Local dose rate effects in implantable cardioverter-defibrillators with flattening filter free and flattened photon radiation
In the beam penumbra of stereotactic body radiotherapy volumes, dose rate effects in implantable cardioverter-defibrillators (ICDs) may be the predominant cause for failures in the absence of neutron-generating photon energies. We investigate such dose rate effects in ICDs and provide evidence for s...
Gespeichert in:
| Hauptverfasser: | , , , , , |
|---|---|
| Dokumenttyp: | Article (Journal) |
| Sprache: | Englisch |
| Veröffentlicht: |
10 March 2022
|
| In: |
Strahlentherapie und Onkologie
Year: 2022, Jahrgang: 198, Heft: 6, Pages: 566-572 |
| ISSN: | 1439-099X |
| DOI: | 10.1007/s00066-022-01911-8 |
| Online-Zugang: | Verlag, kostenfrei, Volltext: https://doi.org/10.1007/s00066-022-01911-8 Verlag, kostenfrei, Volltext: https://link.springer.com/article/10.1007/s00066-022-01911-8 |
| Verfasserangaben: | Benjamin Gauter-Fleckenstein, Erol Tülümen, Boris Rudic, Martin Borggrefe, Martin Polednik, Jens Fleckenstein |
| Zusammenfassung: | In the beam penumbra of stereotactic body radiotherapy volumes, dose rate effects in implantable cardioverter-defibrillators (ICDs) may be the predominant cause for failures in the absence of neutron-generating photon energies. We investigate such dose rate effects in ICDs and provide evidence for safe use of lung tumor stereotactic radioablation with flattening filter free (FFF) and flattened 6 Megavolt (MV) beams in ICD-bearing patients. |
|---|---|