Sample cells for probing solid/liquid interfaces with broadband sum-frequency-generation spectroscopy

Two sample cells designed specifically for sum-frequency-generation (SFG) measurements at the solid/liquid interface were developed: one thin-layer analysis cell allowing measurement of films on reflective metallic surfaces through a micrometer layer of solution and one spectroelectrochemical cell a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Verreault, Dominique (VerfasserIn) , Kurz, Volker L. (VerfasserIn) , Howell, Caitlin (VerfasserIn) , Kölsch, Patrick (VerfasserIn)
Dokumenttyp: Article (Journal)
Sprache:Englisch
Veröffentlicht: 24 June 2010
In: Review of scientific instruments
Year: 2010, Jahrgang: 81, Heft: 6, Pages: 1-10
ISSN:1089-7623
DOI:10.1063/1.3443096
Online-Zugang:Resolving-System, lizenzpflichtig, Volltext: https://doi.org/10.1063/1.3443096
Verlag, lizenzpflichtig, Volltext: https://pubs.aip.org/aip/rsi/article/81/6/063111/352508/Sample-cells-for-probing-solid-liquid-interfaces
Volltext
Verfasserangaben:Dominique Verreault, Volker Kurz, Caitlin Howell, and Patrick Koelsch

MARC

LEADER 00000caa a2200000 c 4500
001 1858789362
003 DE-627
005 20240311124832.0
007 cr uuu---uuuuu
008 230905s2010 xx |||||o 00| ||eng c
024 7 |a 10.1063/1.3443096  |2 doi 
035 |a (DE-627)1858789362 
035 |a (DE-599)KXP1858789362 
035 |a (OCoLC)1425873831 
040 |a DE-627  |b ger  |c DE-627  |e rda 
041 |a eng 
084 |a 32  |2 sdnb 
100 1 |a Verreault, Dominique  |e VerfasserIn  |0 (DE-588)1013552512  |0 (DE-627)664616577  |0 (DE-576)34731998X  |4 aut 
245 1 0 |a Sample cells for probing solid/liquid interfaces with broadband sum-frequency-generation spectroscopy  |c Dominique Verreault, Volker Kurz, Caitlin Howell, and Patrick Koelsch 
246 3 0 |a solid / liquid 
264 1 |c 24 June 2010 
300 |a 10 
336 |a Text  |b txt  |2 rdacontent 
337 |a Computermedien  |b c  |2 rdamedia 
338 |a Online-Ressource  |b cr  |2 rdacarrier 
500 |a Gesehen am 05.09.2023 
520 |a Two sample cells designed specifically for sum-frequency-generation (SFG) measurements at the solid/liquid interface were developed: one thin-layer analysis cell allowing measurement of films on reflective metallic surfaces through a micrometer layer of solution and one spectroelectrochemical cell allowing investigation of processes at the indium tin oxide/solution interface. Both sample cells are described in detail and data illustrating the capabilities of each are shown. To further improve measurements at solid/liquid interfaces, the broadband SFG system was modified to include a reference beam which can be measured simultaneously with the sample signal, permitting background correction of SFG spectra in real time. Sensitivity tests of this system yielded a signal-to-noise ratio of 100 at a surface coverage of 0.2 molecules/nm2. Details on data analysis routines, pulse shaping methods of the visible beam, as well as the design of a purging chamber and sample stage setup are presented. These descriptions will be useful to those planning to set up a SFG spectrometer or seeking to optimize their own SFG systems for measurements of solid/liquid interfaces. 
700 1 |a Kurz, Volker L.  |e VerfasserIn  |0 (DE-588)143809369  |0 (DE-627)704608855  |0 (DE-576)337266654  |4 aut 
700 1 |a Howell, Caitlin  |e VerfasserIn  |0 (DE-588)1016481624  |0 (DE-627)705534642  |0 (DE-576)352262184  |4 aut 
700 1 |a Kölsch, Patrick  |e VerfasserIn  |0 (DE-588)1161222677  |0 (DE-627)102462823X  |0 (DE-576)506417247  |4 aut 
