Comparative evaluation of a surface-based respiratory monitoring system against a pressure sensor for 4DCT image reconstruction in phantoms

Four-dimensional computed tomography (4DCT), which relies on breathing-induced motion, requires realistic surrogate information of breathing variations to reconstruct the tumor trajectory and motion variability of normal tissues accurately. Therefore, the SimRT surface-guided respiratory monitoring...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Qubala, Abdallah (VerfasserIn) , Shafee, Jehad (VerfasserIn) , Batista, Vania (VerfasserIn) , Liermann, Jakob (VerfasserIn) , Winter, Marcus (VerfasserIn) , Piro, Daniel (VerfasserIn) , Jäkel, Oliver (VerfasserIn)
Dokumenttyp: Article (Journal)
Sprache:Englisch
Veröffentlicht: February 2024
In: Journal of applied clinical medical physics
Year: 2024, Jahrgang: 25, Heft: 2, Pages: 1-14
ISSN:1526-9914
DOI:10.1002/acm2.14174
Online-Zugang:Verlag, kostenfrei, Volltext: https://doi.org/10.1002/acm2.14174
Verlag, kostenfrei, Volltext: https://onlinelibrary.wiley.com/doi/abs/10.1002/acm2.14174
Volltext
Verfasserangaben:Abdallah Qubala, Jehad Shafee, Vania Batista, Jakob Liermann, Marcus Winter, Daniel Piro, Oliver Jäkel

