ObjectiveThe objective of this study was to compare different modes of amplitude-integrated EEG (aEEG) assessment (semi-manual vs. automated) in children.MethodsA total of 450 unremarkable pediatric EEGs from children aged 6 months to 17.9 years were converted into aEEGs and the medians and means of...
Gespeichert in:
| Hauptverfasser: | , , , , , , |
|---|---|
| Dokumenttyp: | Article (Journal) |
| Sprache: | Englisch |
| Veröffentlicht: |
08 October 2025
|
| In: |
Frontiers in neurology
Year: 2025, Jahrgang: 16, Pages: 1-8 |
| ISSN: | 1664-2295 |
| DOI: | 10.3389/fneur.2025.1562580 |
| Online-Zugang: | Verlag, lizenzpflichtig, Volltext: https://doi.org/10.3389/fneur.2025.1562580 Verlag, lizenzpflichtig, Volltext: https://www.frontiersin.org/journals/neurology/articles/10.3389/fneur.2025.1562580/full |
| Verfasserangaben: | Sandra Greve, Pia Brensing, Luisa Paul, Christian Dohna-Schwake, Julia Wichlacz, Ursula Felderhoff-Mueser and Nora Bruns |
MARC
| LEADER | 00000naa a2200000 c 4500 | ||
|---|---|---|---|
| 001 | 1950089509 | ||
| 003 | DE-627 | ||
| 005 | 20260123131647.0 | ||
| 007 | cr uuu---uuuuu | ||
| 008 | 260123s2025 xx |||||o 00| ||eng c | ||
| 024 | 7 | |a 10.3389/fneur.2025.1562580 |2 doi | |
| 035 | |a (DE-627)1950089509 | ||
| 035 | |a (DE-599)KXP1950089509 | ||
| 040 | |a DE-627 |b ger |c DE-627 |e rda | ||
| 041 | |a eng | ||
| 084 | |a 33 |2 sdnb | ||
| 100 | 1 | |a Greve, Sandra Julia Nicole |d 1989- |e VerfasserIn |0 (DE-588)1247583589 |0 (DE-627)1782124063 |4 aut | |
| 245 | 1 | 0 | |a Systematically lower aEEG amplitude values in neurologically healthy children using an automated algorithm compared to a semi-manual method |c Sandra Greve, Pia Brensing, Luisa Paul, Christian Dohna-Schwake, Julia Wichlacz, Ursula Felderhoff-Mueser and Nora Bruns |
| 264 | 1 | |c 08 October 2025 | |
| 300 | |b Illustrationen | ||
| 300 | |a 8 | ||
| 336 | |a Text |b txt |2 rdacontent | ||
| 337 | |a Computermedien |b c |2 rdamedia | ||
| 338 | |a Online-Ressource |b cr |2 rdacarrier | ||
| 500 | |a Gesehen am 23.01.2026 | ||
| 520 | |a ObjectiveThe objective of this study was to compare different modes of amplitude-integrated EEG (aEEG) assessment (semi-manual vs. automated) in children.MethodsA total of 450 unremarkable pediatric EEGs from children aged 6 months to 17.9 years were converted into aEEGs and the medians and means of the upper and lower amplitudes (C3-P3, C4-P4, C3-C4, P3-P4, Fp1-Fp2) were determined. The agreement of the semi-manual and automated measurements was assessed via the Pearson correlation coefficients (PCC) and Bland-Altman plots. Mean differences between the methods and age-specific percentiles (5th−95th) were calculated.ResultsSemi-manually measured amplitudes were systematically greater than automated assessments. Mean differences of the means ranged between 23.7 and 29.3 μV for the upper and between 2.4 and 4.4 μV for the lower amplitudes depending on the channel. The PCC ranged between 0.68 and 0.92 for the upper and lower amplitudes of the mean depending on the channel. Age-specific percentiles showed different absolute values but similar trends.ConclusionAEEG amplitude values systematically differ between semi-manual and automated assessment. Age-related trends are evident despite differences in the absolute values. Reference values for different measurement techniques are needed for pediatric aEEG. | ||
| 650 | 4 | |a amplitude-integrated EEG | |
| 650 | 4 | |a children | |
| 650 | 4 | |a infants | |
| 650 | 4 | |a neuromonitoring | |
| 650 | 4 | |a pediatric intensive care | |
| 650 | 4 | |a reference values | |
| 700 | 1 | |a Brensing, Pia Sophie |e VerfasserIn |0 (DE-588)1260418766 |0 (DE-627)1807288315 |4 aut | |
| 700 | 1 | |a Paul, Luisa |d 1990- |e VerfasserIn |0 (DE-588)1196231990 |0 (DE-627)1678054984 |4 aut | |
| 700 | 1 | |a Dohna-Schwake, Christian |e VerfasserIn |0 (DE-588)1033312193 |0 (DE-627)740853341 |0 (DE-576)380895366 |4 aut | |
| 700 | 1 | |a Wichlacz, Julia |e VerfasserIn |4 aut | |
| 700 | 1 | |a Felderhoff-Mueser, Ursula |e VerfasserIn |4 aut | |
| 700 | 1 | |a Bruns, Nora |d 1984- |e VerfasserIn |0 (DE-588)1166334090 |0 (DE-627)1030374090 |0 (DE-576)510723071 |4 aut | |
| 773 | 0 | 8 | |i Enthalten in |t Frontiers in neurology |d Lausanne : Frontiers Research Foundation, 2008 |g 16(2025), Artikel-ID 1562580, Seite 1-8 |h Online-Ressource |w (DE-627)631498753 |w (DE-600)2564214-5 |w (DE-576)326550933 |x 1664-2295 |7 nnas |a Systematically lower aEEG amplitude values in neurologically healthy children using an automated algorithm compared to a semi-manual method |