773 0 8 |i Enthalten in  |t Review of scientific instruments  |d [Melville, NY] : AIP Publishing, 1930  |g 81(2010), 6 vom: Juni, Artikel-ID 063111, Seite 1-10  |h Online-Ressource  |w (DE-627)268759324  |w (DE-600)1472905-2  |w (DE-576)077609948  |x 1089-7623  |7 nnas  |a Sample cells for probing solid/liquid interfaces with broadband sum-frequency-generation spectroscopy 
773 1 8 |g volume:81  |g year:2010  |g number:6  |g month:06  |g elocationid:063111  |g pages:1-10  |g extent:10  |a Sample cells for probing solid/liquid interfaces with broadband sum-frequency-generation spectroscopy 
856 4 0 |u https://doi.org/10.1063/1.3443096  |x Resolving-System  |x Verlag  |z lizenzpflichtig  |3 Volltext 
856 4 0 |u https://pubs.aip.org/aip/rsi/article/81/6/063111/352508/Sample-cells-for-probing-solid-liquid-interfaces  |x Verlag  |z lizenzpflichtig  |3 Volltext 
951 |a AR 
992 |a 20230905 
993 |a Article 
994 |a 2010 
998 |g 1161222677  |a Kölsch, Patrick  |m 1161222677:Kölsch, Patrick  |d 120000  |d 120300  |e 120000PK1161222677  |e 120300PK1161222677  |k 0/120000/  |k 1/120000/120300/  |p 4  |y j 
998 |g 1016481624  |a Howell, Caitlin  |m 1016481624:Howell, Caitlin  |d 120000  |e 120000PH1016481624  |k 0/120000/  |p 3 
998 |g 143809369  |a Kurz, Volker L.  |m 143809369:Kurz, Volker L.  |d 120000  |e 120000PK143809369  |k 0/120000/  |p 2 
998 |g 1013552512  |a Verreault, Dominique  |m 1013552512:Verreault, Dominique  |d 120000  |e 120000PV1013552512  |k 0/120000/  |p 1  |x j 
999 |a KXP-PPN1858789362  |e 4372792336 
BIB |a Y 
SER |a journal 
JSO |a {"physDesc":[{"extent":"10 S."}],"relHost":[{"pubHistory":["1.1930 -"],"part":{"year":"2010","issue":"6","pages":"1-10","text":"81(2010), 6 vom: Juni, Artikel-ID 063111, Seite 1-10","volume":"81","extent":"10"},"disp":"Sample cells for probing solid/liquid interfaces with broadband sum-frequency-generation spectroscopyReview of scientific instruments","note":["Gesehen am 12.07.24"],"type":{"bibl":"periodical","media":"Online-Ressource"},"language":["eng"],"corporate":[{"role":"isb","display":"American Institute of Physics","roleDisplay":"Herausgebendes Organ"}],"recId":"268759324","title":[{"title_sort":"Review of scientific instruments","subtitle":"a monthly journal devoted to scientific instruments, apparatus, and techniques","title":"Review of scientific instruments"}],"physDesc":[{"extent":"Online-Ressource"}],"origin":[{"dateIssuedDisp":"2013-","dateIssuedKey":"2013","publisher":"AIP Publishing ; American Institute of Physics","publisherPlace":"[Melville, NY] ; [Woodbury, Long Island, NY]"}],"id":{"eki":["268759324"],"zdb":["1472905-2"],"issn":["1089-7623"]},"name":{"displayForm":["publ. by the American Institute of Physics"]}}],"name":{"displayForm":["Dominique Verreault, Volker Kurz, Caitlin Howell, and Patrick Koelsch"]},"origin":[{"dateIssuedKey":"2010","dateIssuedDisp":"24 June 2010"}],"id":{"doi":["10.1063/1.3443096"],"eki":["1858789362"]},"note":["Gesehen am 05.09.2023"],"type":{"media":"Online-Ressource","bibl":"article-journal"},"language":["eng"],"recId":"1858789362","person":[{"family":"Verreault","given":"Dominique","display":"Verreault, Dominique","roleDisplay":"VerfasserIn","role":"aut"},{"given":"Volker L.","family":"Kurz","role":"aut","display":"Kurz, Volker L.","roleDisplay":"VerfasserIn"},{"family":"Howell","given":"Caitlin","display":"Howell, Caitlin","roleDisplay":"VerfasserIn","role":"aut"},{"roleDisplay":"VerfasserIn","display":"Kölsch, Patrick","role":"aut","family":"Kölsch","given":"Patrick"}],"title":[{"title":"Sample cells for probing solid/liquid interfaces with broadband sum-frequency-generation spectroscopy","title_sort":"Sample cells for probing solid/liquid interfaces with broadband sum-frequency-generation spectroscopy"}]} 
SRT |a VERREAULTDSAMPLECELL2420