616580  |4 aut 
700 1 |a Meinzer, Hans-Peter  |d 1948-2023  |e VerfasserIn  |0 (DE-588)10927279X  |0 (DE-627)516177990  |0 (DE-576)289636914  |4 aut 
700 1 |a Hallscheidt, Peter  |e VerfasserIn  |0 (DE-588)1072530686  |0 (DE-627)827694156  |0 (DE-576)173264670  |4 aut 
700 1 |a Hadaschik, Boris  |d 1974-  |e VerfasserIn  |0 (DE-588)124300553  |0 (DE-627)085770930  |0 (DE-576)294109013  |4 aut 
700 1 |a Pahernik, Sascha  |d 1968-  |e VerfasserIn  |0 (DE-588)122063163  |0 (DE-627)081713495  |0 (DE-576)293073147  |4 aut 
700 1 |a Hohenfellner, Markus  |d 1958-  |e VerfasserIn  |0 (DE-588)133862518  |0 (DE-627)557857988  |0 (DE-576)300155263  |4 aut 
773 0 8 |i Enthalten in  |t Surgical endoscopy and other interventional techniques  |d New York : Springer-Verlag, 2002  |g 25(2011), 4, Seite 1307-1312  |h Online-Ressource  |w (DE-627)254909620  |w (DE-600)1463171-4  |w (DE-576)074754106  |x 1432-2218  |7 nnas  |a Navigated renal access using electromagnetic tracking an initial experience 
773 1 8 |g volume:25  |g year:2011  |g number:4  |g pages:1307-1312  |g extent:6  |a Navigated renal access using electromagnetic tracking an initial experience 
856 4 0 |u https://doi.org/10.1007/s00464-010-1338-x  |x Verlag  |x Resolving-System  |z lizenzpflichtig  |3 Volltext 
951 |a AR 
992 |a 20220630 
993 |a Article 
994 |a 2011 
998 |g 133862518  |a Hohenfellner, Markus  |m 133862518:Hohenfellner, Markus  |d 910000  |d 910200  |e 910000PH133862518  |e 910200PH133862518  |k 0/910000/  |k 1/910000/910200/  |p 7  |y j 
998 |g 122063163  |a Pahernik, Sascha  |m 122063163:Pahernik, Sascha  |d 910000  |d 910200  |d 50000  |e 910000PP122063163  |e 910200PP122063163  |e 50000PP122063163  |k 0/910000/  |k 1/910000/910200/  |k 0/50000/  |p 6 
998 |g 124300553  |a Hadaschik, Boris  |m 124300553:Hadaschik, Boris  |d 910000  |d 910200  |e 910000PH124300553  |e 910200PH124300553  |k 0/910000/  |k 1/910000/910200/  |p 5 
998 |g 1072530686  |a Hallscheidt, Peter  |m 1072530686:Hallscheidt, Peter  |d 50000  |e 50000PH1072530686  |k 0/50000/  |p 4 
998 |g 10927279X  |a Meinzer, Hans-Peter  |m 10927279X:Meinzer, Hans-Peter  |d 50000  |e 50000PM10927279X  |k 0/50000/  |p 3 
998 |g 137032994  |a Wegner, Ingmar  |m 137032994:Wegner, Ingmar  |d 50000  |e 50000PW137032994  |k 0/50000/  |p 2 
999 |a KXP-PPN180870438X  |e 4159514723 
BIB |a Y 
SER |a journal 
JSO |a {"title":[{"subtitle":"an initial experience","title_sort":"Navigated renal access using electromagnetic tracking","title":"Navigated renal access using electromagnetic tracking"}],"language":["eng"],"person":[{"role":"aut","family":"Huber","display":"Huber, Johannes","given":"Johannes"},{"role":"aut","family":"Wegner","given":"Ingmar","display":"Wegner, Ingmar"},{"display":"Meinzer, Hans-Peter","given":"Hans-Peter","family":"Meinzer","role":"aut"},{"given":"Peter","display":"Hallscheidt, Peter","family":"Hallscheidt","role":"aut"},{"display":"Hadaschik, Boris","given":"Boris","family":"Hadaschik","role":"aut"},{"given":"Sascha","display":"Pahernik, Sascha","family":"Pahernik","role":"aut"},{"role":"aut","given":"Markus","display":"Hohenfellner, Markus","family":"Hohenfellner"}],"origin":[{"dateIssuedKey":"2011","dateIssuedDisp":"2011"}],"name":{"displayForm":["Johannes Huber, Ingmar Wegner, Hans-Peter Meinzer, Peter Hallscheidt, Boris Hadaschik, Sascha Pahernik, Markus Hohenfellner"]},"recId":"180870438X","note":["Published online: 11 September 2010","Gesehen am 30.06.2022"],"relHost":[{"title":[{"title":"Surgical endoscopy and other interventional techniques","subtitle":"official journal of the Society of American Gastrointestinal Endoscopic Surgeons (SAGES) and European Association for Endoscopic Surgery (E.A.E.S.)","title_sort":"Surgical endoscopy and other interventional techniques"}],"origin":[{"publisherPlace":"New York","dateIssuedKey":"2002","dateIssuedDisp":"2002-","publisher":"Springer-Verlag"}],"note":["Gesehen am 19.06.24"],"physDesc":[{"extent":"Online-Ressource"}],"id":{"eki":["254909620"],"issn":["1432-2218"],"zdb":["1463171-4"]},"part":{"year":"2011","pages":"1307-1312","text":"25(2011), 4, Seite 1307-1312","volume":"25","extent":"6","issue":"4"},"language":["eng"],"disp":"Navigated renal access using electromagnetic tracking an initial experienceSurgical endoscopy and other interventional techniques","recId":"254909620","pubHistory":["Volume 16, issue 1 (January 2002)-"],"type":{"bibl":"periodical","media":"Online-Ressource"}}],"physDesc":[{"extent":"6 S."}],"id":{"doi":["10.1007/s00464-010-1338-x"],"eki":["180870438X"]},"type":{"bibl":"article-journal","media":"Online-Ressource"}} 
SRT |a HUBERJOHANNAVIGATEDR2011