Quantification of plan robustness against different uncertainty sources for classical and anatomical robust optimized treatment plans in head and neck cancer proton therapy

Objective:Classical robust optimization (cRO) in intensity-modulated proton therapy (IMPT) considers isocenter position and particle range uncertainties; anatomical robust optimization (aRO) aims to consider additional non-rigid positioning variations. This work compares the influence of different u...

Full description

Saved in:
Bibliographic Details
Main Authors: Cubillos Mesías, Macarena (Author) , Troost, Esther Gera Cornelia (Author) , Lohaus, Fabian (Author)
Format: Article (Journal)
Language:English
Published: 2020
In: BJR case reports
Year: 2019, Volume: 93, Issue: 1107
ISSN:2055-7159
DOI:10.1259/bjr.20190573
Online Access:Verlag, lizenzpflichtig, Volltext: https://doi.org/10.1259/bjr.20190573
Verlag, lizenzpflichtig, Volltext: https://www.birpublications.org/doi/10.1259/bjr.20190573
Get full text
Author Notes:Macarena Cubillos-Mesías, MSc, Esther GC Troost, MD, PhD, Fabian Lohaus, MD, Linda Agolli, MD, Maximilian Rehm, MD, Christian Richter, PhD and Kristin Stützer, PhD
Description
Summary:Objective:Classical robust optimization (cRO) in intensity-modulated proton therapy (IMPT) considers isocenter position and particle range uncertainties; anatomical robust optimization (aRO) aims to consider additional non-rigid positioning variations. This work compares the influence of different uncertainty sources on the robustness of cRO and aRO IMPT plans for head and neck squamous cell carcinoma (HNSCC).Methods:Two IMPT plans were optimized for 20 HNSCC patients who received weekly control CTs (cCT): cRO, using solely the planning CT, and aRO, including 2 additional cCTs. The robustness of the plans in terms of clinical target volume (CTV) coverage and organ at risk (OAR) sparing was analyzed considering stepwise the influence of (1) non-rigid anatomical variations given by the weekly cCT, (2) with fraction-wise added rigid random setup errors and (3) additional systematic proton range uncertainties.Results:cRO plans presented significantly higher nominal CTV coverage but are outperformed by aRO plans when considering non-rigid anatomical variations only, as cRO and aRO plans presented a median target coverage (D98%) decrease for the low-risk/high-risk CTV of 1.8/1.1 percentage points (pp) and −0.2 pp/−0.3 pp, respectively. Setup and range uncertainties had larger influence on cRO CTV coverage, but led to similar OAR dose changes in both plans. Considering all error sources, 10/2 cRO/aRO patients missed the CTV coverage and a limited number exceeded some OAR constraints in both plans.Conclusion:Non-rigid anatomical variations are mainly responsible for critical target coverage loss of cRO plans, whereas the aRO approach was robust against such variations. Both plans provide similar robustness of OAR parameters.Advances in knowledge:The influence of different uncertainty sources was quantified for robust IMPT HNSCC plans.
Item Description:Published online: November 28, 2019
Gesehen am 13.05.2020
Physical Description:Online Resource
ISSN:2055-7159