Oxygenator oxygen transfer performance during cardiopulmonary bypass of the Sorin® Inspire 6F©

Background - Oxygenator oxygen transfer performance is typically measured by manufacturers under standardized, static laboratory conditions. However, these parameters do not adequately reflect the dynamic environment of clinical cardiopulmonary bypass (CPB). To align with modern goal-directed perfus...

Description complète

Enregistré dans:
Détails bibliographiques
Auteurs principaux: Riesterer, David (Auteur) , Eisner, Christoph (Auteur) , Wojdyla, Jacob (Auteur) , Turra, Jan (Auteur)
Format: Article (Journal)
Langue:anglais
Publié: March 24, 2026
In: Perfusion
Year: 2026, Pages: 1-6
ISSN:1477-111X
DOI:10.1177/02676591261438976
Accès en ligne:Verlag, kostenfrei, Volltext: https://doi.org/10.1177/02676591261438976
Accéder au texte intégral
Notes sur l'auteur:David Max Riesterer, Christoph Eisner, Jacob Wojdyla and Jan Turra
Description
Résumé:Background - Oxygenator oxygen transfer performance is typically measured by manufacturers under standardized, static laboratory conditions. However, these parameters do not adequately reflect the dynamic environment of clinical cardiopulmonary bypass (CPB). To align with modern goal-directed perfusion standards - specifically the maintenance of physiological PaO2 levels to reduce oxidative stress - perfusionists require a reliable tool to titrate FiO2 based on real-time oxygen demand. This study aimed to develop a clinical oxygen transfer function for the Sorin® Inspire 6F© oxygenator. - Methods - We retrospectively analyzed 420 blood gas samples and hemodynamic parameters from 54 adult patients undergoing CPB with the Sorin® Inspire 6F©. Oxygen transfer (VO2) was calculated via the Fick principle. Using the methodology established by Hamilton et al., a validation subset (n = 57) of samples within a target PaO2 range of 150 ± 5 mmHg was used to derive a specific correlation function between FiO2 and VO2 via linear regression analysis. - Results - A significant correlation was observed between FiO2 and oxygen transfer. For the validation subset, the derived oxygen transfer function for the Sorin® Inspire 6F© was VO2 = 284.34 x FiO2 (R2 = 0.6185). In contrast, the correlation for the entire clinical dataset (n = 420) yielded an R2 of 0.4583. Comparison with existing literature indicates that the Inspire 6F© has a lower transfer capacity compared to larger membrane surface area oxygenators, as expected. - Conclusions - The newly developed oxygen transfer function provides a practical, evidence-based tool for the clinical setting. It allows perfusionists to adjust the FiO2 accurately to meet patient-specific oxygen requirements and maintain desired PaO2 levels. Implementing this function in routine practice can enhance standardization and improve patient safety by preventing hyperoxygenation.
Description:Zuerst online veröffentlicht: 24. März 2026