Quality assurance and temporal stability of a 1.5 T MRI scanner for MR-guided photon and particle therapy
Purpose - To describe performance measurements, adaptations and time stability over 20 months of a diagnostic MR scanner for integration into MR-guided photon and particle radiotherapy. - Material and methods - For realization of MR-guided photon and particle therapy (MRgRT/MRgPT), a 1.5 T MR scanne...
Saved in:
| Main Authors: | , , , , , , |
|---|---|
| Format: | Article (Journal) |
| Language: | English |
| Published: |
23 May 2025
|
| In: |
Zeitschrift für medizinische Physik
Year: 2025, Volume: 35, Issue: 2, Pages: 204-217 |
| ISSN: | 1876-4436 |
| DOI: | 10.1016/j.zemedi.2023.04.004 |
| Online Access: | Verlag, kostenfrei, Volltext: https://doi.org/10.1016/j.zemedi.2023.04.004 Verlag, kostenfrei, Volltext: https://www.sciencedirect.com/science/article/pii/S0939388923000466 |
| Author Notes: | Stefan Dorsch, Katharina Paul, Cedric Beyer, Christian P Karger, Oliver Jäkel, Jürgen Debus, Sebastian Klüter |
| Summary: | Purpose - To describe performance measurements, adaptations and time stability over 20 months of a diagnostic MR scanner for integration into MR-guided photon and particle radiotherapy. - Material and methods - For realization of MR-guided photon and particle therapy (MRgRT/MRgPT), a 1.5 T MR scanner was installed at the Heidelberg Ion Beam Therapy Center. To integrate MRI into the treatment process, a flat tabletop and dedicated coil holders for flex coils were used, which prevent deformation of the patient external contour and allow for the use of immobilization tools for reproducible positioning. The signal-to-noise ratio (SNR) was compared for the diagnostic and therapy-specific setup using the flat couch top and flexible coils for the a) head & neck and b) abdominal region as well as for different bandwidths and clinical pulse sequences. Additionally, a quality assurance (QA) protocol with monthly measurements of the ACR phantom and measurement of geometric distortions for a large field-of-view (FOV) was implemented to assess the imaging quality parameters of the device over the course of 20 months. - Results - The SNR measurements showed a decreased SNR for the RT-specific as compared to the diagnostic setup of (a) 26% to 34% and (b) 11% to 33%. No significant bandwidth dependency for this ratio was found. The longitudinal assessment of the image quality parameters with the ACR and distortion phantom confirmed the long-term stability of the MRI device. - Conclusion - A diagnostic MRI was commissioned for use in MR-guided particle therapy. Using a radiotherapy specific setup, a high geometric accuracy and signal homogeneity was obtained after some adaptions and the measured parameters were shown to be stable over a period of 20 months. |
|---|---|
| Item Description: | Online verfügbar 5 May 2023, Version des Artikels 23 May 2025 Gesehen am 23.10.2025 |
| Physical Description: | Online Resource |
| ISSN: | 1876-4436 |
| DOI: | 10.1016/j.zemedi.2023.04.004 |