Evaluation of 2D ion chamber arrays for patient specific quality assurance using a static phantom at a 0.35T MR-Linac
Introduction - Patient specific quality assurance (QA) in MR-Linacs can be performed with MR-compatible ion chamber arrays. However, the presence of a static magnetic field can alter the angular response of such arrays substantially. This works investigates the suitability of two ion chamber arrays,...
Gespeichert in:
| Hauptverfasser: | , , , , , , , |
|---|---|
| Dokumenttyp: | Article (Journal) |
| Sprache: | Englisch |
| Veröffentlicht: |
August 2025
|
| In: |
Zeitschrift für medizinische Physik
Year: 2025, Jahrgang: 35, Heft: 3, Pages: 375-384 |
| ISSN: | 1876-4436 |
| DOI: | 10.1016/j.zemedi.2023.12.003 |
| Online-Zugang: | Verlag, kostenfrei, Volltext: https://doi.org/10.1016/j.zemedi.2023.12.003 Verlag, kostenfrei, Volltext: https://www.sciencedirect.com/science/article/pii/S0939388923001502 |
| Verfasserangaben: | C. Katharina Renkamp, Daniela Eulenstein, Moritz Sebald, Fabian Schlüter, Carolin Buchele, Carolin Rippke, Jürgen P. Debus, Sebastian Klüter |
MARC
| LEADER | 00000naa a2200000 c 4500 | ||
|---|---|---|---|
| 001 | 1950430162 | ||
| 003 | DE-627 | ||
| 005 | 20260128101505.0 | ||
| 007 | cr uuu---uuuuu | ||
| 008 | 260128s2025 xx |||||o 00| ||eng c | ||
| 024 | 7 | |a 10.1016/j.zemedi.2023.12.003 |2 doi | |
| 035 | |a (DE-627)1950430162 | ||
| 035 | |a (DE-599)KXP1950430162 | ||
| 040 | |a DE-627 |b ger |c DE-627 |e rda | ||
| 041 | |a eng | ||
| 084 | |a 33 |2 sdnb | ||
| 100 | 1 | |a Renkamp, Claudia Katharina |d 1990- |e VerfasserIn |0 (DE-588)1153576384 |0 (DE-627)101508625X |0 (DE-576)500325286 |4 aut | |
| 245 | 1 | 0 | |a Evaluation of 2D ion chamber arrays for patient specific quality assurance using a static phantom at a 0.35T MR-Linac |c C. Katharina Renkamp, Daniela Eulenstein, Moritz Sebald, Fabian Schlüter, Carolin Buchele, Carolin Rippke, Jürgen P. Debus, Sebastian Klüter |
| 246 | 3 | 0 | |a two |
| 264 | 1 | |c August 2025 | |
| 300 | |b Illustrationen | ||
| 300 | |a 10 | ||
| 336 | |a Text |b txt |2 rdacontent | ||
| 337 | |a Computermedien |b c |2 rdamedia | ||
| 338 | |a Online-Ressource |b cr |2 rdacarrier | ||
| 500 | |a Online verfügbar: 5. Januar 2024, Artikelversion: 15. November 2025 | ||
| 500 | |a Gesehen am 28.01.2026 | ||
| 520 | |a Introduction - Patient specific quality assurance (QA) in MR-Linacs can be performed with MR-compatible ion chamber arrays. However, the presence of a static magnetic field can alter the angular response of such arrays substantially. This works investigates the suitability of two ion chamber arrays, an air-filled and a liquid-filled array, for patient specific QA at a 0.35T MR-Linac using a static phantom. - Methods - In order to study the angular response, the two arrays were placed in a static, solid phantom and irradiated with 9.96×9.96cm2 fields every 10° beam angle at a 0.35T MR-Linac. Measurements were compared to the TPS calculated dose in terms of gamma passing rate and relative dose to the central chamber. 