Metrologically based procedures for the temperature, heat and heat flow rate calibration of DSC

Metrologically based measuring procedures and evaluation methods are recommended as guidance for practical temperature, heat and heat flow rate calibration of DSC instruments which are largely independent of instrumental, test and sample parameters. The relevant terms are defined, the measuring proc...

Full description

Saved in:
Bibliographic Details
Main Authors: Sarge, Stefan M. (Author) , Hemminger, W. (Author) , Gmelin, E. (Author) , Höhne, G. W. H. (Author) , Cammenga, H. K. (Author) , Eysel, Walter (Author)
Format: Article (Journal)
Language:English
Published: 1997
In: Journal of thermal analysis and calorimetry
Year: 1997, Volume: 49, Issue: 2, Pages: 1125-1134
ISSN:1588-2926
DOI:10.1007/BF01996802
Online Access:Verlag, lizenzpflichtig, Volltext: https://doi.org/10.1007/BF01996802
Get full text
Author Notes:S.M. Sarge, W. Hemminger, E. Gmelin, G.W.H. Höhne, H.K. Cammenga, W. Eysel
Description
Summary:Metrologically based measuring procedures and evaluation methods are recommended as guidance for practical temperature, heat and heat flow rate calibration of DSC instruments which are largely independent of instrumental, test and sample parameters. The relevant terms are defined, the measuring procedures and evaluation methods described, calibration materials and their characteristic data stated and guidance for the sample handling provided. Reference is made to three extended papers on calibration. The recommendations were developed by the working group ‘Calibration of Scanning Calorimeters’ of the German Society of Thermal Analysis (GEFTA).
Item Description:Gesehen am 22.11.2022
Physical Description:Online Resource
ISSN:1588-2926
DOI:10.1007/BF01996802