| 773 | 1 | 8 | |g volume:16 |g year:2025 |g elocationid:1562580 |g pages:1-8 |g extent:8 |a Systematically lower aEEG amplitude values in neurologically healthy children using an automated algorithm compared to a semi-manual method |
| 856 | 4 | 0 | |u https://doi.org/10.3389/fneur.2025.1562580 |x Verlag |x Resolving-System |z lizenzpflichtig |3 Volltext |7 1 |
| 856 | 4 | 0 | |u https://www.frontiersin.org/journals/neurology/articles/10.3389/fneur.2025.1562580/full |x Verlag |z lizenzpflichtig |3 Volltext |7 1 |
| 951 | |a AR | ||
| 992 | |a 20260123 | ||
| 993 | |a Article | ||
| 994 | |a 2025 | ||
| 998 | |g 1196231990 |a Paul, Luisa |m 1196231990:Paul, Luisa |d 910000 |d 910500 |e 910000PP1196231990 |e 910500PP1196231990 |k 0/910000/ |k 1/910000/910500/ |p 3 | ||
| 999 | |a KXP-PPN1950089509 |e 4858601382 | ||
| BIB | |a Y | ||
| SER | |a journal | ||
| JSO | |a {"origin":[{"dateIssuedKey":"2025","dateIssuedDisp":"08 October 2025"}],"id":{"doi":["10.3389/fneur.2025.1562580"],"eki":["1950089509"]},"person":[{"family":"Greve","display":"Greve, Sandra Julia Nicole","role":"aut","given":"Sandra Julia Nicole"},{"display":"Brensing, Pia Sophie","role":"aut","family":"Brensing","given":"Pia Sophie"},{"given":"Luisa","role":"aut","display":"Paul, Luisa","family":"Paul"},{"family":"Dohna-Schwake","role":"aut","display":"Dohna-Schwake, Christian","given":"Christian"},{"family":"Wichlacz","role":"aut","display":"Wichlacz, Julia","given":"Julia"},{"given":"Ursula","family":"Felderhoff-Mueser","role":"aut","display":"Felderhoff-Mueser, Ursula"},{"display":"Bruns, Nora","role":"aut","family":"Bruns","given":"Nora"}],"type":{"media":"Online-Ressource","bibl":"article-journal"},"title":[{"title":"Systematically lower aEEG amplitude values in neurologically healthy children using an automated algorithm compared to a semi-manual method","title_sort":"Systematically lower aEEG amplitude values in neurologically healthy children using an automated algorithm compared to a semi-manual method"}],"name":{"displayForm":["Sandra Greve, Pia Brensing, Luisa Paul, Christian Dohna-Schwake, Julia Wichlacz, Ursula Felderhoff-Mueser and Nora Bruns"]},"physDesc":[{"noteIll":"Illustrationen","extent":"8 S."}],"recId":"1950089509","relHost":[{"note":["Gesehen am 18.08.20","Specialty sections: Autonomic Neuroscience; Brain Imaging Methods; Dementia; Endovascular and Interventional Neurology; Epilepsy; Headache Medicine and Facial Pain; Movement Disorders; Multiple Sclerosis and Neuroimmunology; Neurocritical and Neurohospitalist; Care; Neurodegeneration; Neurogenomics; Neurology Education; Neuromuscular Diseases; Neuro-Oncology; Neuro-Ophthalmology; Neuro-Otology; Neuropediatrics; Neuropharmacology; Neuroprosthetics; Neurotrauma; Sleep and Chronobiology; Spinal Cord Medicine; Sports Neurology; Stroke; Teleneurology"],"recId":"631498753","physDesc":[{"extent":"Online-Ressource"}],"type":{"media":"Online-Ressource","bibl":"periodical"},"titleAlt":[{"title":"Autonomic Neuroscience"},{"title":"Brain Imaging Methods"},{"title":"Dementia"},{"title":"Endovascular and Interventional Neurology"},{"title":"Epilepsy"},{"title":"Headache Medicine and Facial Pain"},{"title":"Movement Disorders"},{"title":"Multiple Sclerosis and Neuroimmunology"},{"title":"Neurocritical and Neurohospitalist Care"},{"title":"Neurodegeneration"},{"title":"Neurogenomics"},{"title":"Neurology Education"},{"title":"Neuromuscular Diseases"},{"title":"Neuro-Oncology"},{"title":"Neuro-Ophthalmology"},{"title":"Neuro-Otology"},{"title":"Neuropediatrics"},{"title":"Neuropharmacology"},{"title":"Neuroprosthetics"},{"title":"Neurotrauma"},{"title":"Sleep and Chronobiology"},{"title":"Spinal Cord Medicine"},{"title":"Sports Neurology"},{"title":"Stroke"},{"title":"Teleneurology"}],"pubHistory":["2008 -"],"language":["eng"],"title":[{"title":"Frontiers in neurology","title_sort":"Frontiers in neurology"}],"disp":"Systematically lower aEEG amplitude values in neurologically healthy children using an automated algorithm compared to a semi-manual methodFrontiers in neurology","name":{"displayForm":["Frontiers Research Foundation"]},"origin":[{"dateIssuedKey":"2008","publisherPlace":"Lausanne","publisher":"Frontiers Research Foundation","dateIssuedDisp":"2008-"}],"id":{"eki":["631498753"],"issn":["1664-2295"],"zdb":["2564214-5"]},"part":{"year":"2025","text":"16(2025), Artikel-ID 1562580, Seite 1-8","extent":"8","pages":"1-8","volume":"16"}}],"note":["Gesehen am 23.01.2026"],"language":["eng"]} | ||
| SRT | |a GREVESANDRSYSTEMATIC0820 | ||