20 patient specific quality assurance plans were measured using the liquid-filled array. - Results - The air-filled array showed asymmetric angular response changes of central chamber dose of up to 18% and down to local 3mm/3% gamma rates of 20%, while only minor differences within 3% (excluding parallel irradiation and beams through the couch edges) were found for the liquid-filled ion chamber array without rotating the phantom. Patient plan QA using the liquid-filled array yielded a median local 3mm/3% 3D gamma passing rate of 99.8% (range 96.9%-100%). - Conclusion - A liquid-filled ionization chamber array in combination with a static phantom can be used for efficient patient specific plan QA in a single measurement set-up in a 0.35T MR-Linac, while the air-filled ion chamber array phantom shows large angular response changes and has its limitations regarding patient specific QA measurements. | ||
| 650 | 4 | |a ion chamber array | |
| 650 | 4 | |a MR-Linac | |
| 650 | 4 | |a patient specific quality assurance | |
| 700 | 1 | |a Eulenstein, Daniela |e VerfasserIn |4 aut | |
| 700 | 1 | |a Sebald, Moritz |e VerfasserIn |0 (DE-588)1388203855 |0 (DE-627)1950430510 |4 aut | |
| 700 | 1 | |a Schlüter, Fabian |e VerfasserIn |0 (DE-588)1296060675 |0 (DE-627)1852746041 |4 aut | |
| 700 | 1 | |a Buchele, Carolin |d 1992- |e VerfasserIn |0 (DE-588)1238773230 |0 (DE-627)1766395635 |4 aut | |
| 700 | 1 | |a Rippke, Carolin |d 1993- |e VerfasserIn |0 (DE-588)1238772773 |0 (DE-627)1766392539 |4 aut | |
| 700 | 1 | |a Debus, Jürgen |d 1964- |e VerfasserIn |0 (DE-588)1022671421 |0 (DE-627)717025780 |0 (DE-576)365774944 |4 aut | |
| 700 | 1 | |a Klüter, Sebastian |d 1980- |e VerfasserIn |0 (DE-588)1071160796 |0 (DE-627)825622646 |0 (DE-576)311255248 |4 aut | |
| 773 | 0 | 8 | |i Enthalten in |t Zeitschrift für medizinische Physik |d Amsterdam [u.a.] : Elsevier, 1990 |g 35(2025), 3 vom: Aug., Seite 375-384 |w (DE-627)510617662 |w (DE-600)2231492-1 |w (DE-576)272350818 |x 1876-4436 |7 nnas |a Evaluation of 2D ion chamber arrays for patient specific quality assurance using a static phantom at a 0.35T MR-Linac |
| 773 | 1 | 8 | |g volume:35 |g year:2025 |g number:3 |g month:08 |g pages:375-384 |g extent:10 |a Evaluation of 2D ion chamber arrays for patient specific quality assurance using a static phantom at a 0.35T MR-Linac |
| 856 | 4 | 0 | |u https://doi.org/10.1016/j.zemedi.2023.12.003 |x Verlag |x Resolving-System |z kostenfrei |3 Volltext |7 0 |
| 856 | 4 | 0 | |u https://www.sciencedirect.com/science/article/pii/S0939388923001502 |x Verlag |z kostenfrei |3 Volltext |7 0 |
| 951 | |a AR | ||
| 992 | |a 20260128 | ||
| 993 | |a Article | ||
| 994 | |a 2025 | ||
| 998 | |g 1071160796 |a Klüter, Sebastian |m 1071160796:Klüter, Sebastian |d 910000 |d 911400 |e 910000PK1071160796 |e 911400PK1071160796 |k 0/910000/ |k 1/910000/911400/ |p 8 |y j | ||
| 998 | |g 1022671421 |a Debus, Jürgen |m 1022671421:Debus, Jürgen |d 910000 |d 911400 |e 910000PD1022671421 |e 911400PD1022671421 |k 0/910000/ |k 1/910000/911400/ |p 7 | ||
| 998 | |g 1238772773 |a Rippke, Carolin |m 1238772773:Rippke, Carolin |d 910000 |d 911400 |e 910000PR1238772773 |e 911400PR1238772773 |k 0/910000/ |k 1/910000/911400/ |p 6 | ||
| 998 | |g 1238773230 |a Buchele, Carolin |m 1238773230:Buchele, Carolin |p 5 | ||
