Local application of Hyperthermia in the esophagus with a heatable malleable thermoplastic stent

Objective: This study measures the effect of hyperthermia applied through a heatable stent in the esophagus in order to investigate whether this procedure offers a therapeutic option for tumor treatment. Material and Methods: Thermoplastic malleable stents, with the capacity to be heated after impla...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Freudenberg, Sebastian (Verfasst von) , Rewerk, Stephan (Verfasst von) , Wagner, Achim (Verfasst von) , Gebhard, Martha-Maria (Verfasst von) , Kähler, Georg (Verfasst von)
Dokumenttyp: Article (Journal)
Sprache:Englisch
Veröffentlicht: February 14, 2006
In: European surgical research
Year: 2006, Jahrgang: 38, Heft: 1, Pages: 42-47
ISSN:1421-9921
DOI:10.1159/000091526
Online-Zugang:Verlag, Volltext: https://doi.org/10.1159/000091526
Verlag: https://www.karger.com/Article/FullText/91526
Volltext
Verfasserangaben:S. Freudenberg, S. Rewerk, Fatma Bay, Christiane Al Khouri, A. Wagner, M. Isaac, Martha Maria Gebhard, G. Kähler
Beschreibung
Zusammenfassung:Objective: This study measures the effect of hyperthermia applied through a heatable stent in the esophagus in order to investigate whether this procedure offers a therapeutic option for tumor treatment. Material and Methods: Thermoplastic malleable stents, with the capacity to be heated after implantation, were placed endoscopically in the middle third of the esophagi of 30 pigs. After placement the stents were heated in vivo for 60 min at temperatures ranging from 43 to 52°C. Temperature was measured in the surrounding tissue at various distances from the stent, determining heat penetration. The esophagi were histologically examined after 7 days. Results: The maximal heating temperature tolerated in the esophagi without transmural necrosis was 46.5°C, when applied twice for 60 min with a pause of 48 h. With this procedure a tumor damaging temperature of 42.5°C was achieved at a maximum distance of 12 mm surrounding the stent. Conclusion: Application of hyperthermia through a heatable stent in the esophagus is feasible. The maximal depth of therapeutic temperature achieved by this procedure (12 mm) is not sufficient to heal esophageal cancer, but may be of palliative value.
Beschreibung:Gesehen am 11.12.2019
Beschreibung:Online Resource
ISSN:1421-9921
DOI:10.1159/000091526