MARC

LEADER 00000caa a2200000 c 4500
001 187191972X
003 DE-627
005 20240729084808.0
007 cr uuu---uuuuu
008 231205s2024 xx |||||o 00| ||eng c
024 7 |a 10.1002/acm2.14174  |2 doi 
035 |a (DE-627)187191972X 
035 |a (DE-599)KXP187191972X 
035 |a (OCoLC)1425209053 
040 |a DE-627  |b ger  |c DE-627  |e rda 
041 |a eng 
084 |a 33  |2 sdnb 
100 1 |a Qubala, Abdallah  |d 1990-  |e VerfasserIn  |0 (DE-588)1306941512  |0 (DE-627)1867272601  |4 aut 
245 1 0 |a Comparative evaluation of a surface-based respiratory monitoring system against a pressure sensor for 4DCT image reconstruction in phantoms  |c Abdallah Qubala, Jehad Shafee, Vania Batista, Jakob Liermann, Marcus Winter, Daniel Piro, Oliver Jäkel 
264 1 |c February 2024 
300 |b Illustrationen 
300 |a 14 
336 |a Text  |b txt  |2 rdacontent 
337 |a Computermedien  |b c  |2 rdamedia 
338 |a Online-Ressource  |b cr  |2 rdacarrier 
500 |a Online veröffentlicht: 10. Oktober 2023 
500 |a Gesehen am 05.12.2023 
520 |a Four-dimensional computed tomography (4DCT), which relies on breathing-induced motion, requires realistic surrogate information of breathing variations to reconstruct the tumor trajectory and motion variability of normal tissues accurately. Therefore, the SimRT surface-guided respiratory monitoring system has been installed on a Siemens CT scanner. This work evaluated the temporal and spatial accuracy of SimRT versus our commonly used pressure sensor, AZ-733 V. A dynamic thorax phantom was used to reproduce regular and irregular breathing patterns acquired by SimRT and Anzai. Various parameters of the recorded breathing patterns, including mean absolute deviations (MAD), Pearson correlations (PC), and tagging precision, were investigated and compared to ground-truth. Furthermore, 4DCT reconstructions were analyzed to assess the volume discrepancy, shape deformation and tumor trajectory. Compared to the ground-truth, SimRT more precisely reproduced the breathing patterns with a MAD range of 0.37 ± 0.27 and 0.92 ± 1.02 mm versus Anzai with 1.75 ± 1.54 and 5.85 ± 3.61 mm for regular and irregular breathing patterns, respectively. Additionally, SimRT provided a more robust PC of 0.994 ± 0.009 and 0.936 ± 0.062 for all investigated breathing patterns. Further, the peak and valley recognition were found to be more accurate and stable using SimRT. The comparison of tumor trajectories revealed discrepancies up to 7.2 and 2.3 mm for Anzai and SimRT, respectively. Moreover, volume discrepancies up to 1.71 ± 1.62% and 1.24 ± 2.02% were found for both Anzai and SimRT, respectively. SimRT was validated across various breathing patterns and showed a more precise and stable breathing tracking, (i) independent of the amplitude and period, (ii) and without placing any physical devices on the patient's body. These findings resulted in a more accurate temporal and spatial accuracy, thus leading to a more realistic 4DCT reconstruction and breathing-adapted treatment planning. 
650 4 |a 4DCT reconstruction 
650 4 |a breathing detection 
650 4 |a breathing surrogate 
650 4 |a commissioning 
650 4 |a mobile tumors 
650 4 |a motion artifacts 
650 4 |a respiratory monitoring system 
650 4 |a Surface-guided radiotherapy 
700 1 |a Shafee, Jehad  |e VerfasserIn  |0 (DE-588)1336887850  |0 (DE-627)1896512860  |4 aut 
700 1 |a Batista, Vania  |d 1985-  |e VerfasserIn  |0 (DE-588)1136686053  |0 (DE-627)893460656  |0 (DE-576)490784550  |4 aut 
700 1 |a Liermann, Jakob  |d 1989-  |e VerfasserIn  |0 (DE-588)1147779287  |0 (DE-627)1007181842  |0 (DE-576)496034170  |4 aut 
700 1 |a Winter, Marcus  |e VerfasserIn  |0 (DE-588)116515871X  |0 (DE-627)1029325561  |0 (DE-576)510276121  |4 aut 
700 1 |a Piro, Daniel  |e VerfasserIn  |4 aut 
700 1 |a Jäkel, Oliver  |d 1964-  |e VerfasserIn  |0 (DE-588)1050056302  |0 (DE-627)783260709  |0 (DE-576)404358705  |4 aut 
773 0 8 |i Enthalten in  |t Journal of applied clinical medical physics  |d Reston, Va. : ACMP, 2000  |g 25(2024), 2, Artikel-ID e14174, Seite 1-14  |h Online-Ressource  |w (DE-627)312202091  |w (DE-600)2010347-5  |w (DE-576)281192669  |x 1526-9914  |7 nnas  |a Comparative evaluation of a surface-based respiratory monitoring system against a pressure sensor for 4DCT image reconstruction in phantoms 
773 1 8 |g volume:25  |g year:2024  |g number:2  |g elocationid:e14174  |g pages:1-14  |g extent:14  |a Comparative evaluation of a surface-based respiratory monitoring system against a pressure sensor for 4DCT image reconstruction in phantoms 
856 4 0 |u https://doi.org/10.1002/acm2.14174  |x Verlag  |x Resolving-System  |z kostenfrei  |3 Volltext 
856 4 0 |u https://onlinelibrary.wiley.com/doi/abs/10.1002/acm2.14174  |x Verlag  |z kostenfrei  |3 Volltext 
951 |a AR 
992 |a 20231205 
993 |a Article 
994 |a 2024 
998 |g 1050056302  |a Jäkel, Oliver  |m 1050056302:Jäkel, Oliver  |d 50000  |e 50000PJ1050056302  |k 0/50000/  |p 7  |y j 
998 |g 116515871X  |a Winter, Marcus  |m 116515871X:Winter, Marcus  |d 910000  |d 950000  |d 950400  |e 910000PW116515871X  |e 950000PW116515871X  |e 950400PW116515871X  |k 0/910000/  |k 1/910000/950000/  |k 2/910000/950000/950400/  |p 5 
998 |g 1147779287  |a Liermann, Jakob  |m 1147779287:Liermann, Jakob  |d 910000  |d 911400  |d 50000  |e 910000PL1147779287  |e 911400PL1147779287  |e 50000PL1147779287  |k 0/910000/  |k 1/910000/911400/  |k 0/50000/  |p 4 
998 |g 1136686053  |a Batista, Vania  |m 1136686053:Batista, Vania  |d 50000  |e 50000PB1136686053  |k 0/50000/  |p 3 
998 |g 1336887850  |a Shafee, Jehad  |m 1336887850:Shafee, Jehad  |d 50000  |e 50000PS1336887850  |k 0/50000/  |p 2 
999 |a KXP-PPN187191972X  |e 4425363205 
BIB |a Y 
SER |a journal 
JSO |a {"name":{"displayForm":["Abdallah Qubala, Jehad Shafee, Vania Batista, Jakob Liermann, Marcus Winter, Daniel Piro, Oliver Jäkel"]},"id":{"doi":["10.1002/acm2.14174"],"eki":["187191972X"]},"origin":[{"dateIssuedDisp":"February 2024","dateIssuedKey":"2024"}],"relHost":[{"id":{"zdb":["2010347-5"],"eki":["312202091"],"issn":["1526-9914"]},"part":{"volume":"25","year":"2024","issue":"2","text":"25(2024), 2, Artikel-ID e14174, Seite 1-14","extent":"14","pages":"1-14"},"name":{"displayForm":["American College of Medical Physics"]},"pubHistory":["1.2000 -"],"title":[{"title_sort":"Journal of applied clinical medical physics","title":"Journal of applied clinical medical physics"}],"type":{"bibl":"periodical","media":"Online-Ressource"},"language":["eng"],"note":["Gesehen am 17. Januar 2017"],"origin":[{"dateIssuedKey":"2000","dateIssuedDisp":"2000-","publisherPlace":"Reston, Va.","publisher":"ACMP"}],"disp":"Comparative evaluation of a surface-based respiratory monitoring system against a pressure sensor for 4DCT image reconstruction in phantomsJournal of applied clinical medical physics","recId":"312202091","physDesc":[{"extent":"Online-Ressource"}]}],"person":[{"role":"aut","given":"Abdallah","family":"Qubala","display":"Qubala, Abdallah"},{"given":"Jehad","role":"aut","family":"Shafee","display":"Shafee, Jehad"},{"family":"Batista","given":"Vania","role":"aut","display":"Batista, Vania"},{"display":"Liermann, Jakob","given":"Jakob","role":"aut","family":"Liermann"},{"display":"Winter, Marcus","family":"Winter","role":"aut","given":"Marcus"},{"display":"Piro, Daniel","role":"aut","given":"Daniel","family":"Piro"},{"display":"Jäkel, Oliver","role":"aut","given":"Oliver","family":"Jäkel"}],"title":[{"title":"Comparative evaluation of a surface-based respiratory monitoring system against a pressure sensor for 4DCT image reconstruction in phantoms","title_sort":"Comparative evaluation of a surface-based respiratory monitoring system against a pressure sensor for 4DCT image reconstruction in phantoms"}],"type":{"media":"Online-Ressource","bibl":"article-journal"},"language":["eng"],"note":["Online veröffentlicht: 10. Oktober 2023","Gesehen am 05.12.2023"],"recId":"187191972X","physDesc":[{"extent":"14 S.","noteIll":"Illustrationen"}]} 
SRT |a QUBALAABDACOMPARATIV2024