Robust and versatile assembly for emitter positioning, observation, and heating in atmospheric pressure field desorption mass spectrometry

Atmospheric pressure field desorption (APFD) mass spectrometry (MS) has recently been introduced as a new variant of field desorption (FD) mass spectrometry. The development aimed at providing the basic characteristics of FD-MS in combination with instruments equipped with an atmospheric pressure (A...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Schweinfurth, Jan (VerfasserIn) , Linden, H. Bernhard (VerfasserIn) , Gross, Jürgen H. (VerfasserIn)
Dokumenttyp: Article (Journal)
Sprache:Englisch
Veröffentlicht: April 2024
In: European journal of mass spectrometry
Year: 2024, Jahrgang: 30, Heft: 2, Pages: 103-115
ISSN:1751-6838
DOI:10.1177/14690667241236073
Online-Zugang:Verlag, kostenfrei, Volltext: https://doi.org/10.1177/14690667241236073
Volltext
Verfasserangaben:Jan Schweinfurth, H. Bernhard Linden and Jürgen H. Gross

MARC

LEADER 00000caa a2200000 c 4500
001 1898323267
003 DE-627
005 20241205162018.0
007 cr uuu---uuuuu
008 240809s2024 xx |||||o 00| ||eng c
024 7 |a 10.1177/14690667241236073  |2 doi 
035 |a (DE-627)1898323267 
035 |a (DE-599)KXP1898323267 
035 |a (OCoLC)1475307358 
040 |a DE-627  |b ger  |c DE-627  |e rda 
041 |a eng 
084 |a 30  |2 sdnb 
100 1 |a Schweinfurth, Jan  |e VerfasserIn  |0 (DE-588)1338429930  |0 (DE-627)1898323623  |4 aut 
245 1 0 |a Robust and versatile assembly for emitter positioning, observation, and heating in atmospheric pressure field desorption mass spectrometry  |c Jan Schweinfurth, H. Bernhard Linden and Jürgen H. Gross 
264 1 |c April 2024 
300 |a 13 
336 |a Text  |b txt  |2 rdacontent 
337 |a Computermedien  |b c  |2 rdamedia 
338 |a Online-Ressource  |b cr  |2 rdacarrier 
500 |a Zuerst online veröffentlicht: 15. März 2024 
500 |a Gesehen am 09.08.2024 
520 |a Atmospheric pressure field desorption (APFD) mass spectrometry (MS) has recently been introduced as a new variant of field desorption (FD) mass spectrometry. The development aimed at providing the basic characteristics of FD-MS in combination with instruments equipped with an atmospheric pressure (AP) interface. Hitherto, APFD has been demonstrated to yield both positive and negative even electron ions of highly polar or ionic compounds as well as to enable the generation of positive molecular ions, M+•, of polycyclic aromatic compounds. The prototype setup for APFD was based on a nano-electrospray ionization (nanoESI) source slightly modified to allow for emitter positioning in front of the AP interface of a Fourier transform-ion cyclotron resonance (FT-ICR) mass spectrometer. The entrance electrode of the interface was set to negative or positive high voltage with respect to the emitter at ground potential, thereby permitting the formation of positive or negative ions, respectively. This work describes a custom-built device for quicker and more reproducible sample loading on and positioning of field emitters at the entrance electrode of the atmospheric pressure interface of a mass spectrometer. In addition, the device provides means for observation of the emitter during operation and for resistive emitter heating as employed in traditional FD-MS. Emitter heating both speeds up the desorption of the analytes and allows for the desorption/ionization of analytes of higher molecular weight than without emitter heating. In some cases, the signal-to-noise ratio of APFD mass spectra is improved due to higher ion currents effected by compressing the entire process into shorter periods of spectral acquisition. The new setup enables robust and reliable operation in APFD-MS. Moreover, it has been designed as to allow for use on a range of instruments as it can either be used on an FT-ICR mass spectrometer or in combination with a trapped ion mobility-quadrupole-time-of-flight (TIMS-Q-TOF) instrument. 
