Development of the trigger-controlling system for the proton computed tomography prototype

Proton computed tomography (pCT) is an imaging reconstruction technique that uses protons’ positions, directions, and energy loss to generate a map of an object’s relative stopping power. In this work, the trigger-controlling system was designed and constructed to synchronize between components of o...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Ritjoho, Narongrit (VerfasserIn) , Phumara, P. (VerfasserIn) , Buanill, Y. (VerfasserIn) , Boontueng, Phongnared (VerfasserIn) , Songmoolnak, Arnon (VerfasserIn) , Laojamnongwong, Natthawut (VerfasserIn) , Charoonratana, L. (VerfasserIn) , Sanghangthum, Taweap (VerfasserIn) , Röhrich, Dieter (VerfasserIn) , Schilling, Alexander (VerfasserIn) , Aehle, Markus (VerfasserIn) , Alme, Johannes (VerfasserIn) , Barnaföldi, Gergely Gábor (VerfasserIn) , Bíró, Gábor (VerfasserIn) , Bodova, Tea (VerfasserIn) , Borshchov, Vyacheslav (VerfasserIn) , van den Brink, Anthony (VerfasserIn) , Eikeland, Viljar (VerfasserIn) , Feofilov, Grigory (VerfasserIn) , Garth, Clint (VerfasserIn) , Gauger, Nicolas R. (VerfasserIn) , Grøttvik, Ola (VerfasserIn) , Helstrup, Havard (VerfasserIn) , Igolkin, Sergey (VerfasserIn) , Keidel, Ralf (VerfasserIn) , Kortus, Tobias (VerfasserIn) , Kusch, Lisa (VerfasserIn) , Leonhardt, Volker (VerfasserIn) , Mehendale, Shruti (VerfasserIn) , Mulawade, Raju Ningappa (VerfasserIn) , Odland, Odd Harald (VerfasserIn) , O’Neill, George (VerfasserIn) , Papp, Gabriella. (VerfasserIn) , Peitzmann, Thomas (VerfasserIn) , Pettersen, Helge Egil Seime (VerfasserIn) , Piersimoni, Pierluigi (VerfasserIn) , Pochampalli, Rohit (VerfasserIn) , Protsenko, Maksym (VerfasserIn) , Rauch, Max Philip (VerfasserIn) , Rehman, A. (VerfasserIn) , Richter, M. (VerfasserIn) , Sagebaum, M. (VerfasserIn) , Santana, Jonas Machado (VerfasserIn) , Seco, Joao (VerfasserIn) , Tambave, Ganesh (VerfasserIn) , Tymchuk, Ihor (VerfasserIn) , Ullaland, Kjetil (VerfasserIn) , Varga-Kőfaragó, Monika (VerfasserIn) , Volz, Lauren (VerfasserIn) , Wagner, B. (VerfasserIn) , Wendzel, Steffen (VerfasserIn) , Wiebel, Arnold (VerfasserIn) , Xiao, Renzheng (VerfasserIn) , Yang, S. (VerfasserIn) , Zillien, S. (VerfasserIn) , Kobdaj, Chinorat (VerfasserIn)
Dokumenttyp: Article (Journal)
Sprache:Englisch
Veröffentlicht: 2025
In: Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment
Year: 2025, Jahrgang: 1072, Pages: 1-9
ISSN:1872-9576
DOI:10.1016/j.nima.2024.170125
Online-Zugang:Verlag, lizenzpflichtig, Volltext: https://doi.org/10.1016/j.nima.2024.170125
Verlag, lizenzpflichtig, Volltext: https://www.sciencedirect.com/science/article/pii/S0168900224010519
Volltext
Verfasserangaben:N. Ritjoho, P. Phumara, Y. Buanill, P. Boontueng, A. Songmoolnak, N. Laojamnongwong, L. Charoonratana, T. Sanghangthum, D. Röhrich, A. Schilling, M. Aehle, J. Alme, G. G. Barnaföldi, G. Bíró, T. Bodova, V. Borshchov, A. van den Brink, V. Eikeland, G. Feofilov, C. Garth, N. R. Gauger, O. Grøttvik, H. Helstrup, S. Igolkin, R. Keidel, T. Kortus, L. Kusch, V. Leonhardt, S. Mehendale, R. N. Mulawade, O. H. Odland, G. O’Neill, G. Papp, T. Peitzmann, H. E. S. Pettersen, P. Piersimoni, R. Pochampalli, M. Protsenko, M. Rauch, A. U. Rehman, M. Richter, M. Sagebaum, J. Santana, J. Seco, G. Tambave, I. Tymchuk, K. Ullaland, M. Varga-Kőfaragó, L. Volz, B. Wagner, S. Wendzel, A. Wiebel, R. Xiao, S. Yang, S. Zillien, C. Kobdaj
Beschreibung
Zusammenfassung:Proton computed tomography (pCT) is an imaging reconstruction technique that uses protons’ positions, directions, and energy loss to generate a map of an object’s relative stopping power. In this work, the trigger-controlling system was designed and constructed to synchronize between components of our proposed pCT prototype and a proton beam at King Chulalongkorn Memorial Hospital (KCMH) in Bangkok. The controlling system can control the proton beam gating, a rotational stage, and ALPIDE sensors, which are parts of a tracker and a proton calorimeter of the pCT prototype. The controlling system was constructed based on the MEGA2560 microcontroller unit. The controlling system was verified by measuring the activated rows of the ALPIDE sensor at different proton extraction times. Additionally, the controlling system can control an array of six ALPIDE sensors, which could provide preliminary results of the proton calorimeter prototype.
Beschreibung:Gesehen am 26.01.2026
Online veröffentlicht: 10. Dezember 2024
Beschreibung:Online Resource
ISSN:1872-9576
DOI:10.1016/j.nima.2024.170125