Hybrid algorithms for SAR matrix compression and the impact of post-processing on SAR calculation complexity

Purpose: This study proposes faster virtual observation point (VOP) compression as well as post-processing algorithms for specific absorption rate (SAR) matrix compression. Furthermore, it shows the relation between the number of channels and the computational burden for VOP-based SAR calculation. M...

Full description

Saved in:
Bibliographic Details
Main Authors: Orzada, Stephan (Author) , Fiedler, Thomas M. (Author) , Ladd, Mark E. (Author)
Format: Article (Journal)
Language:English
Published: December 2024
In: Magnetic resonance in medicine
Year: 2024, Volume: 92, Issue: 6, Pages: 2696-2706
ISSN:1522-2594
DOI:10.1002/mrm.30235
Online Access:Verlag, kostenfrei, Volltext: https://doi.org/10.1002/mrm.30235
Verlag, kostenfrei, Volltext: https://onlinelibrary.wiley.com/doi/abs/10.1002/mrm.30235
Get full text
Author Notes:Stephan Orzada, Thomas M. Fiedler, Mark E. Ladd
Description
Summary:Purpose: This study proposes faster virtual observation point (VOP) compression as well as post-processing algorithms for specific absorption rate (SAR) matrix compression. Furthermore, it shows the relation between the number of channels and the computational burden for VOP-based SAR calculation. Methods: The proposed new algorithms combine the respective benefits of two different criteria for determining upper boundedness of SAR matrices by the VOPs. Comparisons of the old and new algorithms are performed for head coil arrays with various channel counts. The new post-processing algorithm is used to post-process the VOP sets of nine arrays, and the number of VOPs for a fixed median relative overestimation is compared. Results: The new algorithms are faster than the old algorithms by a factor of two to more than 10. The compression efficiency (number of VOPs relative to initial number of SAR matrices) is identical. For a fixed median relative overestimation, the number of VOPs increases logarithmically with the number of RF coil channels when post-processing is applied. Conclusion: The new algorithms are much faster than previous algorithms. Post-processing is very beneficial for online SAR supervision of MRI systems with high channel counts, since for a given number of VOPs the relative SAR overestimation can be lowered.
Item Description:Gesehen am 14.01.2025
Physical Description:Online Resource
ISSN:1522-2594
DOI:10.1002/mrm.30235