Quality assurance program on stereotactic radiosurgery: report from a Quality Assurance Task Group

Radiosurgery performed with the "Gamma Knife" or on a linac facility is a specific radiotherapy method to treat certain lesions in the brain with a high single radiation dose. Geometric miss of a radiosurgica target is considered the most likely serious error in a stereotactic radiosurgery...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
1. Verfasser: Hartmann, Günther H. (VerfasserIn)
Dokumenttyp: Buch/Monographie
Sprache:Englisch
Veröffentlicht: Berlin, Heidelberg Springer Berlin Heidelberg 01 July 1995
DOI:10.1007/978-3-642-79718-7
Online-Zugang:Resolving-System, lizenzpflichtig, Volltext: http://dx.doi.org/10.1007/978-3-642-79718-7
Volltext
Verfasserangaben:by Günther H. Hartmann
Beschreibung
Zusammenfassung:Radiosurgery performed with the "Gamma Knife" or on a linac facility is a specific radiotherapy method to treat certain lesions in the brain with a high single radiation dose. Geometric miss of a radiosurgica target is considered the most likely serious error in a stereotactic radiosurgery treaatment. Therefore, the issue of quality assurance in stereotactic radiosurgery is a serious one for neurosurgeons, physicists, patients and those of us who wish to analyze the results of radiosurgery. The purpose of this book is to offer and describe a quality assurance program and procedures of the physical and technical aspects through identifying the sources of possible errors and suggesting quality assurance tests that should be performed in order for a stereotactic system to be used clinically
Beschreibung:Elektronische Reproduktion der Druck-Ausgabe 6. Dezember 2012
Beschreibung:Online Resource
ISBN:9783642797187
DOI:10.1007/978-3-642-79718-7