700 1 |a Linden, H. Bernhard  |e VerfasserIn  |0 (DE-588)1171542607  |0 (DE-627)1040630898  |0 (DE-576)513784713  |4 aut 
700 1 |a Gross, Jürgen H.  |d 1963-  |e VerfasserIn  |0 (DE-588)1028129734  |0 (DE-627)730387143  |0 (DE-576)375656391  |4 aut 
773 0 8 |i Enthalten in  |t European journal of mass spectrometry  |d London : Sage Publishing, 2000  |g 30(2024), 2 vom: Apr., Seite 103-115  |h Online-Ressource  |w (DE-627)320612368  |w (DE-600)2021540-X  |w (DE-576)335274781  |x 1751-6838  |7 nnas  |a Robust and versatile assembly for emitter positioning, observation, and heating in atmospheric pressure field desorption mass spectrometry 
773 1 8 |g volume:30  |g year:2024  |g number:2  |g month:04  |g pages:103-115  |g extent:13  |a Robust and versatile assembly for emitter positioning, observation, and heating in atmospheric pressure field desorption mass spectrometry 
856 4 0 |u https://doi.org/10.1177/14690667241236073  |x Verlag  |x Resolving-System  |z kostenfrei  |3 Volltext 
951 |a AR 
992 |a 20240809 
993 |a Article 
994 |a 2024 
998 |g 1028129734  |a Gross, Jürgen H.  |m 1028129734:Gross, Jürgen H.  |d 120000  |d 120100  |e 120000PG1028129734  |e 120100PG1028129734  |k 0/120000/  |k 1/120000/120100/  |p 3  |y j 
998 |g 1338429930  |a Schweinfurth, Jan  |m 1338429930:Schweinfurth, Jan  |d 120000  |d 120200  |e 120000PS1338429930  |e 120200PS1338429930  |k 0/120000/  |k 1/120000/120200/  |p 1  |x j 
999 |a KXP-PPN1898323267  |e 4564337998 
BIB |a Y 
SER |a journal 
JSO |a {"title":[{"title_sort":"Robust and versatile assembly for emitter positioning, observation, and heating in atmospheric pressure field desorption mass spectrometry","title":"Robust and versatile assembly for emitter positioning, observation, and heating in atmospheric pressure field desorption mass spectrometry"}],"person":[{"family":"Schweinfurth","given":"Jan","display":"Schweinfurth, Jan","roleDisplay":"VerfasserIn","role":"aut"},{"family":"Linden","given":"H. Bernhard","display":"Linden, H. Bernhard","roleDisplay":"VerfasserIn","role":"aut"},{"family":"Gross","given":"Jürgen H.","display":"Gross, Jürgen H.","roleDisplay":"VerfasserIn","role":"aut"}],"recId":"1898323267","language":["eng"],"note":["Zuerst online veröffentlicht: 15. März 2024","Gesehen am 09.08.2024"],"type":{"bibl":"article-journal","media":"Online-Ressource"},"id":{"eki":["1898323267"],"doi":["10.1177/14690667241236073"]},"origin":[{"dateIssuedKey":"2024","dateIssuedDisp":"April 2024"}],"name":{"displayForm":["Jan Schweinfurth, H. Bernhard Linden and Jürgen H. Gross"]},"relHost":[{"id":{"issn":["1751-6838"],"zdb":["2021540-X"],"eki":["320612368"]},"origin":[{"dateIssuedDisp":"2000-","publisher":"Sage Publishing ; IM Publications","dateIssuedKey":"2000","publisherPlace":"London ; Chichester"}],"name":{"displayForm":["published in association with British Mass Spectrometry Society"]},"physDesc":[{"extent":"Online-Ressource"}],"title":[{"title_sort":"European journal of mass spectrometry","subtitle":"EJMS","title":"European journal of mass spectrometry"}],"part":{"volume":"30","text":"30(2024), 2 vom: Apr., Seite 103-115","extent":"13","year":"2024","pages":"103-115","issue":"2"},"titleAlt":[{"title":"EJMS"}],"pubHistory":["Volume 6, issue 1 (February 2000)-"],"language":["eng"],"recId":"320612368","disp":"Robust and versatile assembly for emitter positioning, observation, and heating in atmospheric pressure field desorption mass spectrometryEuropean journal of mass spectrometry","type":{"bibl":"periodical","media":"Online-Ressource"},"note":["Gesehen am 10.06.2022"]}],"physDesc":[{"extent":"13 S."}]} 
SRT |a SCHWEINFURROBUSTANDV2024