Technical note: intrinsic raw data-based CT misalignment correction without redundant data
Purpose CT image reconstruction requires accurate knowledge of the used geometry or image quality might be degraded by misalignment artifacts. To overcome this issue, an intrinsic method, that is, a method not requiring a dedicated calibration phantom, to perform a raw data-based misalignment correc...
Saved in:
| Main Authors: | , , , , |
|---|---|
| Format: | Article (Journal) |
| Language: | English |
| Published: |
2019
|
| In: |
Medical physics
Year: 2018, Volume: 46, Issue: 1, Pages: 173-179 |
| ISSN: | 2473-4209 |
| DOI: | 10.1002/mp.13254 |
| Online Access: | Verlag, Volltext: https://doi.org/10.1002/mp.13254 Verlag, Volltext: https://aapm.onlinelibrary.wiley.com/doi/abs/10.1002/mp.13254 |
| Author Notes: | Stefan Sawall, Andreas Hahn, Joscha Maier, Jan Kuntz, Marc Kachelrieß |
| Summary: | Purpose CT image reconstruction requires accurate knowledge of the used geometry or image quality might be degraded by misalignment artifacts. To overcome this issue, an intrinsic method, that is, a method not requiring a dedicated calibration phantom, to perform a raw data-based misalignment correction for CT is proposed herein that does not require redundant data and hence is applicable to measurements with less than plus fan-angle of data. Methods The forward projection of a volume reconstructed from a misaligned geometry resembles the acquired raw data if no redundant data are used, that is, if less than plus fan-angle are used for image reconstruction. Hence, geometric parameters cannot be deduced from such data by an optimization of the geometry-dependent raw data fidelity. We propose to use a nonlinear transform applied to the reconstructed volume to introduce inconsistencies in the raw data that can be employed to estimate geometric parameters using less than plus fan-angle of data. The proposed method is evaluated using simulations of the FORBILD head phantom and using actual measurements of a contrast-enhanced scan of a mouse acquired using a micro-CT. Results Noisy simulations and actual measurements demonstrate that the proposed method is capable of correcting for artifacts arising from a misaligned geometry without redundant data while ensuring raw data fidelity. Conclusions The proposed method extends intrinsic raw data-based misalignment correction methods to an angular range of or less and is thus applicable to systems with a limited scan range. |
|---|---|
| Item Description: | published 16 November 2018 Gesehen am 23.05.2019 |
| Physical Description: | Online Resource |
| ISSN: | 2473-4209 |
| DOI: | 10.1002/mp.13254 |