Bleeding after cardiovascular surgery caused by detergent residues in laparotomy sponges
On 2 successive days in July 1994, six patients who underwent coronary artery bypass grafting developed postoperative bleeding requiring reexploration. The cause was the use of reprocessed laparotomy sponges with retained detergents, inducing a bleeding disorder.
Gespeichert in:
| Hauptverfasser: | , , |
|---|---|
| Dokumenttyp: | Article (Journal) |
| Sprache: | Englisch |
| Veröffentlicht: |
1997
|
| In: |
Infection control and hospital epidemiology
Year: 1997, Jahrgang: 18, Heft: 8, Pages: 579-581 |
| ISSN: | 1559-6834 |
| DOI: | 10.1086/647677 |
| Online-Zugang: | Verlag, lizenzpflichtig, Volltext: https://doi.org/10.1086/647677 Verlag, lizenzpflichtig, Volltext: https://www.cambridge.org/core/journals/infection-control-and-hospital-epidemiology/article/abs/bleeding-after-cardiovascular-surgery-caused-by-detergent-residues-in-laparotomy-sponges/03FC53C2EE8290610126725EC9F79917# |
| Verfasserangaben: | Heinrich K. Geiss, Judith Schmitt, Stefan C. Frank |
| Zusammenfassung: | On 2 successive days in July 1994, six patients who underwent coronary artery bypass grafting developed postoperative bleeding requiring reexploration. The cause was the use of reprocessed laparotomy sponges with retained detergents, inducing a bleeding disorder. |
|---|---|
| Beschreibung: | Elektronische Reproduktion der Druck-Ausgabe 2. Januar 2015 Gesehen am 31.03.2025 |
| Beschreibung: | Online Resource |
| ISSN: | 1559-6834 |
| DOI: | 10.1086/647677 |