The PETcoil project: PET performance evaluation of two detector modules for a second generation RF-penetrable TOF-PET brain dedicated insert for simultaneous PET/MRI

Objective. We are developing a portable, ‘RF-penetrable’, brain-dedicated time of flight (TOF)-PET insert (PETcoil) for simultaneous PET/MRI. Approach. In this paper, we evaluate the PET performance of two fully assembled detector modules for this insert design outside the MR room. Main results. The...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Dong, Qian (VerfasserIn) , Chang, Chen-Ming (VerfasserIn) , Lee, Brian J. (VerfasserIn) , Sacco, Ilaria (VerfasserIn) , Sajedi, Salar (VerfasserIn) , Adams, Zander (VerfasserIn) , Fischer, Peter (VerfasserIn) , Levin, Craig S. (VerfasserIn)
Dokumenttyp: Article (Journal)
Sprache:Englisch
Veröffentlicht: 5 April 2023
In: Physics in medicine and biology
Year: 2023, Jahrgang: 68, Heft: 8, Pages: 1-15
ISSN:1361-6560
DOI:10.1088/1361-6560/acc3f2
Online-Zugang:Verlag, lizenzpflichtig, Volltext: https://doi.org/10.1088/1361-6560/acc3f2
Verlag, lizenzpflichtig, Volltext: https://dx.doi.org/10.1088/1361-6560/acc3f2
Volltext
Verfasserangaben:Qian Dong, Chen-Ming Chang, Brian J Lee, Ilaria Sacco, Salar Sajedi, Zander Adams, Peter Fischer and Craig S Levin
Beschreibung
Zusammenfassung:Objective. We are developing a portable, ‘RF-penetrable’, brain-dedicated time of flight (TOF)-PET insert (PETcoil) for simultaneous PET/MRI. Approach. In this paper, we evaluate the PET performance of two fully assembled detector modules for this insert design outside the MR room. Main results. The global coincidence time resolution, global 511 keV energy resolution, coincidence count rate, and detector temperature achieved over 2 h data collection were 242.2 ± 0.4 ps full width at half maximum (FWHM), 11.19% ± 0.02% FWHM, 22.0 ± 0.1 kcps, and 23.5 °C ± 0.3 °C, respectively. The intrinsic spatial resolutions in the axial and transaxial directions were 2.74 ± 0.01 mm FWHM and 2.88 ± 0.03 mm FWHM, respectively. Significance. These results demonstrate excellent TOF capability and the performance and stability necessary for scaling up to a full ring comprising 16 detector modules.
Beschreibung:Gesehen am 24.05.2023
Beschreibung:Online Resource
ISSN:1361-6560
DOI:10.1088/1361-6560/acc3f2