| 998 | |g 1296060675 |a Schlüter, Fabian |m 1296060675:Schlüter, Fabian |d 910000 |d 911400 |e 910000PS1296060675 |e 911400PS1296060675 |k 0/910000/ |k 1/910000/911400/ |p 4 | ||
| 998 | |g 1388203855 |a Sebald, Moritz |m 1388203855:Sebald, Moritz |p 3 | ||
| 998 | |g 1153576384 |a Renkamp, Claudia Katharina |m 1153576384:Renkamp, Claudia Katharina |d 910000 |d 911400 |e 910000PR1153576384 |e 911400PR1153576384 |k 0/910000/ |k 1/910000/911400/ |p 1 |x j | ||
| 999 | |a KXP-PPN1950430162 |e 4863024576 | ||
| BIB | |a Y | ||
| SER | |a journal | ||
| JSO | |a {"physDesc":[{"extent":"10 S.","noteIll":"Illustrationen"}],"recId":"1950430162","note":["Online verfügbar: 5. Januar 2024, Artikelversion: 15. November 2025","Gesehen am 28.01.2026"],"type":{"bibl":"article-journal","media":"Online-Ressource"},"language":["eng"],"title":[{"title":"Evaluation of 2D ion chamber arrays for patient specific quality assurance using a static phantom at a 0.35T MR-Linac","title_sort":"Evaluation of 2D ion chamber arrays for patient specific quality assurance using a static phantom at a 0.35T MR-Linac"}],"relHost":[{"corporate":[{"role":"isb","display":"Deutsche Gesellschaft für Medizinische Physik"},{"display":"Österreichische Gesellschaft für Medizinische Physik","role":"isb"}],"pubHistory":["Volume 1, supplement (1990) ; volume 1, issue 1 (1991)-"],"titleAlt":[{"title":"Medizinische Physik"}],"part":{"volume":"35","year":"2025","text":"35(2025), 3 vom: Aug., Seite 375-384","extent":"10","issue":"3","pages":"375-384"},"id":{"issn":["1876-4436"],"eki":["510617662"],"zdb":["2231492-1"]},"origin":[{"dateIssuedKey":"1990","publisherPlace":"Amsterdam [u.a.]","dateIssuedDisp":"1990-","publisher":"Elsevier"}],"note":["Gesehen am 31.01.2022"],"type":{"bibl":"periodical","media":"Online-Ressource"},"language":["ger"],"title":[{"title_sort":"Zeitschrift für medizinische Physik","title":"Zeitschrift für medizinische Physik","subtitle":"offizielles Organ der Deutschen, Österreichischen und Schweizerischen Gesellschaft für Medizinische Physik"}],"recId":"510617662","disp":"Evaluation of 2D ion chamber arrays for patient specific quality assurance using a static phantom at a 0.35T MR-LinacZeitschrift für medizinische Physik"}],"person":[{"display":"Renkamp, Claudia Katharina","family":"Renkamp","role":"aut","given":"Claudia Katharina"},{"display":"Eulenstein, Daniela","family":"Eulenstein","role":"aut","given":"Daniela"},{"family":"Sebald","role":"aut","given":"Moritz","display":"Sebald, Moritz"},{"display":"Schlüter, Fabian","given":"Fabian","role":"aut","family":"Schlüter"},{"family":"Buchele","role":"aut","given":"Carolin","display":"Buchele, Carolin"},{"display":"Rippke, Carolin","family":"Rippke","given":"Carolin","role":"aut"},{"given":"Jürgen","role":"aut","family":"Debus","display":"Debus, Jürgen"},{"given":"Sebastian","role":"aut","family":"Klüter","display":"Klüter, Sebastian"}],"origin":[{"dateIssuedKey":"2025","dateIssuedDisp":"August 2025"}],"id":{"eki":["1950430162"],"doi":["10.1016/j.zemedi.2023.12.003"]},"name":{"displayForm":["C. Katharina Renkamp, Daniela Eulenstein, Moritz Sebald, Fabian Schlüter, Carolin Buchele, Carolin Rippke, Jürgen P. Debus, Sebastian Klüter"]}} | ||
| SRT | |a RENKAMPCLAEVALUATION2025 | ||